Design and performance evaluation of a high resolution IRI-microPET preclinical scanner

Abstract PET for small animal, IRI-microPET, was designed and built at the NSTRI. The scanner is made of four detectors positioned on a rotating gantry at a distance 50 mm from the center. Each detector consists of a 10×10 crystal matrix of 2×2×10 mm 3 directly coupled to a PS-PMT. A position encoding circuit for specific PS-PMT has been designed, built and tested with a PD-MFS-2MS/s-8/14 data acquisition board. After implementing reconstruction algorithms (FBP, MLEM and SART) on sinograms, images quality and system performance were evaluated by energy resolution, timing resolution, spatial resolution, scatter fraction, sensitivity, RMS contrast and SNR parameters. The energy spectra were obtained for the crystals with an energy window of 300–700 keV. The energy resolution in 511 keV averaged over all modules, detectors, and crystals, was 23.5%. A timing resolution of 2.4 ns FWHM obtained by coincidence timing spectrum was measured with crystal LYSO. The radial and tangential resolutions for 18 F (1.15-mm inner diameter) at the center of the field of view were 1.81 mm and 1.90 mm, respectively. At a radial offset of 5 mm, the FWHM values were 1.96 and 2.06 mm. The system scatter fraction was 7.1% for the mouse phantom. The sensitivity was measured for different energy windows, leading to a sensitivity of 1.74% at the center of FOV. Also, images quality was evaluated by RMS contrast and SNR factors, and the results show that the reconstructed images by MLEM algorithm have the best RMS contrast, and SNR. The IRI-microPET presents high image resolution, low scatter fraction values and improved SNR for animal studies.

[1]  A. Del Guerra,et al.  YAP-PET: first results of a small animal positron emission tomograph based on YAP:Ce finger crystals , 1998 .

[2]  James F. Young,et al.  MicroPET: a high resolution PET scanner for imaging small animals , 1996, IEEE Nuclear Science Symposium Conference Record.

[3]  Jogeshwar Mukherjee,et al.  Performance evaluation of an Inveon PET preclinical scanner , 2008, Physics in medicine and biology.

[4]  R. N. Goble,et al.  Performance evaluation of the microPET P4: a PET system dedicated to animal imaging. , 2001, Physics in medicine and biology.

[5]  H. Zaidi,et al.  Advances in PET Image Reconstruction. , 2007, PET clinics.

[6]  Jeffrey T. Yap,et al.  Performance evaluation of the microPET-Focus - F120 , 2004 .

[7]  Ming Jiang,et al.  Convergence of the simultaneous algebraic reconstruction technique (SART) , 2003, Conference Record of Thirty-Fifth Asilomar Conference on Signals, Systems and Computers (Cat.No.01CH37256).

[8]  R. Malaguti,et al.  Preliminary test results on the new electronic readout of the YAP(S)PET small animal scanner , 2001 .

[9]  Gabor T. Herman,et al.  Fundamentals of Computerized Tomography: Image Reconstruction from Projections , 2009, Advances in Pattern Recognition.

[10]  Magdalena Rafecas,et al.  A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals , 2001, European Journal of Nuclear Medicine.

[11]  W. V. Mayneord PHYSICS IN MEDICINE , 1945 .

[12]  Peter Bruyndonckx,et al.  Monolithic scintillator PET detectors with intrinsic depth-of-interaction correction , 2009, Physics in medicine and biology.

[13]  Manuel Desco,et al.  NEMA NU 4-2008 Performance Measurements of Two Commercial Small-Animal PET Scanners: ClearPET and rPET-1 , 2011, IEEE Transactions on Nuclear Science.

[14]  S. Siegel,et al.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[15]  R. N. Goble,et al.  Performance evaluation of the microPET R4 PET scanner for rodents , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[16]  J. S. Lee,et al.  Performance Measurement of the microPET Focus 120 Scanner , 2007, Journal of Nuclear Medicine.

[17]  Simone Weber,et al.  Small animal PET: aspects of performance assessment , 2004, European Journal of Nuclear Medicine and Molecular Imaging.

[18]  Takuya Hayashi,et al.  Use of a compact pixellated gamma camera for small animal pinhole SPECT imaging , 2006, Annals of nuclear medicine.

[19]  C Damiani,et al.  Use of a fast EM algorithm for 3D image reconstruction with the YAP-PET tomograph. , 2002, Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society.

[20]  Frederic H Fahey,et al.  Data acquisition in PET imaging. , 2002, Journal of nuclear medicine technology.

[21]  S. Cherry,et al.  MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. , 2003, Physics in medicine and biology.

[22]  Yuan-Chuan Tai,et al.  Evaluation of Scatter Fraction and Noise Equivalent Count Rate Performance of the microPET F-220 Scanner Integrated with the Micro Insert Device , 2010 .

[23]  Oliver Thews,et al.  Preliminary assessment of the imaging capability of the YAP-(S)PET small animal scanner in neuroscience , 2006 .

[24]  E. Kim,et al.  Molecular imaging and gene therapy. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[25]  Habib Zaidi,et al.  Current Trends in Preclinical PET System Design. , 2007, PET clinics.

[26]  Jurgen Seidel,et al.  Resolution uniformity and sensitivity of the NIH ATLAS small animal PET scanner: comparison to simulated LSO scanners without depth-of-interaction capability , 2001 .

[27]  R. Leahy,et al.  Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging , 2004, Physics in medicine and biology.

[28]  Jae-Seung Kim,et al.  Impact of system design parameters on image figures of merit for a mouse PET scanner , 2004, IEEE Transactions on Nuclear Science.

[29]  A Rodríguez-Ruano,et al.  Design and performance evaluation of a coplanar multimodality scanner for rodent imaging. , 2009, Physics in medicine and biology.

[30]  Arion F. Chatziioannou,et al.  Molecular imaging of small animals with dedicated PET tomographs , 2001, European Journal of Nuclear Medicine and Molecular Imaging.

[31]  Martin A Lodge,et al.  Simultaneous measurement of noise and spatial resolution in PET phantom images. , 2010, Physics in medicine and biology.

[32]  Carlos Lacasta,et al.  Performance evaluation of a very high resolution small animal PET imager using silicon scatter detectors , 2007, Physics in medicine and biology.

[33]  Jing-Huei Lee,et al.  In vivo comparative imaging of dopamine D2 knockout and wild-type mice with (11)C-raclopride and microPET. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[34]  R E Carson,et al.  Noise reduction in oncology FDG PET images by iterative reconstruction: a quantitative assessment. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.