Development of a New Disposable Pulsatile Pump for Cardiopulmonary Bypass: Computational Fluid-Dynamic Design and In Vitro Tests

A newly conceived blood pump for pulsatile cardiopulmonary bypass (CPB) is presented. The new device’s main design features (fully disposable pumping head with ring shaped valves) were intended to overcome the factors that today limit the use of pulsatile blood pumps, i.e., the complexity and costs of devices. The pump was designed and analyzed by means of three-dimensional computational models, including solid computer assisted design of the pumping head and computational fluid-dynamic (CFD) analyses of the fluid domain and of its interaction with deformable components. A prototype of the device, integrated with the venous reservoir, was built to perform hydraulic in vitro tests with aims of both validating CFD results and verifying the new device’s pumping behavior. Functional evaluation of the pump was carried out by using the device in a model circuit made with standard CPB components plus a mock hydraulic bench representing an adult patient’s systemic circulation. A roller pump used in pulsatile mode (RP-PM) was used for comparison. At a 5 L/min flow rate, the pulsatile hydraulic power () delivered to the patient was ∼15 mW for the RP-PM. The new pump proved to be able to deliver up to 40 mW, thus providing a more physiological condition, closer to the delivered by the natural heart (90–140 mW).

[1]  G Pennati,et al.  Influence of Membrane Oxygenators on the Pulsatile flow in Extracorporeal Circuits: An Experimental Analysis , 1997, The International journal of artificial organs.

[2]  G Wright,et al.  The assessment of pulsatile blood flow , 1995, Perfusion.

[3]  G B Fiore,et al.  The pumping oxygenator: design criteria and first in vitro results. , 2000, Artificial organs.

[4]  R Fumero,et al.  Clinical experience with a new pulsatile pump for infant and pediatric cardiopulmonary bypass. , 1989, The International journal of artificial organs.

[5]  A. Furness,et al.  PULSATILE PUMPS FOR OPEN-HEART SURGERY , 1978, The Lancet.

[6]  Y Nosé,et al.  The hemolysis test of the Gyro C1E3 pump in pulsatile mode. , 1997, Artificial organs.

[7]  JJ Driessen,et al.  Pulsatile compared with nonpulsatile perfusion using a centrifugal pump for cardiopulmonary bypass during coronary artery bypass grafting. Effects on systemic haemodynamics, oxygenation, and inflammatory response parameters , 1995, Perfusion.

[8]  Jp. Squifflet,et al.  Ganciclovir for Cytomegalo-virus Infection in Renal-transplant Recipients , 1989 .

[9]  M Endo,et al.  Clinical evaluation of pulsatile flow mode of Terumo Capiox centrifugal pump. , 1997, Artificial organs.

[10]  S. E. Ottmers,et al.  Design of a Physiologic Pulsatile Flow Cardiopulmonary Bypass System for Neonates and Infants , 1996, The International journal of artificial organs.

[11]  P. Hickey,et al.  Pulsatile and nonpulsatile cardiopulmonary bypass: review of a counterproductive controversy. , 1983, The Annals of thoracic surgery.

[12]  Yüksel Atay,et al.  Effects of flow types in cardiopulmonary bypass on gastric intramucosal pH , 1998, Perfusion.

[13]  G Wright,et al.  Computation of haemodynamic power and input impedance in the ascending aorta of patients undergoing open heart surgery. , 1988, Cardiovascular research.

[14]  R B Shepard,et al.  Energy equivalent pressure. , 1966, Archives of surgery.

[15]  J R Montiès,et al.  Experimental use of a semipulsatile rotary blood pump for cardiopulmonary bypass. , 1995, Artificial organs.

[16]  A. Stammers,et al.  Clinical comparison of pulsatile and nonpulsatile perfusion during cardiopulmonary bypass. , 1997, The journal of extra-corporeal technology.

[17]  S Sheppard,et al.  Pulsatile flow during cardiopulmonary bypass speeds thermal energy transfer: a possible explanation for the reduced afterdrop , 1995, Perfusion.

[18]  F Dexter,et al.  Pulsatile Versus Nonpulsatile Flow: No Difference in Cerebral Blood Flow or Metabolism during Normothermic Cardiopulmonary Bypass in Rabbits , 1995, Anesthesiology.

[19]  E. Grossi,et al.  Pulsatile flow. , 1985, The Annals of thoracic surgery.

[20]  A Undar,et al.  Balloon pump-induced pulsatile perfusion during cardiopulmonary bypass does not improve brain oxygenation. , 2000, The Journal of thoracic and cardiovascular surgery.

[21]  Gordon Wright,et al.  The hydraulic power outputs of pulsatile and nonpulsatile cardiopulmonary bypass pumps , 1988 .

[22]  G K Jett Physiology of nonpulsatile circulation: acute versus chronic support. , 1999, ASAIO journal.

[23]  A Undar,et al.  Evaluation of a physiologic pulsatile pump system for neonate-infant cardiopulmonary bypass support. , 1999, ASAIO journal.

[24]  R. Beissinger,et al.  Low‐stress hemolysis in laminar blood flow: Bulk and surface effects in capillaries , 1987 .

[25]  G Wright,et al.  Hemodynamic analysis could resolve the pulsatile blood flow controversy. , 1994, The Annals of thoracic surgery.

[26]  A Undar,et al.  The type of aortic cannula and membrane oxygenator affect the pulsatile waveform morphology produced by a neonate-infant cardiopulmonary bypass system in vivo. , 1998, Artificial organs.

[27]  A Undar,et al.  Effects of perfusion mode on regional and global organ blood flow in a neonatal piglet model. , 1999, The Annals of thoracic surgery.

[28]  R Fumero,et al.  A new pulsatile blood pump for adult cardiopulmonary bypass: design criteria and preliminary fluid dynamic evaluation. , 1996, The International journal of artificial organs.

[29]  T McCarthy,et al.  Long-term blood pumping: an overview , 1995, Perfusion.

[30]  F Iwaya,et al.  In vitro evaluation of a pulsatile assist device for a centrifugal pump using a new principle. , 1995, Artificial organs.

[31]  S. Saito,et al.  Balloon pump-induced pulsatile perfusion during cardiopulmonary bypass does not improve brain oxygenation. , 1999, The Journal of thoracic and cardiovascular surgery.

[32]  K. Nakata,et al.  Cytokine and endothelial damage in pulsatile and nonpulsatile cardiopulmonary bypass. , 1999, Artificial organs.

[33]  R. Koerfer,et al.  The influence of pulsatile and nonpulsatile extracorporeal circulation on fluid retention following coronary artery bypass grafting , 1992, Perfusion.