A SiPM-Readout ASIC for SPECT Applications

In this paper, we present ANGUS, an application specific integrated circuit (ASIC) for silicon photomultiplier (SiPM) readout designed for low-energy gamma imaging (140 keV), specifically targeting clinical single photon emission computed tomography (SPECT) applications based on Anger-type detectors. The chip is composed of 36 channels, each featuring a current conveyor input followed by an RC filter with a time constant programmable from 200 ns up to 10 <inline-formula> <tex-math notation="LaTeX">$\mu \text{s}$ </tex-math></inline-formula>, a baseline holder and a peak stretcher stage. Multichannel readout is performed with a <italic>polling</italic> strategy thanks to two analog multiplexers. The chip has been designed to cope with input capacitance values of several nanofarad, like the ones featured by the large-sized SiPM pixels (8 mm <inline-formula> <tex-math notation="LaTeX">${\times }$ </tex-math></inline-formula> 8 mm) of the application. Despite being conceived for operation with slow scintillator crystals like cesium-iodide (CsI), this ASIC can successfully manage the readout of faster crystals like lutetium-yttrium oxyorthosilicate. Measurements of the ASIC carried out on a fully equipped SPECT detection module have shown an energy resolution of 13.7% at the 140-keV line of <sup>99m</sup>Tc, for a 5 cm <inline-formula> <tex-math notation="LaTeX">${\times }$ </tex-math></inline-formula> 5 cm gamma camera consisting of a continuous CsI crystal and a SiPM detection plane cooled at 0 °C.

[1]  N. H. Lazar,et al.  Intrinsic Scintillator Resolution , 1956, IRE Transactions on Nuclear Science.

[2]  David Gascon,et al.  PACIFIC: A 64-channel ASIC for scintillating fiber tracking in LHCb upgrade , 2015 .

[3]  E Auffray,et al.  Time-Based Readout of a Silicon Photomultiplier (SiPM) for Time of Flight Positron Emission Tomography (TOF-PET) , 2011, IEEE Transactions on Nuclear Science.

[4]  F. Dulucq,et al.  SPIROC (SiPM Integrated Read-Out Chip): Dedicated very front-end electronics for an ILC prototype hadronic calorimeter with SiPM read-out , 2007 .

[5]  Dirk Meier,et al.  An ASIC for SiPM/MPPC readout , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[6]  F. Corsi,et al.  Current-mode front-end electronics for silicon photo-multiplier detectors , 2007, 2007 2nd International Workshop on Advances in Sensors and Interface.

[7]  L. Raux,et al.  SPIROC: Design and performance of a dedicated very front-end for an ILC prototype Hadronic Calorimeter with SiPM , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[8]  Nepomuk Otte,et al.  The Silicon Photomultiplier - A new device for High Energy Physics, Astroparticle Physics, Industrial and Medical Applications , 2006 .

[9]  H. Anger,et al.  Use of a Gamma-Ray Pinhole Camera for in vivo Studies , 1952, Nature.

[10]  G. Matarrese,et al.  ASIC development for SiPM readout , 2009 .

[11]  Carlo Fiorini,et al.  A CMOS self-triggered gated integrator circuit for SiPM readout in SPECT applications , 2015, 2015 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[12]  C. Piemonte,et al.  Experimental Evaluation of a SiPM-Based Scintillation Detector for MR-Compatible SPECT Systems , 2015, IEEE Transactions on Nuclear Science.

[13]  F. Licciulli,et al.  BASIC32_ADC, a front-end ASIC for SiPM detectors , 2013, 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC).

[14]  Behnoush Sanaei,et al.  Characterization of a New Silicon Photomultiplier in Comparison with a Conventional Photomultiplier Tube , 2015 .

[15]  H. Anger,et al.  MAPPING THE DISTRIBUTION OF GAMMA-RAY-EMITTING ISOTOPES WITH THE SCINTILLATION CAMERA , 1958 .

[16]  M. Moszynski,et al.  A Comparative Study of Silicon Drift Detectors With Photomultipliers, Avalanche Photodiodes and PIN Photodiodes in Gamma Spectrometry With LaBr$_{3}$ Crystals , 2009, IEEE Transactions on Nuclear Science.

[17]  Y. Liu,et al.  TIMPIC-II: The second version time-based-readout ASIC for SSPM based PET applications , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[18]  Gabriele Bigongiari,et al.  A custom front-end ASIC for the readout and timing of 64 SiPM photosensors , 2011 .

[19]  H. Traff,et al.  Novel approach to high speed CMOS current comparators , 1992 .

[20]  E. Garutti Dedicated front-end electronics for an ILC prototype hadronic calorimeter with SiPM readout , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[21]  A. Del Guerra,et al.  BASIC: An 8-channel front-end ASIC for Silicon Photomultiplier detectors , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[22]  C. Taille,et al.  SPIROC: design and performances of a dedicated very front-end electronics for an ILC Analog Hadronic CALorimeter (AHCAL) prototype with SiPM read-out , 2013 .

[23]  G. De Geronimo,et al.  A CMOS baseline holder (BLH) for readout ASICs , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[24]  G. Temes,et al.  Fast CMOS current amplifier and buffer stage , 1987 .

[25]  G. Knoll Radiation detection and measurement , 1979 .