Transcutaneous RF-Powered Implantable Minipump Driven by a Class-E Transmitter

We describe the design and testing of an inductive coupling system used to power an implantable minipump for applications in ambulating rats. A 2 MHz class-E oscillator driver powered a coil transmitter wound around a 33-cm-diameter rat cage. A receiver coil, a filtered rectifier, and a voltage-sensitive switch powered the implant. The implant DC current at the center of the primary coil (5.1 V) exceeded the level required to activate the solenoid valve in the pump. The variations of the implant current in the volume of the primary coil reflected the variations of the estimated coupling coefficient between the two coils. The pump could be activated in-vivo, while accommodating the vertical and horizontal movements of the animal. Advantages of this design include a weight reduction for the implant, an operation independent from a finite power source, and a remote activation/deactivation

[1]  Derek A. Linkens,et al.  Hierarchical rule-based and self-organizing fuzzy logic control for depth of anaesthesia , 1999, IEEE Trans. Syst. Man Cybern. Part C.

[2]  Nigel H. Lovell,et al.  CMOS neurostimulation ASIC with 100 channels, scaleable output, and bidirectional radio-frequency telemetry , 2001, IEEE Transactions on Biomedical Engineering.

[3]  O. Scremin,et al.  An implantable bolus infusion pump for use in freely moving, nontethered rats. , 2002, American journal of physiology. Heart and circulatory physiology.

[4]  C.M. Zierhofer,et al.  High-efficiency coupling-insensitive transcutaneous power and data transmission via an inductive link , 1990, IEEE Transactions on Biomedical Engineering.

[5]  C.M. Zierhofer,et al.  Geometric approach for coupling enhancement of magnetically coupled coils , 1996, IEEE Transactions on Biomedical Engineering.

[6]  J. Yang,et al.  Activation of cerebral cortex during acoustic challenge or acute foot-shock in freely moving, nontethered rats , 2004, Neuroscience Letters.

[7]  P. Dimbylow FDTD calculations of the whole-body averaged SAR in an anatomically realistic voxel model of the human body from 1 MHz to 1 GHz. , 1997, Physics in medicine and biology.

[8]  N. de N. Donaldson,et al.  Analysis of resonant coupled coils in the design of radio frequency transcutaneous links , 1983, Medical and Biological Engineering and Computing.

[9]  No Sokal,et al.  CLASS-E - NEW CLASS OF HIGH-EFFICIENCY TUNED SINGLE-ENDED SWITCHING POWER AMPLIFIERS , 1975 .

[10]  O. Scremin,et al.  Functional Brain Mapping in Freely Moving Rats during Treadmill Walking , 2003, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[11]  F G Shellock,et al.  Radiofrequency Energy‐Induced Heating During MR Procedures: A Review , 2000, Journal of magnetic resonance imaging : JMRI.