Traceability of Laser Frequency Calibrations at MIKES

We describe the traceability chain of length measurements at Centre for Metrology and Accreditation (MIKES) from atomic clocks to the lasers of primary interferometers. Crucial part of the traceability chain, an optical frequency comb generator linking optical frequencies to atomic clocks, is described in detail. The frequency comb generator is used in frequency calibrations of iodine-stabilized lasers, which are operated in compliance with the recommendations of the Practical Realization of the Definition of the Meter. Measured absolute frequencies of iodine-stabilized lasers, time records of the measurements, and the respective Allan deviations demonstrate the solid performance of the MIKES laser frequency standards. The results are in good agreement with the recommended values, as well as with the independent characterizations of the measured lasers.

[1]  T J Quinn,et al.  Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards (2001) , 2003 .

[2]  A Lassila,et al.  Interferometer for calibration of graduated line scales with a moving CCD camera as a line detector. , 1994, Applied optics.

[3]  Development of iodine stabilized HeNe lasers at MRI , 1995 .

[4]  Hall,et al.  Carrier-envelope phase control of femtosecond mode-locked lasers and direct optical frequency synthesis , 2000, Science.

[5]  Mikko Merimaa,et al.  Frequency stabilization of a diode-pumped Nd:YAG laser at 532 nm to iodine by using third-harmonic technique , 2002, CPEM 2002.

[6]  Mikko Merimaa,et al.  Optical amplifier for femtosecond frequency comb measurements near 633 nm , 2005 .

[7]  Hall,et al.  Direct link between microwave and optical frequencies with a 300 THz femtosecond laser comb , 2000, Physical review letters.

[8]  H R Simonsen,et al.  International comparison of He-Ne lasers stabilized with 127I2 at λ = 543 nm (December 1999) - Northern European lasers , 2000 .

[9]  Erkki Ikonen,et al.  Gauge-block Interferometer based on one Stabilized Laser and a White-light Source , 1993 .

[10]  Robert Windeler,et al.  International comparisons of femtosecond laser frequency combs , 2005, IEEE Transactions on Instrumentation and Measurement.

[11]  Theodor W. Hänsch,et al.  Absolute Optical Frequency Measurement of the Cesium D 1 Line with a Mode-Locked Laser , 1999 .