Grayscale calibration and quality assurance of diagnostic monitors in a PACS system

Purpose.The aim of this study was to calibrate monitors used in soft-copy review of diagnostic images in a pictures archiving and communication system (PACS) and to assess critical quality assurance (QA) parameters through appropriate checks.Materials and methods.Barco [cathode ray tube (CRT) and liquid crystal display (LCD)] and EIZO (LCD) monitors were evaluated. Calibration and QA controls were carried out during acceptance tests on the systems and every 6 months according to the Task Group 18 (TG18) report by the American Association of Physicists in Medicine (AAPM). The parameters in question include: maximum luminance, contrast ratio, luminance response, spatial resolution and angular response. A subjective evaluation of image quality was also conducted by a number of radiologists.Results.Barco medical monitors’ results were well within tolerances, with significant parameters persisting over time. EIZO nonmedical monitors showed rapid performance deterioration below the minimum requirements. Calibration had to be repeated only in a few cases. Radiologists’ evaluations showed that monitor quality is equal to or even better than that of conventional films.Conclusions.Medical monitors turned out to be fully adequate to the task. Periodic QA tests are, however, absolutely necessary to ensure constant quality levels.

[1]  E Krupinski,et al.  Influence of film and monitor display luminance on observer performance and visual search. , 1999, Academic radiology.

[2]  Joel E. Gray,et al.  Acceptance and Use of the SMPTE Medical Diagnostic Imaging Test Pattern for Television Monitors and Hard-Copy Recording Cameras , 1990 .

[3]  Emily Seto,et al.  Image Quality Assurance of Soft Copy Display Systems , 2005, Journal of Digital Imaging.

[4]  F. Verdun,et al.  A strategy to qualify the performance of radiographic monitors. , 2005, Radiation protection dosimetry.

[5]  Ehsan Samei,et al.  Assessment of flat panel LCD primary class display performance based on AAPM TG 18 acceptance protocol. , 2004, Medical physics.

[6]  Ehsan Samei,et al.  Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. , 2005, Medical physics.

[7]  Jin Mo Goo,et al.  Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs. , 2004, Radiology.

[8]  E A Krupinski,et al.  The influence of a perceptually linearized display on observer performance and visual search. , 2000, Academic radiology.

[9]  Peter G. J. Barten,et al.  Physical model for the contrast sensitivity of the human eye , 1992, Electronic Imaging.

[10]  Peter M. Steven,et al.  Practical aspects of grayscale calibration of display systems , 2001, SPIE Medical Imaging.

[11]  P F Judy,et al.  Evaluation of video gray-scale display. , 1992, Medical physics.