Development of an intraluminal intestinal photoplethysmography sensor

Intestinal ischemia is a serious medical condition and can lead to life threatening sepsis. Currently, there are no reliable techniques available for directly monitoring intestinal viability for prolonged periods of time, and intraoperatively, the majority of the surgeons still rely on subjective methods, such as visual inspection to assess viability of the intestine. The development of an intraluminal optical sensor for monitoring intestinal viability is being proposed. The sensor will continuously monitor changes in blood volume and oxygen saturation. The developed reflectance photoplethysmography/pulse oximetry sensor comprises of two emitters (red and infrared) and a photodiode. A photoplethysmography processing and data acquisition system was also utilized. The prototype sensor was evaluated in a pilot study in the buccal mucosa of 12 healthy volunteers, given the locations similarity to the intestinal mucosa and its easy accessibility. Good quality photoplethysmography signals with high signal-to-noise ratio were acquired from the buccal mucosa in all the volunteers. Preliminary blood oxygen saturation values from the intraluminal sensor were in broad agreement with the standard finger pulse oximeter probes.

[1]  Omar Faiz,et al.  Intraoperative assessment of colorectal anastomotic integrity: a systematic review , 2014, Surgical Endoscopy.

[2]  K Budidha,et al.  The human ear canal: investigation of its suitability for monitoring photoplethysmographs and arterial oxygen saturation , 2014, Physiological measurement.

[3]  Steven D. Mills,et al.  The use of indocyanine green fluorescence to assess anastomotic perfusion during robotic assisted laparoscopic rectal surgery , 2013, Surgical Endoscopy.

[4]  J. Carmichael,et al.  Management of Ischemic Colitis , 2012, Clinics in Colon and Rectal Surgery.

[5]  N. Hyman Managing anastomotic leaks from intestinal anastomoses. , 2009, The surgeon : journal of the Royal Colleges of Surgeons of Edinburgh and Ireland.

[6]  T. Kingham,et al.  Colonic anastomotic leak: risk factors, diagnosis, and treatment. , 2009, Journal of the American College of Surgeons.

[7]  John G. Webster,et al.  Design of Pulse Oximeters , 1997 .

[8]  R. Mason,et al.  Intraoperative assessment of colonic perfusion using scanning laser Doppler flowmetry during colonic resection. , 2000, Journal of the American College of Surgeons.

[9]  Eugene S. Lee,et al.  Intraoperative colon mucosal oxygen saturation during aortic surgery. , 2007, Journal of Surgical Research.

[10]  Michelle Hickey,et al.  Measurement of splanchnic photoplethysmographic signals using a new reflectance fiber optic sensor. , 2010, Journal of biomedical optics.

[11]  Panayiotis A Kyriacou,et al.  Direct Pulse Oximetry Within the Esophagus, on the Surface of Abdominal Viscera, and on Free Flaps , 2013, Anesthesia and analgesia.