Development of a multifrequency conductance catheter-based system to determine LV function in mice.

Transgenic mice offer a valuable way to relate gene products to phenotype, but the ability to assess the cardiovascular phenotype with pressure-volume analysis has lagged. Conductance measurement offers a method to generate an instantaneous left ventricular (LV) volume signal in the mouse but has been limited by the volume signal being a combination of blood and LV muscle. We hypothesized that by developing a mouse conductance system that operates at several simultaneous frequencies, we could identify and correct for the myocardial contribution to the instantaneous volume signal. This hypothesis is based on the assumption that mouse myocardial conductivity will vary with frequency, whereas mouse blood conductivity will not. Consistent with this hypothesis, we demonstrated that at higher excitation frequency, greater end-diastolic and end-systolic conductance are detected, as well as a smaller difference between the two. We then empirically solved for LV blood volume using two frequencies. We combined measured resistivity of mouse myocardium with an analytic approach and extracted an estimate of LV blood volume from the raw conductance signal. Development of a multifrequency catheter-based system to determine LV function could be a tool to assess cardiovascular phenotype in transgenic mice.

[1]  S A Glantz,et al.  Left ventricular volume measurement by conductance catheter in intact dogs. Parallel conductance volume depends on left ventricular size. , 1989, Circulation.

[2]  H. Schwan,et al.  Specific Resistance of Body Tissues , 1956, Circulation research.

[3]  C. H. Chen,et al.  Special Communication , 2004 .

[4]  B Buis,et al.  Continuous measurement of left ventricular volume in animals and humans by conductance catheter. , 1984, Circulation.

[5]  T. Skalak,et al.  Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance , 1997, IEEE Transactions on Biomedical Engineering.

[6]  A. Feldman,et al.  New insight into the role of enhanced adrenergic receptor-effector coupling in the heart. , 1999, Circulation.

[7]  S. Rush,et al.  Resistivity of Body Tissues at Low Frequencies , 1963, Circulation research.

[8]  L. Mickleborough,et al.  Conductance catheter measurements of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size. , 1994, Cardiovascular research.

[9]  M. F. Fuller,et al.  Practical Nonparametric Statistics; Nonparametric Statistical Inference , 1973 .

[10]  S. Weidmann Electrical constants of trabecular muscle from mammalian heart , 1970, The Journal of physiology.

[11]  J. G. Webster,et al.  Impedance of Skeletal Muscle from 1 Hz to 1 MHz , 1984, IEEE Transactions on Biomedical Engineering.

[12]  D. Teitel,et al.  The Conductance Volume Catheter Technique for Measurement of Left Ventricular Volume in Young Piglets , 1992, Pediatric Research.

[13]  T J Gawne,et al.  Estimating left ventricular offset volume using dual-frequency conductance catheters. , 1987, Journal of applied physiology.

[14]  N Reichek,et al.  Reliable estimation of peak left ventricular systolic pressure by M-mode echographic-determined end-diastolic relative wall thickness: identification of severe valvular aortic stenosis in adult patients. , 1982, American heart journal.

[15]  D. Kass,et al.  Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique. , 1986, Circulation.

[16]  B. Lofgren The electrical impedance of a complex tissue and its relation to changes in volume and fluid distribution; a study on rat kidneys. , 1951, Acta physiologica Scandinavica. Supplementum.

[17]  K. Chien,et al.  Physiological assessment of complex cardiac phenotypes in genetically engineered mice. , 1997, The American journal of physiology.

[18]  A. Redington,et al.  Left ventricular parallel conductance during cardiac cycle in children with congenital heart disease. , 1997, American Journal of Physiology.

[19]  K R Foster,et al.  Myocardial Electrical Impedance Mapping of Ischemic Sheep Hearts and Healing Aneurysms , 1993, Circulation.

[20]  H. Spotnitz,et al.  Electrical Isolation of the Heart Stabilizing Parallel Conductance for Left Ventricular Volume Measurement , 1997, ASAIO journal.

[21]  D. Laurent,et al.  Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within cardiac cycle. , 1995, The American journal of physiology.

[22]  D. Kass,et al.  Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts. , 1985, Circulation.

[23]  P. Steendijk,et al.  The four-electrode resistivity technique in anisotropic media: theoretical analysis and application on myocardial tissue in vivo , 1993, IEEE Transactions on Biomedical Engineering.

[24]  J. Laragh,et al.  Patterns of left ventricular hypertrophy and geometric remodeling in essential hypertension. , 1992, Journal of the American College of Cardiology.

[25]  A A Shoukas,et al.  Does volume catheter parallel conductance vary during a cardiac cycle? , 1990, The American journal of physiology.

[26]  T E Hewett,et al.  Cardiac physiology in transgenic mice. , 1998, Circulation research.

[27]  Ronald S. Newbower,et al.  A Practical Analysis of the Electrical Conductivity of Blood , 1983, IEEE Transactions on Biomedical Engineering.

[28]  Jan Baan,et al.  Computation of the Input Impedances of a Catheter for Cardiac Volumetry , 1984, IEEE Transactions on Biomedical Engineering.