Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure

Object Self-measurement of blood pressure (BP) might offer some advantages in diagnosis and therapeutic evaluation and in patient management of hypertension. Recently, wrist-cuff devices for self-measurement of BP have gained more than one-third of the world market share. In the present study, we validated wrist-cuff devices and compared the results between wrist- and arm-cuff devices. The factors affecting the accuracy of wrist-cuff devices were also studied. Method The research group to assess the validity of automated blood pressure measuring device consisted of 13 institutes in Japan, which validated two wrist-cuff devices (WC-1 and WC-2) and two arm-cuff devices (AC-1 and AC-2). They used a crossover method, where the comparison was done between auscultation, by two observers by means of a double stethoscope on one arm and the device on the opposite arm or wrist. Results There was good inter-observer agreement for the auscultation method in each institute (systolic blood pressure (SBP), −0.1 ± 2.8 mmHg; diastolic blood pressure (DBP), −0.1 ± 2.6 mmHg, n = 498). The mean difference between auscultation and the device was minimal both in arm-cuff devices (mean difference for AC-1, 2.2/1.9 mmHg, n = 97 and for AC-2, 5.1/2.9 mmHg, n = 136, SBP/DBP) and wrist-cuff devices (mean difference for WC-1, −2.1/1.2 mmHg, n = 173 mmHg and for WC-2, −2.3/−5.6 mmHg, n = 92). The standard deviation of the difference (SDD) in wrist-cuff devices, however (SDD for WC-1, 9.7/7.3 mmHg and for WC-2, 10.2/8.6 mmHg), was larger than that of the arm-cuff devices (SDD for AC-1, 5.6/6.6 mmHg and for AC-2, 6.3/5.1 mmHg). Grading of AC-1 and AC-2 based on criteria of British Hypertension Society was A/A and B/A, respectively, while that of WC-1 and WC-2 was C/B and D/B, respectively. Using the same validation protocol, the results of validation for one device were divergent in each institute. In wrist-cuff devices, the BP value obtained in palmar flexion was significantly higher and that obtained in palmar dorsiflexion was significantly lower than that in palmar extension. In some cases, finger plethysmogram did not disappear during maximum inflation of the wrist-cuff (≅ 250 mmHg), even in palmar extension and especially in palmar flexion, suggesting incomplete obstruction of radial and/or ulnar arteries during inflation. Conclusion The results suggest that wrist-cuff devices in the present form are inadequate for self-measurement of blood pressure and, thus, are inadequate for general use or clinical and practical use. However, there is much possibility in wrist-cuff device and the accuracy and reliability of wrist-cuff device are warranted by an improvement of technology.

[1]  A. Coats,et al.  Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. , 1994, Journal of human hypertension.

[2]  W. Vetter,et al.  Self-measurement of blood pressure improves the accuracy and reduces the number of subjects in clinical trials. , 1991, Journal of hypertension. Supplement : official journal of the International Society of Hypertension.

[3]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[4]  J. Ménard,et al.  Crossover design to test antihypertensive drugs with self-recorded blood pressure. , 1988, Hypertension.

[5]  A. Hedayat,et al.  Statistical Methods in Assessing Agreement , 2002 .

[6]  M. Thümmler,et al.  Ein neues oszillometrisches Blutdruckmeßgerät mit Handgelenkmanschette - Erste klinische Vergleichsuntersuchung , 1993 .

[7]  Dubroca,et al.  Superiority of home blood pressure measurements over office measurements for testing antihypertensive drugs. , 1998, Blood pressure monitoring.

[8]  L A Geddes,et al.  The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. 1. , 1969, Cardiovascular Research Center bulletin.

[9]  Bruce Neal,et al.  1999 World Health Organization-International Society of Hypertension Guidelines for the Management of Hypertension. Guidelines Subcommittee. , 1999, Journal of hypertension.

[10]  F. Mee,et al.  Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. , 1996, Blood pressure monitoring.

[11]  Y. Imai,et al.  Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods , 1989, Journal of hypertension.

[12]  Eckert,et al.  Validation of the OMRON R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation. , 1997, Blood pressure monitoring.

[13]  F. Mee,et al.  Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure. , 1990, Journal of hypertension.

[14]  Shigeru Hisamichi,et al.  Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population‐based observation in Ohasama, Japan , 1998, Journal of hypertension.

[15]  A. Zanchetti,et al.  Guidelines for the use of self-blood pressure monitoring: a summary report of the First International Consensus Conference. Groupe Evaluation & Measure of the French Society of Hypertension. , 2000, Journal of hypertension.

[16]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[17]  E. O’Brien,et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. , 1990, Journal of hypertension.

[18]  Self-measurement of blood pressure. , 1988, Lancet.

[19]  E van der Does,et al.  Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. , 1995, Journal of human hypertension.

[20]  T. Ohkubo,et al.  Proposal of reference values for home blood pressure measurement: prognostic criteria based on a prospective observation of the general population in Ohasama, Japan. , 1997, American journal of hypertension.