Feedforward correction of mirror misalignment fluctuations for the GEO 600 gravitational wave detector

The core instrument of the GEO 600 gravitational wave detector is a Michelson interferometer with folded arms. The five main optics that form this interferometer are suspended in vacuum by triple pendulums with quasi-monolithic lower stages of fused silica. After installation of these pendulums in early 2003, a larger than expected coupling of longitudinal ground motion to tilt misalignment of the suspended optics was observed. Because of this, the uncontrolled misalignment of the optics during average conditions was several µrad Hz−1/2 in the frequency band around the pendulum resonance frequencies (0.5–4 Hz). In addition, it was found that longitudinal control signals applied to the intermediate pendulum stages also resulted in excessive mirror tilt. The resulting misalignment exceeded the level tolerable for stable operation of GEO 600. In order to reduce the level of mirror tilt, a bipartite feedforward system was implemented. One part feeds signals derived from seismic measurements to piezo-electric crystals in the stacks supporting the suspensions, reducing the longitudinal motion of the uppermost suspension points. The other applies tilt correction signals, derived from longitudinal control signals, to the intermediate level of the suspensions. The seismic feedforward correction reduces the root-mean-squared tilt misalignment of each main optic between 0.1 and 5 Hz by about 10 dB, typically. The intermediate-mass feedforward correction reduces the differential tilt misalignment of the Michelson interferometer by about 10 dB between 0.1 and 0.8 Hz, typically.

[1]  M. M. Casey,et al.  Commissioning, characterization and operation of the dual-recycled GEO 600 , 2004 .

[2]  Daniel Enard,et al.  Status of VIRGO , 2004, SPIE Astronomical Telescopes + Instrumentation.

[3]  Martin M. Fejer,et al.  Mechanical quality factor measurements of monolithically suspended fused silica test masses of the GEO 600 gravitational wave detector , 2004 .

[4]  Karsten Danzmann,et al.  Damping and tuning of the fibre violin modes in monolithic silica suspensions , 2004 .

[5]  Karsten Danzmann,et al.  Thermal correction of the radii of curvature of mirrors for GEO 600 , 2004 .

[6]  M. M. Casey,et al.  Alignment control of GEO 600 , 2004 .

[7]  B Willke,et al.  Frequency-domain interferometer simulation with higher-order spatial modes , 2004 .

[8]  Joshua R. Smith,et al.  Mechanical loss associated with silicate bonding of fused silica , 2003 .

[9]  M. Fejer,et al.  The intrinsic mechanical loss factor of hydroxy-catalysis bonds for use in the mirror suspensions of gravitational wave detectors , 2003 .

[10]  R. Schilling,et al.  Frequency domain interferometer simulation with higher-order spatial modes , 2003, gr-qc/0309012.

[11]  R. Abbott,et al.  Feedforward reduction of the microseism disturbance in a long-baseline interferometric gravitational-wave detector , 2002 .

[12]  K. Arai Status of TAMA300 , 2002 .

[13]  V. Mitrofanov,et al.  Investigations of the dynamics and mechanical dissipation of a fused silica suspension , 2002 .

[14]  Daniel Sigg,et al.  Commissioning of the LIGO detectors , 2002 .

[15]  Hiroaki Yamamoto,et al.  Mirror suspension system for the TAMA SAS , 2002 .

[16]  Benno Willke,et al.  Dual recycling for GEO 600 , 2002 .

[17]  K. Arai,et al.  Stable operation of a 300-m laser interferometer with sufficient sensitivity to detect gravitational-wave events within our galaxy. , 2001, Physical review letters.

[18]  Bernard F. Schutz,et al.  The GEO 600 gravitational wave detector , 2002 .

[19]  Kenneth A. Strain,et al.  GEO 600 triple pendulum suspension system: Seismic isolation and control , 2000 .

[20]  M. Thattai,et al.  Increased thermal noise in nonuniform fiber suspensions , 1999 .