System accuracy evaluation of systems for point-of-care testing of blood glucose: a comparison of a patient-use system with six professional-use systems

Abstract Background: Point-of-care testing (POCT) of blood glucose (BG) is performed by medical personnel in clinical settings as well as by patients themselves for self-monitoring of blood glucose (SMBG) at home. We investigated if a system mainly intended for SMBG by people with diabetes, but also suitable for BG measurements by medical personnel, can achieve measurement accuracy on capillary blood samples comparable with professional-use POCT systems. Methods: System accuracy was evaluated under standardised conditions, following the ISO standard 15197:2003. For each system (one SMBG system with three test strip lots and six professional-use systems), measurement results from capillary blood samples of 100 subjects were compared with a standardised laboratory glucose oxidase method (YSI 2300 glucose analyser). Results: The seven evaluated systems showed 99.5% or 100% of the measurement results within the required system accuracy limits of ISO 15197:2003 (±0.83 mmol/L at BG concentrations <4.2 mmol/L and ±20% at BG concentrations ≥4.2 mmol/L). Applying the more stringent requirements of the revision ISO 15197:2013, the systems showed between 99% and 100% of the measurement results within the accuracy limits (±0.83 mmol/L at BG concentrations <5.55 mmol/L and ±15% at BG concentrations ≥5.55 mmol/L) and between 82% and 98% when even more restrictive limits were applied (±0.56 mmol/L and ±10%, respectively). Conclusions: Data from this study, which focused on system accuracy, suggest that SMBG systems can achieve system accuracy that is comparable with professional-use systems when measurements are performed on capillary blood samples by trained personnel in a standardised and controlled setting.

[1]  Boyd,et al.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. , 2002, Clinical chemistry.

[2]  D. Nathan,et al.  Point‐of‐Care Testing for Diabetes , 2002, Critical care nursing quarterly.

[3]  Joyce Y. Wu,et al.  Can one point-of-care glucose meter be used for all pediatric and adult hospital patients? Evaluation of three meters, including recently modified test strips. , 2011, Diabetes technology & therapeutics.

[4]  Frederick L. Kiechle,et al.  Blood Glucose: Measurement in the Point-of-Care Setting , 2000 .

[5]  P. Twomey,et al.  Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640 , 2004, Journal of Clinical Pathology.

[6]  G J Kost,et al.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. , 2000, Archives of pathology & laboratory medicine.

[7]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[8]  S. Sandberg,et al.  Effect of ambient temperature on analytical performance of self-monitoring blood glucose systems. , 2011, Diabetes technology & therapeutics.

[9]  P. Sharp,et al.  Tighter Accuracy Standards within Point-of-Care Blood Glucose Monitoring: How Six Commonly Used Systems Compare , 2012, Journal of diabetes science and technology.

[10]  M. Lyon,et al.  Interference studies with two hospital-grade and two home-grade glucose meters. , 2009, Diabetes technology & therapeutics.

[11]  D. Christensen,et al.  Effect of high altitude on blood glucose meter performance. , 2002, Diabetes technology & therapeutics.

[12]  C. Ostenson,et al.  Performance of glucose dehydrogenase-and glucose oxidase-based blood glucose meters at high altitude and low temperature. , 2005, Diabetes care.

[13]  M. Walkenstein,et al.  Inaccurate glucose determination by fingerstick in a patient with peripheral arterial disease. , 2002, Annals of internal medicine.

[14]  W. Kerner,et al.  Definition, Klassifikation und Diagnostik des Diabetes mellitus , 2006 .

[15]  E. Ipp,et al.  Accuracy of Plasma Glucose Measurements in the Hypoglycemic Range , 1994, Diabetes Care.

[16]  Jochen Sieber,et al.  Determination of Hematocrit Interference in Blood Samples Derived from Patients with Different Blood Glucose Concentrations , 2013, Journal of diabetes science and technology.

[17]  S. Maule,et al.  Effects of exposure at an altitude of 3,000 m on performance of glucose meters. , 2000, Diabetes care.

[18]  Suzanne H. Atkin,et al.  Fingerstick glucose determination in shock. , 1991, Annals of internal medicine.

[19]  T. Veneman,et al.  Comparison of the accuracy of the HemoCue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia. , 2005, European journal of endocrinology.

[20]  W. Bablok,et al.  A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part I , 1983, Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie.

[21]  M. Roden,et al.  Precision and accuracy of blood glucose measurements using three different instruments , 2012, Diabetic medicine : a journal of the British Diabetic Association.

[22]  J. Mahoney,et al.  Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management , 2007, Diabetes Care.

[23]  Lutz Heinemann,et al.  Quality of glucose measurement with blood glucose meters at the point-of-care: relevance of interfering factors. , 2010, Diabetes technology & therapeutics.

[24]  Amy Callahan,et al.  Accuracy of bedside capillary blood glucose measurements in critically ill patients , 2006, Intensive Care Medicine.