Technical Note: Quantitative accuracy evaluation for spectral images from a detector-based spectral CT scanner using an iodine phantom.

PURPOSE The purpose of this study was to evaluate the quantitative accuracy of spectral images from a detector-based spectral CT scanner using a phantom with iodine-loaded inserts. METHODS A 40-cm long-body phantom with seven iodine inserts (2-20 mg/ml of iodine) was used in the study. The inserts could be placed at 5.5 or 10.5 cm from the phantom axis. The phantom was scanned five times for each insert configuration using 120 kVp tube voltage. A set of iodine, virtual noncontrast, effective atomic number, and virtual monoenergetic spectral CT images were generated and measurements were made for all the iodine rods. Measured values were compared with reference values calculated from the chemical composition information provided by the phantom manufacturer. Radiation dose from the spectral CT was compared to a conventional CT using a CTDI (32 cm) phantom. RESULTS Good agreement between measurements and reference values was achieved for all types of spectral images. The differences ranged from -0.46 to 0.1 mg/ml for iodine concentration, -9.95 to 6.41 HU for virtual noncontrast images, 0.12 to 0.35 for effective Z images, and -17.7 to 55.7 HU for virtual monoenergetic images. For a similar CTDIvol, image noise from the conventional CT was 10% lower than the spectral CT. CONCLUSIONS The detector-based spectral CT can achieve accurate spectral measurements on iodine concentration, virtual non-contrast images, effective atomic numbers, and virtual monoenergetic images.

[1]  A Macovski,et al.  Energy dependent reconstruction in X-ray computerized tomography. , 1976, Computers in biology and medicine.

[2]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[3]  W. Kalender,et al.  Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. , 1986, Medical physics.

[4]  G. Hounsfield Computerized transverse axial scanning (tomography): Part I. Description of system. 1973. , 1973, The British journal of radiology.

[5]  K. Stierstorfer,et al.  Density and atomic number measurements with spectral x-ray attenuation method , 2003 .

[6]  K. Stierstorfer,et al.  First performance evaluation of a dual-source CT (DSCT) system , 2006, European Radiology.

[7]  M. Reiser,et al.  Material differentiation by dual energy CT: initial experience , 2007, European Radiology.

[8]  William H. Judy,et al.  The Measurement, Reporting, and Management of Radiation Dose in CT , 2008 .

[9]  Guowei Zhang,et al.  A practical reconstruction method for dual energy computed tomography , 2008 .

[10]  C. McCollough,et al.  Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition. , 2009, Medical physics.

[11]  Xinhua Li,et al.  Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode. , 2011, Medical physics.

[12]  M. Macari,et al.  Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. , 2011, AJR. American journal of roentgenology.

[13]  Mitchell M Goodsitt,et al.  Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner. , 2011, Medical physics.

[14]  C. McCollough,et al.  Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications. , 2015, Radiology.

[15]  Accuracy of dual‐energy computed tomography for the quantification of iodine in a soft tissue‐mimicking phantom , 2015, Journal of applied clinical medical physics.

[16]  Julien Milles,et al.  Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT , 2017, European Radiology.