3-D Residual Eddy Current Field Characterisation: Applied to Diffusion Weighted Magnetic Resonance Imaging

Clinical use of the Stejskal-Tanner diffusion weighted images is hampered by the geometric distortions that result from the large residual 3-D eddy current field induced. In this work, we aimed to predict, using linear response theory, the residual 3-D eddy current field required for geometric distortion correction based on phantom eddy current field measurements. The predicted 3-D eddy current field induced by the diffusion-weighting gradients was able to reduce the root mean square error of the residual eddy current field to ~1 Hz. The model's performance was tested on diffusion weighted images of four normal volunteers, following distortion correction, the quality of the Stejskal-Tanner diffusion-weighted images was found to have comparable quality to image registration based corrections (FSL) at low b-values. Unlike registration techniques the correction was not hindered by low SNR at high b-values, and results in improved image quality relative to FSL. Characterization of the 3-D eddy current field with linear response theory enables the prediction of the 3-D eddy current field required to correct eddy current induced geometric distortions for a wide range of clinical and high b-value protocols.

[1]  M A Horsfield,et al.  Mapping eddy current induced fields for the correction of diffusion-weighted echo planar images. , 1999, Magnetic resonance imaging.

[2]  Bin Chen,et al.  Correction for direction-dependent distortions in diffusion tensor imaging using matched magnetic field maps , 2006, NeuroImage.

[3]  J. C. Gore,et al.  Motion and distortion correction in diffusion-weighted MRI of the breast at 3 T , 2009 .

[4]  Mark E Bastin,et al.  A quantitative comparison of two methods to correct eddy current-induced distortions in DT-MRI. , 2007, Magnetic resonance imaging.

[5]  Michael Deppe,et al.  Correcting eddy current and motion effects by affine whole‐brain registrations: Evaluation of three‐dimensional distortions and comparison with slicewise correction , 2010, Magnetic resonance in medicine.

[6]  Alan Connelly,et al.  MRtrix: Diffusion tractography in crossing fiber regions , 2012, Int. J. Imaging Syst. Technol..

[7]  Trong-Kha Truong,et al.  Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging , 2015, BioMed research international.

[8]  P. Jezzard,et al.  Correction for geometric distortion in echo planar images from B0 field variations , 1995, Magnetic resonance in medicine.

[9]  Markus Nilsson,et al.  A Novel Robust Algorithm to Correct for Eddy Current Distortions in High b-value Diffusion MRI , 2010 .

[10]  Jon Bardin,et al.  Neuroscience: Making connections , 2012, Nature.

[11]  Bin Chen,et al.  Integrated SENSE DTI with correction of susceptibility- and eddy current-induced geometric distortions , 2008, NeuroImage.

[12]  Dongrong Xu,et al.  Correction of eddy‐current distortions in diffusion tensor images using the known directions and strengths of diffusion gradients , 2006, Journal of magnetic resonance imaging : JMRI.

[13]  Wilson Fong Handbook of MRI Pulse Sequences , 2005 .

[14]  Paul Mansfield,et al.  LETTER TO THE EDITOR: Double active magnetic screening of coils in NMR , 1986 .

[15]  R. Ordidge,et al.  Correction of motional artifacts in diffusion-weighted MR images using navigator echoes. , 1994, Magnetic resonance imaging.

[16]  Derek K. Jones,et al.  Motion correction and registration of high b‐value diffusion weighted images , 2012, Magnetic resonance in medicine.

[17]  C. Boesch,et al.  Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systems , 1991, Magnetic resonance in medicine.

[18]  Thomas Benner,et al.  Diffusion imaging with prospective motion correction and reacquisition , 2011, Magnetic resonance in medicine.

[19]  N G Papadakis,et al.  Gradient preemphasis calibration in diffusion‐weighted echo‐planar imaging , 2000, Magnetic resonance in medicine.

[20]  N. Dombrowski,et al.  Linear Response Theory for Random Schr , 2011 .

[21]  Jason Berwick,et al.  k‐space correction of eddy‐current‐induced distortions in diffusion‐weighted echo‐planar imaging , 2005, Magnetic resonance in medicine.

[22]  Jurek A Nordmeyer-Massner,et al.  NMR probes for measuring magnetic fields and field dynamics in MR systems , 2008, Magnetic resonance in medicine.

[23]  Andrew S. Glassner,et al.  Graphics Gems , 1990 .

[24]  R. Gruetter,et al.  A 3 D eddy current model for the prediction of geometric image distortions in Stejkal-Tanner diffusion weighted EPI , 2010 .

[25]  J. E. Tanner,et al.  Spin diffusion measurements : spin echoes in the presence of a time-dependent field gradient , 1965 .

[26]  Mark W. Woolrich,et al.  Advances in functional and structural MR image analysis and implementation as FSL , 2004, NeuroImage.

[27]  D. Tuch Q‐ball imaging , 2004, Magnetic resonance in medicine.

[28]  D. Le Bihan,et al.  Artifacts and pitfalls in diffusion MRI , 2006, Journal of magnetic resonance imaging : JMRI.

[29]  Stefan Skare,et al.  A Model-Based Method for Retrospective Correction of Geometric Distortions in Diffusion-Weighted EPI , 2002, NeuroImage.

[30]  Thomas E Yankeelov,et al.  Motion correction in diffusion‐weighted MRI of the breast at 3T , 2011, Journal of magnetic resonance imaging : JMRI.

[31]  R Gruetter,et al.  Localized eddy current compensation using quantitative field mapping. , 1998, Journal of magnetic resonance.

[32]  C. Pierpaoli,et al.  Characterization of and correction for eddy current artifacts in echo planar diffusion imaging , 1998, Magnetic resonance in medicine.

[33]  Tobias Kober,et al.  Prospective and retrospective motion correction in diffusion magnetic resonance imaging of the human brain , 2012, NeuroImage.

[34]  David Atkinson,et al.  High-resolution diffusion tensor imaging of the human kidneys using a free-breathing, multi-slice, targeted field of view approach , 2014, NMR in biomedicine.

[35]  Anthonie Hendrik Bergman,et al.  Optimization of eddy-current compensation☆ , 1990 .

[36]  Yuji Shen,et al.  Correction of high‐order eddy current induced geometric distortion in diffusion‐weighted echo‐planar images , 2004, Magnetic resonance in medicine.

[37]  Alan Connelly,et al.  Robust determination of the fibre orientation distribution in diffusion MRI: Non-negativity constrained super-resolved spherical deconvolution , 2007, NeuroImage.

[38]  J. Finsterbusch Eddy‐current compensated diffusion weighting with a single refocusing RF pulse , 2009, Magnetic resonance in medicine.

[39]  Haruyasu Yamada,et al.  Efficacy of Distortion Correction on Diffusion Imaging: Comparison of FSL Eddy and Eddy_Correct Using 30 and 60 Directions Diffusion Encoding , 2014, PloS one.

[40]  P. Jehenson,et al.  Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systems , 1990 .

[41]  J. Mazziotta,et al.  Rapid Automated Algorithm for Aligning and Reslicing PET Images , 1992, Journal of computer assisted tomography.

[42]  Louis A. Romero,et al.  Robust two-dimensional weighted and unweighted phase unwrapping that uses fast transforms and iterative methods , 1994 .

[43]  K. Pruessmann,et al.  Spatiotemporal magnetic field monitoring for MR , 2008, Magnetic resonance in medicine.

[44]  James R. Moore,et al.  Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficient , 1996, Magnetic resonance in medicine.

[45]  Mark Jenkinson,et al.  Reducing distortions in diffusion‐weighted echo planar imaging with a dual‐echo blip‐reversed sequence , 2010, Magnetic resonance in medicine.

[46]  V. Wedeen,et al.  Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echo , 2003, Magnetic resonance in medicine.

[47]  Derek K. Jones,et al.  Diffusion‐tensor MRI: theory, experimental design and data analysis – a technical review , 2002 .