Improving Fat-Suppressed T2-Weighted Imaging of the Head and Neck with 2 Fast Spin-Echo Dixon Techniques: Initial Experiences

SUMMARY: Two modified fast spin-echo (FSE) techniques (a 2-point and a single-scan triple-echo Dixon) were used for T2-weighted imaging of the head and neck in 7 patients along with conventional FSE with fat saturation. Both Dixon techniques provided consistent and more uniform fat suppression (FS) than conventional FSE. The 2-point Dixon technique was noted to be more susceptible to motion artifacts. The triple-echo Dixon technique offered the best scan time efficiency and overall image quality.

[1]  Jean A. Tkach,et al.  Separation of fat and water in fast spin‐echo MR imaging with the three‐point dixon technique , 1995, Journal of magnetic resonance imaging : JMRI.

[2]  R V Mulkern,et al.  Fast three-point Dixon MR imaging of the retrobulbar space with low-resolution images for phase correction: comparison with fast spin-echo inversion recovery imaging. , 2001, AJNR. American journal of neuroradiology.

[3]  Jingfei Ma,et al.  Fast spin‐echo triple‐echo dixon (fTED) technique for efficient T2‐weighted water and fat imaging , 2007, Magnetic resonance in medicine.

[4]  Michael Markl,et al.  Multicoil Dixon chemical species separation with an iterative least‐squares estimation method , 2004, Magnetic resonance in medicine.

[5]  W. T. Dixon Simple proton spectroscopic imaging. , 1984, Radiology.

[6]  J Hennig,et al.  RARE imaging: A fast imaging method for clinical MR , 1986, Magnetic resonance in medicine.

[7]  Jingfei Ma,et al.  Breath‐hold water and fat imaging using a dual‐echo two‐point dixon technique with an efficient and robust phase‐correction algorithm , 2004, Magnetic resonance in medicine.

[8]  A V Barger,et al.  Fat signal suppression in head and neck imaging using fast spin-echo-IDEAL technique. , 2006, AJNR. American journal of neuroradiology.

[9]  Jingfei Ma,et al.  T2‐weighted spine imaging with a fast three‐point dixon technique: Comparison with chemical shift selective fat suppression , 2004, Journal of magnetic resonance imaging : JMRI.

[10]  G H Glover,et al.  Three‐point dixon technique for true water/fat decomposition with B0 inhomogeneity correction , 1991, Magnetic resonance in medicine.

[11]  Jingfei Ma,et al.  A fast spin echo two-point Dixon technique and its combination with sensitivity encoding for efficient T2-weighted imaging. , 2005, Magnetic resonance imaging.