Cramér–Rao lower bound of basis image noise in multiple-energy x-ray imaging

We present an analytical method to compute the basis image noise in the context of multi-energy x-ray imaging based on the Cramér-Rao lower bound (CRLB). The proposed formalism extends the original idea of Alvarez and Macovski (1976 Phys. Med. Biol. 21 733) to estimate the noise in the photo-effect and Compton-effect basis images in the case of dual-energy imaging. It includes an arbitrary number of independent, spectrally distinct attenuation measurements and also goes beyond the two-dimensional decomposition of the attenuation, including, e.g., a contrast agent as a third basis material. To illustrate our method, we consider three simple applications. The first application is to study the influence of the exact values for the energy thresholds on the basis image noise for a binned photon-counting system. The second application relates to the same detector system as the first and is an investigation of the dependence of the basis image noise on the energy resolution of the detector system. Finally, the third application provides an example for the case of an energy-integrating detector: the aim is to optimize the front-scintillator layer thickness of a dual-crystal detector for dual-energy imaging. The CRLB is used to minimize the noise of a photo-effect/Compton-effect basis material decomposition.

[1]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[2]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[3]  S. Riederer,et al.  Selective iodine imaging using K-edge energies in computerized x-ray tomography. , 1977, Medical physics.

[4]  R. Brooks,et al.  Split-detector computed tomography: a preliminary report. , 1978, Radiology.

[5]  P. Joseph,et al.  Noise considerations in dual energy CT scanning. , 1979, Medical physics.

[6]  G. Knoll Radiation detection and measurement , 1979 .

[7]  A. Macovski,et al.  Generalized image combinations in dual KVP digital radiography. , 1981, Medical physics.

[8]  S J Riederer,et al.  Dual-energy projection radiography: initial clinical experience. , 1981, AJR. American journal of roentgenology.

[9]  G T Barnes,et al.  Detector for dual-energy digital radiography. , 1985, Radiology.

[10]  M Ishida,et al.  Breast imaging: dual-energy projection radiography with digital radiography. , 1987, Radiology.

[11]  M Ishida,et al.  Breast imaging: dual-energy projection radiography with digital radiography. , 1987, Radiology.

[12]  J. H. Hubbell,et al.  XCOM: Photon cross sections on a personal computer , 1987 .

[13]  G T Barnes,et al.  Noise correlations in images acquired simultaneously with a dual-energy sandwich detector. , 1989, Medical physics.

[14]  G. Barnes,et al.  Semiempirical model for generating tungsten target x-ray spectra. , 1991, Medical physics.

[15]  D. Gauntt,et al.  X-ray tube potential, filtration, and detector considerations in dual-energy chest radiography. , 1994, Medical physics.

[16]  R E Alvarez Active energy selective image detector for dual-energy computed radiography. , 1996, Medical physics.

[17]  D. Mcgregor,et al.  Room-temperature compound semiconductor radiation detectors , 1997 .

[18]  G. Cowan Statistical data analysis , 1998 .

[19]  N J Pelc,et al.  Depth-segmented detector for x-ray absorptiometry. , 2000, Medical physics.

[20]  Steven Kay,et al.  Fundamentals Of Statistical Signal Processing , 2001 .

[21]  R. Alvarez,et al.  Comparison of dual energy detector system performance. , 2004, Medical physics.

[22]  R. Carmi,et al.  Material separation with dual-layer CT , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[23]  C. Herrmann,et al.  Counting and integrating readout for direct conversion X-ray imaging concept, realization and first prototype measurements , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[24]  E. Roessl,et al.  Optimal energy threshold arrangement in photon-counting spectral x-ray imaging , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[25]  K. Stierstorfer,et al.  First performance evaluation of a dual-source CT (DSCT) system , 2006, European Radiology.

[26]  E. Roessl,et al.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors , 2007, Physics in medicine and biology.

[27]  E. Roessl,et al.  On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging. , 2007, Medical physics.

[28]  J. Schlomka,et al.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography , 2008, Physics in medicine and biology.

[29]  Sensitivity of photon-counting K-edge imaging: Dependence on atomic number and object size , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.