Evaluating performance of a pixel array semiconductor SPECT system for small animal imaging

ObjectivesSmall animal imaging has recently been focused on basic nuclear medicine. We have designed and built a small animal SPECT imaging system using a semiconductor camera and a newly designed collimator. We assess the performance of this system for small object imaging.MethodsWe employed an MGC1500 (Acrorad Co.) camera including a CdTe semiconductor. The pixel size was 1.4 mm/pixel. We designed and produced a parallel-hole collimator with 20-mm hole length. Our SPECT system consisted of a semiconductor camera with the subject holder set on an electric rotating stage controlled by a computer. We compared this system with a conventional small animal SPECT system comprising a SPECT-2000H scanner with four Anger type cameras and pinhole collimators. The count rate linearity for estimation of the scatter was evaluated for a pie-chart phantom containing different concentrations of99mTc. We measured the FWHM of the99mTc SPECT line source along with scatter. The system volume sensitivity was examined using a flood source phantom which was 35 mm long with a 32-mm inside diameter. Additionally, anin vivo myocardial perfusion SPECT study was performed with a rat.ResultsWith regards to energy resolution, the semiconductor camera (5.6%) was superior to the conventional Anger type camera (9.8%). In the count rate linearity evaluation, the regression lines of the SPECT values werey = 0.0 9x+ 0.031 (r2 = 0.999) for our system andy = 0.018* + 0.060 (r2 = 0.997) for the conventional system. Thus, the scatter count using the semiconductor camera was less than that using the conventional camera. FWHMs of our system and the conventional system were 2.9 ± 0.1 and 2.0 ± 0.1 mm, respectively. Moreover, the system volume sensitivity of our system [0.51 kcps/(MBq/ ml)/cm] was superior to that of the conventional system [0.44 kcps/(MBq/m/)/cm]. Our system provided clear images of the rat myocardium, sufficient for practical use in small animal imaging.ConclusionsOur SPECT system, utilizing a semiconductor camera, permits high quantitative analysis by virtue of its low scatter radiation and high sensitivity. Therefore, this system may contribute to molecular imaging of small animals and basic medical research. Key words: semiconductor detectors, small animal imaging, single photon emission computed tomography

[1]  I. Blevis,et al.  Evaluation of a small cadmium zinc telluride detector for scintimammography. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[2]  R. Moyer A low-energy multihole converging collimator compared with a pinhole collimator. , 1974, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[3]  K S Nikita,et al.  A 3D high-resolution gamma camera for radiopharmaceutical studies with small animals. , 2003, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[4]  K Doi,et al.  Investigation of basic imaging properties in digital radiography. I. Modulation transfer function. , 1984, Medical physics.

[5]  G A Clarke,et al.  Physical performance evaluation of the Toshiba GCA-9300A triple-headed system. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[6]  Y. Yonekura,et al.  Ultra-high resolution SPECT system using four pinhole collimators for small animal studies. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[7]  J. Ballinger,et al.  Indexing glomerular filtration rate to suit children. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  A. Kojima,et al.  Effect of spatial resolution on SPECT quantification values. , 1989, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  Michael W Dae,et al.  Pinhole single-photon emission computed tomography for myocardial perfusion imaging of mice. , 2003, Journal of the American College of Cardiology.

[10]  T. Yamamoto,et al.  Implementation and quantitative evaluation of analytical methods for attenuation correction in SPECT: a phantom study. , 1999, Physics in medicine and biology.

[11]  Paul D Acton,et al.  Small animal imaging with high resolution single photon emission tomography. , 2003, Nuclear medicine and biology.

[12]  M L Giger,et al.  Investigation of basic imaging properties in digital radiography. 6. MTFs of II-TV digital imaging systems. , 1985, Medical physics.

[13]  E. Hoffman,et al.  Quantitation in Positron Emission Computed Tomography: 1. Effect of Object Size , 1979, Journal of computer assisted tomography.

[14]  Tomio Inoue,et al.  Performance evaluation of a hand-held, semiconductor (CdZnTe)-based gamma camera , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[15]  Michael W Dae,et al.  Performance evaluation of a pinhole SPECT system for myocardial perfusion imaging of mice. , 2002, Medical physics.

[16]  B. Hasegawa,et al.  A bound on the energy resolution required for quantitative SPECT. , 1996, Medical physics.

[17]  G. Entine,et al.  Time-dependent polarization of CdTe gamma-ray detectors , 1974 .

[18]  H. Anger SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS. , 1964, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[20]  F. Fazio,et al.  Single-photon emission tomographic quantification in spherical objects: effects of object size and background , 1996, European Journal of Nuclear Medicine.

[21]  N. Motomura,et al.  The CdTe detector module and its imaging performance , 2001, Annals of nuclear medicine.

[22]  A. Todd-Pokropek,et al.  Solid state detectors in nuclear medicine. , 2002, The quarterly journal of nuclear medicine : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology.

[23]  H. Arai,et al.  Validation of left ventricular function from gated single photon computed emission tomography by using a scintillator-photodiode camera: a dynamic myocardial phantom study , 2002, Nuclear medicine communications.

[24]  J. Booij,et al.  Evaluation of high-resolution pinhole SPECT using a small rotating animal. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[25]  Einar Nygard,et al.  A prototype small CdTe gamma camera for radioguided surgery and other imaging applications , 2003, European Journal of Nuclear Medicine and Molecular Imaging.