Accuracy of cortical and trabecular bone measurements with peripheral quantitative computed tomography (pQCT).

In order to assess the accuracy of peripheral QCT (Stratec XCT 960) we analysed scans of the European Forearm Phantom and another phantom consisting of K2HPO4 encased in aluminium tubes to simulate cortical walls. Additionally 14 cadaveric forearm specimen scans were compared to CT scans acquired on a GE9800Q. The accuracy for density assessment of the European Forearm Phantom was better than 3%. A small increase in density was observed with increasing thickness of the aluminium wall (10% for each mm). Density measurements within the wall were confounded by limited spatial resolution. For a thickness of less than 4 mm, the density within the wall was underestimated by up to 40%. The measurement of mineral content was not influenced by this effect and showed an accuracy error of less than 6%. The agreement of density measurements on the different CT systems was very strong (R2 > 0.96; RMSE < 6.2%). Our findings suggest that the Stratec pQCT scanner very accurately measures volumetric trabecular and total bone mineral densities at the distal radius while the assessment of cortical density is associated with considerable inaccuracies due to limited spatial resolution.

[1]  P M Joseph,et al.  The exponential edge-gradient effect in x-ray computed tomography. , 1981, Physics in medicine and biology.

[2]  P. Rüegsegger,et al.  In vivo reproducibility of three‐dimensional structural properties of noninvasive bone biopsies using 3D‐pQCT , 1996, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[3]  M. Osteaux,et al.  Cortical and total bone mineral content of the radius: accuracy of peripheral computed tomography. , 1996, Bone.

[4]  M. Bouxsein,et al.  Tibial ultrasound velocity measured in situ predicts the material properties of tibial cortical bone. , 1997, Bone.

[5]  A. Miyauchi,et al.  Transmenopausal change of trabecular bone density and structural pattern assessed by peripheral quantitative computed tomography in Japanese women , 1995, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[6]  H. Genant,et al.  Impact of marrow fat on accuracy of quantitative CT. , 1989, Journal of computer assisted tomography.

[7]  M. Osteaux,et al.  Cortical mineral content of the radius assessed by peripheral QCT predicts compressive strength on biomechanical testing. , 1995, Bone.

[8]  D. Reid,et al.  Appendicular measurements in screening women for low axial bone mineral density. , 1996, The British journal of radiology.

[9]  A G Patwardhan,et al.  Cross-sectional geometrical properties and bone mineral contents of the human radius and ulna. , 1993, Journal of biomechanics.

[10]  P Rüegsegger,et al.  Quantification of bone mineralization using computed tomography. , 1976, Radiology.

[11]  B. Allolio,et al.  Multicenter German reference data base for peripheral quantitative computer tomography. , 1995, Technology and health care : official journal of the European Society for Engineering and Medicine.

[12]  T N Hangartner Correction of scatter in computed tomography images of bone. , 1987, Medical physics.

[13]  H K Genant,et al.  Assessment of bone mineral at appendicular sites in females with fractures of the proximal femur. , 1998, Bone.

[14]  M. Jergas,et al.  Assessment of the skeletal status by peripheral quantitative computed tomography of the forearm: Short‐term precision in vivo and comparison to dual X‐ray absorptiometry , 1995, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[15]  B. Allolio,et al.  Determinants of Radial Bone Density as Measured by PQCT in Pre‐ and Postmenopausal Women: The Role of Bone Size , 1997, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[16]  C C Glüer,et al.  Comparisons of Noninvasive Bone Mineral Measurements in Assessing Age‐Related Loss, Fracture Discrimination, and Diagnostic Classification , 1997, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[17]  G. Wheeler,et al.  Bone mass measurements in the distal forearm using dual‐energy X‐ray absorptiometry and γ‐ray computed tomography: A longitudinal, in vivo comparative study , 1992, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[18]  S Rathee,et al.  A special purpose x-ray fan-beam CT scanner for trabecular bone density measurement in the appendicular skeleton. , 1990, Physics in medicine and biology.

[19]  W. Kalender,et al.  A phantom for standardization and quality control in peripheral bone measurements by PQCT and DXA , 1993 .

[20]  J. Adachi,et al.  In vivo assessment of trabecular bone structure at the distal radius from high-resolution computed tomography images. , 1996, Physics in medicine and biology.