Capillary electrophoresis artifact due to eosin: implications for the interpretation of molecular diagnostic assays.

Capillary electrophoresis (CE) is a commonly used tool in the analysis of fluorescently labeled PCR amplification products. We have identified a CE artifact caused by the tissue stain eosin that can complicate the interpretation of CE data. The artifact was detected during routine analysis of a DNA sample isolated from a formalin-fixed, paraffin-embedded tissue sample considered histologically suspicious for a B-cell neoplasm. A standard clinical PCR and CE assay for immunoglobulin heavy chain (IGH) gene rearrangement revealed a weak polyclonal population of rearranged IGH genes and a 71 base peak suspicious for IGH clonality. The spectral properties of the 71 base peak were unusual in that although the dominant fluorescence of the peak was blue, it also fluoresced in green and yellow (blue>green>yellow), raising the suspicion that the peak might represent an artifact. CE analysis of the genomic DNA sample without PCR amplification demonstrated the presence of the 71 base peak, suggesting that the artifact was caused by a contaminant within the DNA sample itself. We demonstrate that eosin, which was used to stain the formalin-fixed tissue during processing, yields a discrete 71 base peak of similar morphology to the contaminant peak on CE analysis. The data suggest that eosin in the fixed tissue was not completely eliminated during nucleic acid extraction, resulting in the artifact peak. We discuss the implications of this potentially common contaminant on the interpretation of CE data and demonstrate that artifacts caused by eosin can be avoided by using more stringent DNA purification steps. Histological dyes may fluoresce, and artifacts from them should be considered when primary peaks contain additional underlying peaks of other colors.

[1]  J. Farley,et al.  Determining eosin as a groundwater migration tracer by capillary electrophoresis/laser‐induced fluorescence using a multiwavelength laser , 2003, Electrophoresis.

[2]  T. Lion,et al.  Sequential monitoring of chimerism and detection of minimal residual disease after allogeneic blood stem cell transplantation (BSCT) using multiplex PCR amplification of short tandem repeat-markers , 2001, Leukemia.

[3]  J. Eshleman,et al.  Detection of microsatellite instability by fluorescence multiplex polymerase chain reaction. , 2000, The Journal of molecular diagnostics : JMD.

[4]  J. Eshleman,et al.  Pseudo-spikes are common in histologically benign lymphoid tissues. , 2000, The Journal of molecular diagnostics : JMD.

[5]  J. Eshleman,et al.  Molecular Diagnosis of Oligodendroglioma in Paraffin Sections , 2003, Laboratory Investigation.

[6]  D. Clark,et al.  Inhibitors of the polymerase chain reaction in Papanicolaou stain. Removal with a simple destaining procedure. , 1996, Acta cytologica.

[7]  N. Beaubier,et al.  Comparison of Capillary Electrophoresis and Polyacrylamide Gel Electrophoresis for the Evaluation of T and B Cell Clonality by Polymerase Chain Reaction , 2000, Diagnostic molecular pathology : the American journal of surgical pathology, part B.

[8]  M. Gulley,et al.  Comparison of histologic stains for use in PCR analysis of microdissected, paraffin-embedded tissues. , 1998, BioTechniques.

[9]  L. Medeiros,et al.  Automated high-resolution polymerase chain reaction fragment analysis: a method for detecting T-cell receptor gamma-chain gene rearrangements in lymphoproliferative diseases. , 1998, The American journal of pathology.

[10]  Kevin J. Trainor,et al.  Monoclonality in B-lymphoproliferative disorders detected at the DNA level. , 1990, Blood.

[11]  A. Morley,et al.  Development of a highly sensitive assay, based on the polymerase chain reaction, for rare B‐lymphocyte clones in a polyclonal population , 1990, British Journal of Haematology.

[12]  Á. Ríos,et al.  Use of eosin as a fluorophore in capillary electrophoresis with laser detection. , 2001, Journal of chromatography. A.

[13]  J. Landers,et al.  Evaluation of capillary electrophoresis in polymer solutions with laser‐induced fluorescence detection for the automated detection of T‐cell gene rearrangements in lymphoproliferative disorders , 1996, Electrophoresis.

[14]  T. Lion,et al.  Interference of dye-associated fluorescence signals with quantitative analysis of chimerism by capillary electrophoresis , 2003, Leukemia.