Photon arrival timing with sub-camera exposure time resolution in wide-field time-resolved photon counting imaging.

We demonstrate that an ultra-fast CMOS camera combined with a photon counting image intensifier can be used to determine photon arrival times well below the exposure time of the camera. We can obtain a time resolution down to around 1% of the exposure time, i.e. of the order of 40 ns with microsecond exposure times. This is achieved by exploiting the invariant phosphor decay of the image intensifier's phosphor screen: Developing a suitable mathematical framework, we show that the relative intensities of the phosphor decay in successive frames following the photon detection uniquely determine the photon arrival time. This approach opens a way to measuring fast luminescence decays in parallel in many pixels. Possible applications include oxygen and ion concentration imaging using probes with luminescence lifetimes in the range of 100 ns to microseconds.

[1]  Klaus Suhling,et al.  Optimisation of centroiding algorithms for photon event counting imaging , 1999 .

[2]  John B. Hutchings,et al.  Photon Event Centroiding with UV Photon‐counting Detectors , 2007 .

[3]  J. Siegel,et al.  Wide-Field Time-Resolved Fluorescence Anisotropy Imaging (TR-FAIM) , 2003 .

[4]  S. Weiss,et al.  High Speed Multichannel Charge Sensitive Data Acquisition System With Self-Triggered Event Timing , 2009, IEEE Transactions on Nuclear Science.

[5]  R. Yasukuni,et al.  Scanning‐less wide‐field single‐photon counting device for fluorescence intensity, lifetime and time‐resolved anisotropy imaging microscopy , 2008, Journal of microscopy.

[6]  D. Delpy,et al.  A 32-channel time-resolved instrument for medical optical tomography , 2000 .

[7]  K. Kemnitz,et al.  Wide-field photon counting fluorescence lifetime imaging microscopy: application to photosynthesizing systems , 2009, Photosynthesis Research.

[8]  Barry J. Kent,et al.  Uses of microchannel plate intensified detectors for imaging applications in the X-ray, EUV and visible wavelength regions , 1997 .

[9]  Yuji Yamamoto,et al.  New luminescent europium(III) chelates for DNA labeling. , 2006, Inorganic chemistry.

[10]  S. Ameer-Beg,et al.  Fluorescence lifetime and polarization-resolved imaging in cell biology. , 2009, Current opinion in biotechnology.

[11]  Stanley W Botchway,et al.  Time-resolved and two-photon emission imaging microscopy of live cells with inert platinum complexes , 2008, Proceedings of the National Academy of Sciences.

[12]  Edoardo Charbon,et al.  Real-time fluorescence lifetime imaging system with a 32 x 32 0.13microm CMOS low dark-count single-photon avalanche diode array. , 2010, Optics express.

[13]  Franco Zappa,et al.  Monolithic array of 32 SPAD pixels for single-photon imaging at high frame rates , 2009 .

[14]  A Zelić,et al.  Analytical approach to the recovery of short fluorescence lifetimes from fluorescence decay curves. , 1995, Biophysical journal.

[15]  M. Yaffe,et al.  The effect of phosphor persistence on image quality in digital x-ray scanning systems. , 1998, Medical physics.

[16]  T. Ng,et al.  The CS Award for chemical analysis and instrumentation , 1980 .

[17]  Martinus H V Werts,et al.  Making sense of Lanthanide Luminescence , 2005, Science progress.

[18]  J. Bünzli Lanthanide luminescence for biomedical analyses and imaging. , 2010, Chemical reviews.

[19]  B Vojnovic,et al.  Measurement of oxygen tension in tumours by time-resolved fluorescence. , 1996, The British journal of cancer. Supplement.

[20]  C G Morgan,et al.  Direct modulation of the effective sensitivity of a CCD detector: a new approach to time‐resolved fluorescence imaging , 2002, Journal of microscopy.

[21]  David Lloyd,et al.  Probing intracellular oxygen by quenched phosphorescence lifetimes of nanoparticles containing polyacrylamide-embedded [Ru(dpp(SO_3Na)_2)_3]Cl_2 , 2010, Photochemical & photobiological sciences : Official journal of the European Photochemistry Association and the European Society for Photobiology.

[22]  Masahiro Ito,et al.  Sodium Dodecyl Sulfate Polyacrylamide Slab Gel Electrophoresis and Hydroxyethyl Cellurose Gel Capillary Electrophoresis of Luminescent Lanthanide Chelate-labeled Proteins with Time-Resolved Detection , 2009, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[23]  Marion Peter,et al.  Imaging molecular interactions by multiphoton FLIM , 2004, Biology of the cell.

[24]  J Alexander Liddle,et al.  Fast, bias-free algorithm for tracking single particles with variable size and shape. , 2008, Optics express.

[25]  Andrew M. Wallace,et al.  Multiple wavelength time-of-flight sensor based on time-correlated single-photon counting , 2004 .

[26]  Charles L. Joseph,et al.  UV image sensors and associated technologies , 1995 .

[27]  Peter Kohl,et al.  Temporal Pixel Multiplexing for simultaneous high-speed high-resolution imaging , 2010, Nature Methods.

[28]  Paul R Selvin,et al.  Principles and biophysical applications of lanthanide-based probes. , 2002, Annual review of biophysics and biomolecular structure.

[29]  P. Schwille,et al.  Fluctuations as a source of information in fluorescence microscopy , 2009, Journal of The Royal Society Interface.

[30]  H. Offerhaus,et al.  Application of a time-resolved event counting technique in velocity map imaging , 2002 .

[31]  S. Weiss,et al.  Single-quantum dot imaging with a photon counting camera. , 2009, Current pharmaceutical biotechnology.

[32]  G. W. Fraser,et al.  Efficiency enhancements for MCP-based beta autoradiography imaging , 2002 .

[33]  I. Hemmilä,et al.  Progress in Lanthanides as Luminescent Probes , 2005, Journal of Fluorescence.

[34]  Klaus Suhling,et al.  Minimization of fixed pattern noise in photon event counting imaging , 2002 .

[35]  David J. S. Birch,et al.  REVIEW ARTICLE: Single-photon timing detectors for fluorescence lifetime spectroscopy , 1996 .

[36]  P. French,et al.  Time-resolved fluorescence microscopy , 2005 .

[37]  N. Pailer,et al.  Faint object camera: European contribution to the Hubble Space Telescope☆ , 1992 .

[38]  T. Ng,et al.  Imaging proteins in vivo using fluorescence lifetime microscopy. , 2007, Molecular bioSystems.

[39]  T M Jovin,et al.  Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging. , 1991, Biophysical journal.

[40]  M. Gersbach,et al.  A 128 $\times$ 128 Single-Photon Image Sensor With Column-Level 10-Bit Time-to-Digital Converter Array , 2008, IEEE Journal of Solid-State Circuits.

[41]  O. Jagutzki,et al.  Multiple hit read-out of a microchannel plate detector with a three-layer delay-line anode , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[42]  Neveen A. Hosny,et al.  Extracellular oxygen concentration mapping with a confocal multiphoton laser scanning microscope and TCSPC card , 2010, BiOS.

[43]  M. Neil,et al.  Real-time time-domain fluorescence lifetime imaging including single-shot acquisition with a segmented optical image intensifier , 2004 .

[44]  P. Selvin,et al.  Temporally and spectrally resolved imaging microscopy of lanthanide chelates. , 1998, Biophysical journal.

[45]  S. Weiss,et al.  Detectors for single-molecule fluorescence imaging and spectroscopy , 2007, Journal of modern optics.

[46]  N. A. Sharp Millisecond time resolution with the Kitt Peak Photon-Counting Array , 1992 .

[47]  Keith O. Mason,et al.  The Swift Ultra-Violet/Optical Telescope , 2002 .

[48]  P. Selvin,et al.  Progress in Lanthanides as Luminescent Probes , 2005 .

[49]  Klaus Suhling,et al.  A position-sensitive photon event counting detector applied to fluorescence imaging of dyes in sol-gel matrices , 2001 .