BISTRO: An outer‐volume suppression method that tolerates RF field inhomogeneity

A technique is described for performing frequency‐selective signal suppression with a high degree of tolerance to RF field inhomogeneity. The method is called B1‐insensitive train to obliterate signal (BISTRO). BISTRO consists of multiple amplitude‐ and frequency‐modulated (FM) pulses interleaved with spoiler gradients. BISTRO was developed for the purpose of accomplishing band‐selective signal removal, as in water suppression and outer‐volume suppression (OVS), in applications requiring the use of an inhomogeneous RF transmitter, such as a surface coil. In the present work, Bloch simulations were used to illustrate the principles and theoretical performance of BISTRO. Its performance for OVS was evaluated experimentally using MRI and spectroscopic imaging of phantoms and in vivo animal and human brain. By using FM pulses featuring offset‐independent adiabaticity, BISTRO permitted high‐quality, broadband suppression with one (or two) discrete borders demarcating the edge(s) of the suppression band. Simulations and experiments demonstrated the ability to operate BISTRO with reasonably attainable peak RF power levels and with average RF energy deposition similar to other multipulse OVS techniques. Magn Reson Med 45:1095–1102, 2001. © 2001 Wiley‐Liss, Inc.

[1]  K Uğurbil,et al.  Observation of resolved glucose signals in 1H NMR spectra of the human brain at 4 Tesla , 1996, Magnetic resonance in medicine.

[2]  M. Garwood,et al.  Fast broadband inversion by adiabatic pulses. , 1998, Journal of magnetic resonance.

[3]  Minoru Toda,et al.  Springer Series in Solid-State Sciences , 1989 .

[4]  K Ugurbil,et al.  Increase of creatine kinase activity in the visual cortex of human brain during visual stimulation: A 31p NMR magnetization transfer study , 1997, Magnetic resonance in medicine.

[5]  Rolf Gruetter,et al.  Noninvasive Measurements of [1‐13C] Glycogen Concentrations and Metabolism in Rat Brain In Vivo , 1999, Journal of neurochemistry.

[6]  R. B. Kingsley,et al.  WET, a T1- and B1-insensitive water-suppression method for in vivo localized 1H NMR spectroscopy. , 1994, Journal of magnetic resonance. Series B.

[7]  R. Freeman,et al.  Optimized Adiabatic Pulses for Wideband Spin Inversion , 1996 .

[8]  D Matthaei,et al.  1H NMR chemical shift selective (CHESS) imaging. , 1985, Physics in medicine and biology.

[9]  D. Shungu,et al.  Sensitivity and localization enhancement in multinuclear in vivo NMR spectroscopy by outer volume presaturation , 1993, Magnetic resonance in medicine.

[10]  J. Kurhanewicz,et al.  Very selective suppression pulses for clinical MRSI studies of brain and prostate cancer , 2000, Magnetic resonance in medicine.

[11]  A Haase,et al.  Localization of unaffected spins in NMR imaging and spectroscopy (LOCUS spectroscopy) , 1986, Magnetic resonance in medicine.

[12]  M. S. Silver,et al.  Highly selective {π}/{2} and π pulse generation , 1984 .

[13]  G C McKinnon,et al.  Optimized outer volume suppression for single‐shot fast spin‐echo cardiac imaging , 1998, Journal of magnetic resonance imaging : JMRI.

[14]  Pines,et al.  Broadband and adiabatic inversion of a two-level system by phase-modulated pulses. , 1985, Physical review. A, General physics.

[15]  W. Brooks,et al.  The utilization of two frequency‐shifted sinc pulses for performing volume‐selected in vivo NMR spectroscopy , 1986, Magnetic resonance in medicine.

[16]  M. Garwood,et al.  B1-insensitive, single-shot localization and water suppression. , 1996, Journal of magnetic resonance. Series B.

[17]  J H Duyn,et al.  Multisection proton MR spectroscopic imaging of the brain. , 1993, Radiology.

[18]  M. Garwood,et al.  Adiabatic pulses , 1997, NMR in biomedicine.

[19]  M. Garwood,et al.  A new localization method using an adiabatic pulse, BIR-4. , 1995, Journal of magnetic resonance. Series B.

[20]  G. Adriany,et al.  A half-volume coil for efficient proton decoupling in humans at 4 tesla. , 1997, Journal of magnetic resonance.

[21]  Brian K. Rutt,et al.  Projection presaturation: A fast and accurate technique for multidimensional spatial localization , 1990 .

[22]  K. Nicolay,et al.  In vivo 1H MR spectroscopic imaging and difusion weighted mri in experimental hydrocephalus , 1998, Magnetic resonance in medicine.

[23]  M. Garwood,et al.  Localized detection of glioma glycolysis using edited 1H MRS , 1993, Magnetic resonance in medicine.

[24]  H. Merkle,et al.  Spatially localized in vivo 1H magnetic resonance spectroscopy of an intracerebral rat glioma , 1992, Magnetic resonance in medicine.

[25]  W. R. Brody,et al.  Projection Presaturation. III. Accurate Selective Excitation or Presaturation of the Regions of Tailored Shape in the Presence of Short-T1 Species , 1993 .

[26]  Michael Garwood,et al.  Improved Performance of Frequency-Swept Pulses Using Offset-Independent Adiabaticity , 1996 .

[27]  M. Garwood,et al.  Asymmetric adiabatic pulses for NH selection. , 1999, Journal of magnetic resonance.

[28]  Brian K. Rutt,et al.  Projection presaturation. II. Single-shot localization of multiple regions of interest , 1990 .