A new method to determine the projected coordinate origin of a cone-beam CT system using elliptical projection

In order to determine the projected coordinate origin in the cone-beam CT scanning system with respect to the Feldkamp-Davis-Kress (FDK) algorithm, we propose a simple yet feasible method to accurately measure the projected coordinate origin. This method was established on the basis of the theory that the projection of a spherical object in the cone-beam field is an ellipse. We first utilized image processing and the least square estimation method to get each major axis of the elliptical Digital Radiography (DR) projections of a group of spherical objects. Then we determined the intersection point of the group of major axis by solving an over-determined equation set that was composed by the major axis equations of all the elliptical projections. Based on the experimental results, this new method was proved to be easy to implement in practical scanning systems with high accuracy and anti-noise capability.

[1]  Richard L. Van Metter,et al.  Factors influencing image quality in digital radiographic systems , 2001, SPIE Medical Imaging.

[2]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[3]  Ronald J. Jaszczak,et al.  Determination of both mechanical and electronic shifts in cone beam SPECT , 1992, IEEE Conference on Nuclear Science Symposium and Medical Imaging.

[4]  R Clackdoyle,et al.  Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography. , 2000, Physics in medicine and biology.

[5]  Kai Yang,et al.  A geometric calibration method for cone beam CT systems. , 2006, Medical physics.

[6]  John M Boone,et al.  An improved method for flat-field correction of flat panel x-ray detector. , 2006, Medical physics.

[7]  E. Ritman Micro-computed tomography-current status and developments. , 2004, Annual review of biomedical engineering.

[8]  Ying Hou,et al.  A calibration method for misaligned scanner geometry in cone-beam computed tomography , 2006 .

[9]  G T Gullberg,et al.  Estimation of geometrical parameters and collimator evaluation for cone beam tomography. , 1990, Medical physics.

[10]  Patrick Dupont,et al.  Characterization of pinhole SPECT acquisition geometry , 2003, IEEE Transactions on Medical Imaging.

[11]  Min Hyoung Cho,et al.  A flat-panel detector based micro-CT system: performance evaluation for small-animal imaging. , 2003, Physics in medicine and biology.

[12]  Shu Xiao,et al.  Fast Feldkamp algorithm for cone-beam computer tomography , 2003, Proceedings 2003 International Conference on Image Processing (Cat. No.03CH37429).