Modeling and performance evaluation of a robotic treatment couch for tumor tracking

Abstract Tumor motion during radiation therapy increases the irradiation of healthy tissue. However, this problem may be mitigated by moving the patient via the treatment couch such that the tumor motion relative to the beam is minimized. The treatment couch poses limitations to the potential mitigation, thus the performance of the Protura (CIVCO) treatment couch was characterized and numerically modeled. The unknown parameters were identified using chirp signals and verified with one-dimensional tumor tracking. The Protura tracked chirp signals well up to 0.2 Hz in both longitudinal and vertical directions. If only the vertical or only the longitudinal direction was tracked, the Protura tracked well up to 0.3 Hz. However, there was unintentional yet substantial lateral motion in the former case. And during vertical motion, the extension caused rotation of the Protura around the lateral axis. The numerical model matched the Protura up to 0.3 Hz. Even though the Protura was designed for static positioning, it was able to reduce the tumor motion by 69% (median). The correlation coefficient between the tumor motion reductions of the Protura and the model was 0.99. Therefore, the model allows tumor-tracking results of the Protura to be predicted.

[1]  Daniel Paluszczyszyn,et al.  Couch-based motion compensation: modelling, simulation and real-time experiments , 2012, Physics in medicine and biology.

[2]  Sonja Dieterich,et al.  Tumor motion ranges due to respiration and respiratory motion characteristics , 2007 .

[3]  Bhaskar Dasgupta,et al.  Closed-Form Dynamic Equations of the General Stewart Platform through the Newton–Euler Approach , 1998 .

[4]  Roberto Zanasi,et al.  Time-optimal regulation of a chain of integrators with saturated input and internal variables: an application to trajectory planning , 2010 .

[5]  M. V. van Herk,et al.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[6]  J. Kuipers Quaternions and Rotation Sequences , 1998 .

[7]  Steve Webb,et al.  A novel respiratory motion compensation strategy combining gated beam delivery and mean target position concept --a compromise between small safety margins and long duty cycles. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  Yan Yu,et al.  Tumor Motion Prediction and Tracking in Adaptive Radiotherapy , 2010, 2010 IEEE International Conference on BioInformatics and BioEngineering.

[9]  Gregory C Sharp,et al.  Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy. , 2006, International journal of radiation oncology, biology, physics.

[10]  Gregor Ochsner,et al.  Development and evaluation of a prototype tracking system using the treatment couch. , 2014, Medical physics.

[11]  S. Sastryvedam GEOMETRIC ACCURACY OF A REAL-TIME TARGET TRACKING SYSTEM WITH DYNAMIC MULTILEAF COLLIMATOR TRACKING SYSTEM , 2006 .

[12]  Y Yu,et al.  A robotic approach to 4D real-time tumor tracking for radiotherapy , 2011, Physics in medicine and biology.

[13]  M. Hoogeman,et al.  Clinical accuracy of the respiratory tumor tracking system of the cyberknife: assessment by analysis of log files. , 2009, International journal of radiation oncology, biology, physics.

[14]  Klaus Schilling,et al.  Modeling a HexaPOD for Tumor Motion Compensation in Robot Assisted Radiotherapy , 2010, ISR/ROBOTIK.

[15]  Klaus Schilling,et al.  Control of a HexaPOD treatment couch for robot-assisted radiotherapy , 2012, Biomedizinische Technik. Biomedical engineering.

[16]  Thierry Gevaert,et al.  Geometric accuracy of a novel gimbals based radiation therapy tumor tracking system. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  Sarah Quirk,et al.  External respiratory motion analysis and statistics for patients and volunteers , 2013, Journal of applied clinical medical physics.

[18]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[19]  W. O'Dell,et al.  Modeling liver motion and deformation during the respiratory cycle using intensity-based nonrigid registration of gated MR images. , 2004, Medical physics.

[20]  Warren D D'Souza,et al.  An analysis of the treatment couch and control system dynamics for respiration-induced motion compensation. , 2006, Medical physics.