Absolute backscatter coefficient estimates of tissue-mimicking phantoms in the 5-50 MHz frequency range.

Absolute backscatter coefficients in tissue-mimicking phantoms were experimentally determined in the 5-50 MHz frequency range using a broadband technique. A focused broadband transducer from a commercial research system, the VisualSonics Vevo 770, was used with two tissue-mimicking phantoms. The phantoms differed regarding the thin layers covering their surfaces to prevent desiccation and regarding glass bead concentrations and diameter distributions. Ultrasound scanning of these phantoms was performed through the thin layer. To avoid signal saturation, the power spectra obtained from the backscattered radio frequency signals were calibrated by using the signal from a liquid planar reflector, a water-brominated hydrocarbon interface with acoustic impedance close to that of water. Experimental values of absolute backscatter coefficients were compared with those predicted by the Faran scattering model over the frequency range 5-50 MHz. The mean percent difference and standard deviation was 54% ± 45% for the phantom with a mean glass bead diameter of 5.40 μm and was 47% ± 28% for the phantom with 5.16 μm mean diameter beads.

[1]  Martin O Culjat,et al.  A review of tissue substitutes for ultrasound imaging. , 2010, Ultrasound in medicine & biology.

[2]  M. Birch,et al.  The acoustic properties, centered on 20 MHZ, of an IEC agar-based tissue-mimicking material and its temperature, frequency and age dependence. , 2008, Ultrasound in medicine & biology.

[3]  J. G. Miller,et al.  Interlaboratory comparison of ultrasonic backscatter, attenuation, and speed measurements. , 1999, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[4]  E. Madsen,et al.  Tests of backscatter coefficient measurement using broadband pulses , 1993, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  K. Wear Group velocity, phase velocity, and dispersion in human calcaneus in vivo. , 2007, The Journal of the Acoustical Society of America.

[6]  C. Moran,et al.  Characterising the performance of a high resolution ultrasound scanner for pre-clinical ultrasound imaging , 2008, 2008 IEEE Ultrasonics Symposium.

[7]  E. Madsen,et al.  Low-reflection-coefficient liquid interfaces for system characterization. , 2001, Ultrasound in medicine & biology.

[8]  E. Madsen,et al.  Interlaboratory Comparison of Ultrasonic Backscatter Coefficient Measurements From 2 to 9 MHz , 2005, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[9]  M A Srinivasan,et al.  High-frequency ultrasonic attenuation and backscatter coefficients of in vivo normal human dermis and subcutaneous fat. , 2001, Ultrasound in medicine & biology.

[10]  E. Parzen On Estimation of a Probability Density Function and Mode , 1962 .

[11]  M. Rosenblatt Remarks on Some Nonparametric Estimates of a Density Function , 1956 .

[12]  Ernest L Madsen,et al.  Anechoic sphere phantoms for estimating 3-D resolution of very-high-frequency ultrasound scanners , 2010, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[13]  J. Faran Sound Scattering by Solid Cylinders and Spheres , 1951 .

[14]  V. A. D. Grosso,et al.  Speed of Sound in Pure Water , 1972 .

[15]  L. X. Yao,et al.  Backscatter Coefficient Measurements Using a Reference Phantom to Extract Depth-Dependent Instrumentation Factors , 1990, Ultrasonic imaging.

[16]  E. Madsen,et al.  Method of data reduction for accurate determination of acoustic backscatter coefficients. , 1984, The Journal of the Acoustical Society of America.