Clinical Utility of a Fingerstick Technology to Identify Individuals with Abnormal Blood Lipids and High-Sensitivity C-Reactive Protein Levels

Purpose. The purpose of this study was to determine the ability of a commonly used fingerstick technology to identify individuals with abnormal blood levels of total cholesterol (TC), calculated low-density lipoprotein cholesterol (LDL-C), high-density lipoprotein cholesterol (HDL-C), triglycerides (TG), and high-sensitivity C-reactive protein (hsCRP) compared with a standardized laboratory. Methods. Participants (n = 250; mean age, 48.0 + 13.5 years; 66% female; 36% nonwhite) were eligible for primary prevention of cardiovascular disease (CVD). Blood lipids and hsCKP were measured simultaneously by (1) fingerstick analyzed by Cholestech LDX analyzers and (2) fresh venous blood that was analyzed by Columbia University General Clinical Research Center (GCRC) Core Laboratory. Pearson correlation coefficients, kappa, sensitivity, and specificity were calculated for fingerstick versus GCRC laboratory values for lipids and hsCRP. Results. The correlations between fingerstick and core laboratory for TC, LDL-C, HDL-C, TG, and hsCRP were .91, .88, .77, .93, and .81, respectively (all p < .01). Sensitivity and specificity of the fingerstick to identify those with abnormal lipids and hsCRP ≥ 1 mg/L were all ≥ 75%. Conclusion. Fingerstick screening is accurate and has good clinical utility to identify persons with abnormal blood lipids and hsCRP at the point of care in a diverse population that is eligible for primary prevention of CVD. These results may not be generalizable to patients at high risk for CVD or who have known hyperlipidemia.

[1]  P. Bachorik,et al.  National Cholesterol Education Program recommendations for measurement of low-density lipoprotein cholesterol: executive summary. The National Cholesterol Education Program Working Group on Lipoprotein Measurement. , 1995, Clinical chemistry.

[2]  Sidney C. Smith,et al.  MARKERS OF INFLAMMATION AND CARDIOVASCULAR DISEASE: APPLICATION TO CLINICAL AND PUBLIC HEALTH PRACTICE: A STATEMENT FOR HEALTHCARE PROFESSIONALS FROM THE CENTERS FOR DISEASE CONTROL AND PREVENTION AND THE AMERICAN HEART ASSOCIATION , 2003 .

[3]  J. Lindemans,et al.  Evaluation of the Cholestech L.D.X. desktop analyser for cholesterol, HDL-cholesterol, and triacylglycerols in heparinized venous blood. , 1994, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies.

[4]  I. Ockene,et al.  Evaluation of seven Cholestech L.D.X analyzers for total cholesterol determinations. , 1993, Clinical chemistry.

[5]  K. Furie,et al.  Heart disease and stroke statistics--2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. , 2007, Circulation.

[6]  C. Carlsson,et al.  Inaccuracy of lipid measurements with the portable Cholestech L.D.X analyzer in patients with hypercholesterolemia. , 2002, Clinical chemistry.

[7]  K. Furie,et al.  Heart disease and stroke statistics--2007 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. , 2008, Circulation.

[8]  R. Bard,et al.  Evaluation of lipid profile measurements obtained from the Cholestech L.D.X analyzer. , 1997, Journal of cardiopulmonary rehabilitation.

[9]  J. Mckenney,et al.  National Cholesterol Education Program Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III) , 2002 .

[10]  F. Raal,et al.  Performance of the CardioChek PA and Cholestech LDX point-of-care analysers compared to clinical diagnostic laboratory methods for the measurement of lipids. , 2005, Cardiovascular journal of South Africa : official journal for Southern Africa Cardiac Society [and] South African Society of Cardiac Practitioners.

[11]  D. Mozaffarian,et al.  Heart disease and stroke statistics--2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. , 2009, Circulation.

[12]  G. Boran,et al.  Validation of a point of care lipid analyser using a hospital based reference laboratory , 2006, Irish journal of medical science.