Dosimetry of cone-defined stereotactic radiosurgery fields with a commercial synthetic diamond detector.

PURPOSE Small field x-ray beam dosimetry is difficult due to lack of lateral electronic equilibrium, source occlusion, high dose gradients, and detector volume averaging. Currently, there is no single definitive detector recommended for small field dosimetry. The objective of this work was to evaluate the performance of a new commercial synthetic diamond detector, namely, the PTW 60019 microDiamond, for the dosimetry of small x-ray fields as used in stereotactic radiosurgery (SRS). METHODS Small field sizes were defined by BrainLAB circular cones (4-30 mm diameter) on a Novalis Trilogy linear accelerator and using the 6 MV SRS x-ray beam mode for all measurements. Percentage depth doses (PDDs) were measured and compared to an IBA SFD and a PTW 60012 E diode. Cross profiles were measured and compared to an IBA SFD diode. Field factors, ΩQclin,Qmsr (fclin,fmsr) , were calculated by Monte Carlo methods using BEAMnrc and correction factors, kQclin,Qmsr (fclin,fmsr) , were derived for the PTW 60019 microDiamond detector. RESULTS For the small fields of 4-30 mm diameter, there were dose differences in the PDDs of up to 1.5% when compared to an IBA SFD and PTW 60012 E diode detector. For the cross profile measurements the penumbra values varied, depending upon the orientation of the detector. The field factors, ΩQclin,Qmsr (fclin,fmsr) , were calculated for these field diameters at a depth of 1.4 cm in water and they were within 2.7% of published values for a similar linear accelerator. The corrections factors, kQclin,Qmsr (fclin,fmsr) , were derived for the PTW 60019 microDiamond detector. CONCLUSIONS The authors conclude that the new PTW 60019 microDiamond detector is generally suitable for relative dosimetry in small 6 MV SRS beams for a Novalis Trilogy linear equipped with circular cones.

[1]  J. Seuntjens,et al.  A new formalism for reference dosimetry of small and nonstandard fields. , 2008, Medical physics.

[2]  W. Curran,et al.  Dosimetric performance of the new high‐definition multileaf collimator for intracranial stereotactic radiosurgery , 2010, Journal of applied clinical medical physics.

[3]  Hui Yan,et al.  Dosimetric characteristics of novalis Tx system with high definition multileaf collimator. , 2008, Medical physics.

[4]  Arnulfo Martínez-Dávalos,et al.  Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  J. Conway,et al.  CVD diamond detectors as dosimeters for radiotherapy , 1997 .

[6]  I. Kawrakow,et al.  Technical note: overprediction of dose with default PRESTA-I boundary crossing in DOSXYZnrc and BEAMnrc. , 2007, Medical physics.

[7]  L D Lunsford,et al.  Stereotactic radiosurgery for arteriovenous malformations of the brain. , 1991, Journal of neurosurgery.

[8]  Natalka Suchowerska,et al.  Small field diode correction factors derived using an air core fibre optic scintillation dosimeter and EBT2 film , 2012, Physics in medicine and biology.

[9]  D I Thwaites,et al.  Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterization , 2012, Physics in medicine and biology.

[10]  P Francescon,et al.  Monte Carlo simulated correction factors for output factor measurement with the CyberKnife system—results for new detectors and correction factor dependence on measurement distance and detector orientation , 2014, Physics in medicine and biology.

[11]  M.,et al.  Spinal lesions treated with Novalis shaped beam intensity-modulated radiosurgery and stereotactic radiotherapy , 2004 .

[12]  T Kairn,et al.  A practical and theoretical definition of very small field size for radiotherapy output factor measurements. , 2014, Medical physics.

[13]  A. Chalkley,et al.  Evaluation of a synthetic single-crystal diamond detector for relative dosimetry measurements on a CyberKnife. , 2014, The British journal of radiology.

[14]  M. Bullock,et al.  Normobaric hyperoxia--induced improvement in cerebral metabolism and reduction in intracranial pressure in patients with severe head injury: a prospective historical cohort-matched study. , 2004, Journal of neurosurgery.

[15]  I. Das,et al.  Small fields: nonequilibrium radiation dosimetry. , 2007, Medical physics.

[16]  Z. Kuncic,et al.  Practical considerations for reporting surface dose in external beam radiotherapy: a 6 MV X-ray beam study , 2012, Australasian Physical & Engineering Sciences in Medicine.

[17]  Carlo Cavedon,et al.  Application of a Monte Carlo‐based method for total scatter factors of small beams to new solid state micro‐detectors , 2009, Journal of applied clinical medical physics.

[18]  L D Lunsford,et al.  Stereotactic radiosurgery of meningiomas. , 1991, Journal of neurosurgery.

[19]  T. Kairn,et al.  A comparison of surface doses for very small field size x-ray beams: Monte Carlo calculations and radiochromic film measurements , 2014, Australasian Physical & Engineering Sciences in Medicine.

[20]  Josep Sempau,et al.  Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: a PENELOPE Monte Carlo study. , 2014, Medical physics.

[21]  S Derreumaux,et al.  Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones. , 2013, Medical physics.

[22]  David I Thwaites,et al.  Head scatter factors for small MV photon fields. Part II: the effects of source size and detector. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  S. Dieterich,et al.  Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors. , 2011, Medical physics.

[24]  C. De Wagter,et al.  The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapy. , 2000, Physics in medicine and biology.

[25]  D. Thwaites,et al.  Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisition. , 1999, Physics in medicine and biology.

[26]  J. Conway,et al.  An assessment of radiotherapy dosimeters based on CVD grown diamond , 2001 .

[27]  D I Thwaites,et al.  Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors. , 2011, Medical physics.

[28]  Carlo Cavedon,et al.  Total scatter factors of small beams: a multidetector and Monte Carlo study. , 2008, Medical physics.

[29]  Marco Marinelli,et al.  Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams. , 2012, Medical physics.

[30]  Z. Kuncic,et al.  An evaluation of calculation parameters in the EGSnrc/BEAMnrc Monte Carlo codes and their effect on surface dose calculation , 2012, Physics in medicine and biology.

[31]  L. Reinisch,et al.  Clinical investigations of a CVD diamond detector for radiotherapy dosimetry. , 2012, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[32]  F. Yin,et al.  Dosimetric characteristics of Novalis shaped beam surgery unit. , 2002, Medical physics.