Advanced two-dimensional thermal neutron detectors for scattering studies

Advances in neutron scattering studies will be given a large boost with the advent of new spallation and reactor sources at present under consideration or construction. An important element for future experiments is a commensurate improvement in neutron detection techniques. At Brookhaven, a development program is under way for greatly increasing the angular coverage, rate capability and resolution of detectors for scattering studies. For example, a curved detector with angular coverage of 120° by 15° has recently been developed for protein crystallography at a spallation source. Based on neutron detection using 3He, the detector has the following major, new attributes: eight identical proportional wire segments operating in parallel, a single gas volume with seamless readout at segment boundaries, parallax errors eliminated in the horizontal plane by the detector's appropriate radius of curvature, high-throughput front-end electronics, position decoding based on high performance digital signal processing. The detector has a global rate capability greater than 1 million per second, position resolution less than 1.5 mm FWHM, timing resolution about 1 μs, efficiency of 50% and 90% at 1Å and 4 Å respectively, and an active area 1.5 m x 20 cm.