Microscopic investigation of erythrocyte deformation dynamics.

The understanding of erythrocyte deformation under conditions of high shear stress and short exposure time is central to the study of hemorheology and hemolysis within prosthetic blood contacting devices. A combined computational and experimental microscopic study was conducted to investigate the erythrocyte deformation and its relation to transient stress fields. A microfluidic channel system with small channels fabricated using polydimethylsiloxane on the order of 100 mum was designed to generate transient stress fields through which the erythrocytes were forced to flow. The shear stress fields were analyzed by three-dimensional computational fluid dynamics. Microscopic images of deforming erythrocytes were experimentally recorded to obtain the changes in cell morphology over a wide range of fluid dynamic stresses. The erythrocyte elongation index (EI) increased from 0 to 0.54 with increasing shear stress up to 123 Pa. In this shear stress range, erythrocytes behaved like fluid droplets, and deformed and flowed following the surrounding fluid. Cells exposed to shear stress beyond 123 Pa (up to 5170 Pa) did not exhibit additional elongation beyond EI=0.54. Two-stage deformation of erythrocytes in response to shear stress was observed: an initial linear elongation with increasing shear stress and a plateau beyond a critical shear stress.

[1]  Howard A. Stone,et al.  Dynamics of Drop Deformation and Breakup in Viscous Fluids , 1994 .

[2]  Mechanical haemolysis in shock wave lithotripsy (SWL): I. Analysis of cell deformation due to SWL flow-fields. , 2001, Physics in medicine and biology.

[3]  G Laufer,et al.  First clinical experience with the DeBakey VAD continuous-axial-flow pump for bridge to transplantation. , 2000, Circulation.

[4]  R. Paul,et al.  Shear stress related blood damage in laminar couette flow. , 2003, Artificial organs.

[5]  J F Antaki,et al.  Computational fluid dynamics as a development tool for rotary blood pumps. , 2001, Artificial organs.

[6]  R. Waugh,et al.  The cooperative role of membrane skeleton and bilayer in the mechanical behaviour of red blood cells. , 2004, Bioelectrochemistry.

[7]  G. Caimi,et al.  Techniques to evaluate erythrocyte deformability in diabetes mellitus , 2004, Acta Diabetologica.

[8]  Sehyun Shin,et al.  Rapid cell-deformability sensing system based on slit-flow laser diffractometry with decreasing pressure differential. , 2005, Biosensors & bioelectronics.

[9]  Nelson J. Groom,et al.  Development Progress of the University of Pittsburgh Streamliner: a Mixed Flow Blood Pump with Magnetic Bearings , 2000 .

[10]  J F Antaki,et al.  Decrease in red blood cell deformability caused by hypothermia, hemodilution, and mechanical stress: factors related to cardiopulmonary bypass. , 1999, ASAIO journal.

[11]  G. Whitesides,et al.  Rapid Prototyping of Microfluidic Systems in Poly(dimethylsiloxane). , 1998, Analytical chemistry.

[12]  E. Evans,et al.  Structure and deformation properties of red blood cells: concepts and quantitative methods. , 1989, Methods in enzymology.

[13]  J D Hellums,et al.  Red blood cell damage by shear stress. , 1972, Biophysical journal.

[14]  M Grigioni,et al.  A discussion on the threshold limit for hemolysis related to Reynolds shear stress. , 1999, Journal of biomechanics.

[15]  C. Lim,et al.  Mechanics of the human red blood cell deformed by optical tweezers , 2003 .

[16]  Y. Fung,et al.  Biomechanics: Mechanical Properties of Living Tissues , 1981 .

[17]  Mitsuo Umezu,et al.  Fluid Dynamic Investigation of Mechanical Blood Hemolysis , 1996 .

[18]  C Bludszuweit,et al.  Model for a general mechanical blood damage prediction. , 1995, Artificial organs.

[19]  T. Sugiura,et al.  Evaluation of intravascular hemolysis with erythrocyte creatine in patients with cardiac valve prostheses. , 2004, Chest.

[20]  André A. Adams,et al.  Cell transport via electromigration in polymer-based microfluidic devices. , 2004, Lab on a chip.

[21]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[22]  H. Madarame,et al.  Velocity measurement of both red blood cells and plasma of in vitro blood flow using high-speed micro PIV technique , 2005 .

[23]  A. Pries,et al.  Rheology of the microcirculation. , 2003, Clinical hemorheology and microcirculation.

[24]  A. Popel,et al.  A two-phase model for flow of blood in narrow tubes with increased effective viscosity near the wall. , 2001, Biorheology.

[25]  K. Jensen,et al.  A microfluidic electroporation device for cell lysis. , 2005, Lab on a chip.

[26]  J. Stoltz,et al.  Measurement of erythrocyte deformability by two laser diffraction methods. , 1999, Clinical hemorheology and microcirculation.

[27]  C. Allman,et al.  An unusual cause of hemolysis in a patient with an aortic valved conduit replacement. , 2003, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[28]  R. Opitz,et al.  A Couette viscometer for short time shearing of blood. , 1980, Biorheology.

[29]  J. Kutter,et al.  MICRO PIV ON BLOOD FLOW IN A MICROCHANNEL , 2003 .

[30]  S. Hénon,et al.  A new determination of the shear modulus of the human erythrocyte membrane using optical tweezers. , 1999, Biophysical journal.

[31]  Dominique Barthès-Biesel,et al.  Effect of constitutive laws for two-dimensional membranes on flow-induced capsule deformation , 2002, Journal of Fluid Mechanics.

[32]  M. Williams,et al.  Hemolysis in capillary flow. , 1970, The Journal of laboratory and clinical medicine.

[33]  F. White Viscous Fluid Flow , 1974 .

[34]  Kyung Hyun Ahn,et al.  Shear induced damage of red blood cells monitored by the decrease of their deformability , 2004 .

[35]  E. Evans,et al.  Intrinsic material properties of the erythrocyte membrane indicated by mechanical analysis of deformation , 1975 .

[36]  O. Baskurt,et al.  Analyzing shear stress-elongation index curves: comparison of two approaches to simplify data presentation. , 2004, Clinical hemorheology and microcirculation.

[37]  E. Evans,et al.  New membrane concept applied to the analysis of fluid shear- and micropipette-deformed red blood cells. , 1973, Biophysical journal.