Mechanical properties of porcine pulmonary valve leaflets: how do they differ from aortic leaflets?

This study measured the biaxial properties of pulmonary and aortic leaflets in extension in the fresh state and then in the same samples after fixation with glutaraldehyde. The results showed that when fresh, the valves had a similar response to load in the circumferential direction, but the pulmonary leaflets were more extensible in the radial direction. They were also less stiff. Fixation decreased the tissue extensibility and increased the stiffness of the pulmonary leaflets much less than in the aortic leaflets. This was interpreted to mean that the collagen content of the pulmonary leaflets is significantly less than that in the aortic leaflets. Reduced collagen content would be expected to enhance hemodynamic performance because of increased leaflet stretch and reduced stiffness. However, lower collagen levels may reduce implant durability.

[1]  G. Christie,et al.  The stentless bioprosthesis: surgical challenges and implications for long-term durability. , 1992, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[2]  G. Christie,et al.  Identification of a Failure Mode of the Antibiotic Sterilized Aortic Allograft After 10 Years: Implications for Their Long‐Term Survival , 1991, Journal of cardiac surgery.

[3]  Y Lanir,et al.  A structural theory for the homogeneous biaxial stress-strain relationships in flat collagenous tissues. , 1979, Journal of biomechanics.

[4]  M. Lane,et al.  Pulmonary autograft in children: realized growth potential. , 1994, The Annals of thoracic surgery.

[5]  G. Christie,et al.  Anatomy of aortic heart valve leaflets: the influence of glutaraldehyde fixation on function. , 1992, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[6]  F. J. Thomson,et al.  Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis. , 1979, Thorax.

[7]  B H Smaill,et al.  An assessment of the mechanical properties of leaflets from four second-generation porcine bioprostheses with biaxial testing techniques. , 1989, The Journal of thoracic and cardiovascular surgery.

[8]  M. Dobke,et al.  Biomechanics of the pulmonary autograft valve in the aortic position. , 1982, Thorax.

[9]  G. Gerosa,et al.  The Pulmonary Valve: Is It Mechanically Suitable for Use as an Aortic Valve Replacement? , 1994, ASAIO journal.