Feasibility of imaging epileptic seizure onset with EIT and depth electrodes

&NA; Imaging ictal and interictal activity with Electrical Impedance Tomography (EIT) using intracranial electrode mats has been demonstrated in animal models of epilepsy. In human epilepsy subjects undergoing presurgical evaluation, depth electrodes are often preferred. The purpose of this work was to evaluate the feasibility of using EIT to localise epileptogenic areas with intracranial electrodes in humans. The accuracy of localisation of the ictal onset zone was evaluated in computer simulations using 9M element FEM models derived from three subjects. 5 mm radius perturbations imitating a single seizure onset event were placed in several locations forming two groups: under depth electrode coverage and in the contralateral hemisphere. Simulations were made for impedance changes of 1% expected for neuronal depolarisation over milliseconds and 10% for cell swelling over seconds. Reconstructions were compared with EEG source modelling for a radially orientated dipole with respect to the closest EEG recording contact. The best accuracy of EIT was obtained using all depth and 32 scalp electrodes, greater than the equivalent accuracy with EEG inverse source modelling. The localisation error was 5.2 ± 1.8, 4.3 ± 0 and 46.2 ± 25.8 mm for perturbations within the volume enclosed by depth electrodes and 29.6 ± 38.7, 26.1 ± 36.2, 54.0 ± 26.2 mm for those without (EIT 1%, 10% change, EEG source modelling, n = 15 in 3 subjects, p < 0.01). As EIT was insensitive to source dipole orientation, all 15 perturbations within the volume enclosed by depth electrodes were localised, whereas the standard clinical method of visual inspection of EEG voltages, only localised 8 out of 15 cases. This suggests that adding EIT to SEEG measurements could be beneficial in localising the onset of seizures. HighlightsA new method is proposed to use EIT to localise and ictal activity in patients with depth electrodes.The location accuracy was improved with the best EIT protocol than EEG inverse source or SEEG detection in simulations.EIT was not sensitive to dipole orientation, while EEG detection varied with the field angle demonstrated in modelling.A combination of EIT and SEEG can potentially improve the diagnostic yield in epilepsy.

[1]  P. Ryvlin,et al.  Predicting epilepsy surgery outcome. , 2016, Current opinion in neurology.

[2]  W. R. Adey,et al.  Impedance Changes during Epileptic Seizures , 1966, Epilepsia.

[3]  W. R. Adey,et al.  Impedance measurements in brain tissue of animals using microvolt singals. , 1962, Experimental neurology.

[4]  David S. Holder,et al.  Electrical Impedance Tomography : Methods, History and Applications , 2004 .

[5]  Martin Schuettler,et al.  A novel method for recording neuronal depolarization with recording at 125–825 Hz: implications for imaging fast neural activity in the brain with electrical impedance tomography , 2011, Medical & Biological Engineering & Computing.

[6]  D.B. McCreery,et al.  Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation , 1990, IEEE Transactions on Biomedical Engineering.

[7]  Kirill Y Aristovich,et al.  A method for reconstructing tomographic images of evoked neural activity with electrical impedance tomography using intracranial planar arrays , 2014, Physiological measurement.

[8]  J. Pernier,et al.  A systematic evaluation of the spherical model accuracy in EEG dipole localization. , 1997, Electroencephalography and clinical neurophysiology.

[9]  J. Schadé,et al.  Changes in the electrical conductivity of cerebral cortex during seizure activity. , 1962, Experimental neurology.

[10]  Lior Horesh,et al.  Some novel approaches in modelling and image reconstruction for multi-frequency Electrical Impedance Tomography of the human brain , 2006 .

[11]  John S Duncan,et al.  Adult epilepsy , 2006, The Lancet.

[12]  Kazutomo Yunokuchi,et al.  Tests of EEG localization accuracy using implanted sources in the human brain , 1991, Annals of neurology.

[13]  John S Duncan,et al.  The long-term outcome of adult epilepsy surgery, patterns of seizure remission, and relapse: a cohort study , 2011, The Lancet.

[14]  S. Smith EEG in the diagnosis, classification, and management of patients with epilepsy , 2005, Journal of Neurology, Neurosurgery & Psychiatry.

[15]  Robert Galambos,et al.  Resistance Shifts Accompanying the Evoked Cortical Response in the Cat , 1967, Science.

[16]  G. Regesta,et al.  Clinical aspects and biological bases of drug-resistant epilepsies , 1999, Epilepsy Research.

[17]  Robert Oostenveld,et al.  The five percent electrode system for high-resolution EEG and ERP measurements , 2001, Clinical Neurophysiology.

[18]  Anna N. Vongerichten,et al.  Characterisation and imaging of cortical impedance changes during interictal and ictal activity in the anaesthetised rat , 2016, NeuroImage.

[19]  Andreas Schulze-Bonhage,et al.  Source reconstruction based on subdural EEG recordings adds to the presurgical evaluation in refractory frontal lobe epilepsy , 2013, Clinical Neurophysiology.

[20]  Maja O’Connor LONGITUDINAL STUDY , 2013 .

[21]  P Kahane,et al.  Intracranial EEG and human brain mapping , 2003, Journal of Physiology-Paris.

[22]  Bruce J Fisch,et al.  Epilepsy and Intensive Care Monitoring: Principles and Practice , 2009 .

[23]  M. Cook,et al.  Surgery for temporal lobe epilepsy associated with mesial temporal sclerosis in the older patient: A long‐term follow‐up , 2010, Epilepsia.

[24]  E. Niedermeyer The epilepsies : diagnosis and management , 1990 .

[25]  I. Najm,et al.  A longitudinal study of surgical outcome and its determinants following posterior cortex epilepsy surgery , 2009, Epilepsia.

[26]  Bart Vanrumste,et al.  Journal of Neuroengineering and Rehabilitation Open Access Review on Solving the Inverse Problem in Eeg Source Analysis , 2022 .

[27]  David Holder,et al.  Are patient specific meshes required for EIT head imaging? , 2016, Physiological measurement.

[28]  D. Djajaputra Electrical Impedance Tomography: Methods, History and Applications , 2005 .

[29]  H. Lüders,et al.  Textbook of epilepsy surgery , 2008 .

[30]  Jorge Gonzalez-Martinez,et al.  Is SEEG safe? A systematic review and meta‐analysis of stereo‐electroencephalography–related complications , 2016, Epilepsia.

[31]  John S. Duncan,et al.  Selecting patients for epilepsy surgery: Synthesis of data , 2011, Epilepsy & Behavior.

[32]  David S. Holder,et al.  Imaging fast electrical activity in the brain with electrical impedance tomography , 2016, NeuroImage.

[33]  Anthony M. Norcia,et al.  In‐vivo measurements of human brain tissue conductivity using focal electrical current injection through intracerebral multicontact electrodes , 2017, Human brain mapping.

[34]  Christian Bottomley,et al.  Estimation of the burden of active and life-time epilepsy: A meta-analytic approach , 2010, Epilepsia.

[35]  K. Holthoff,et al.  A Novel Role of Vasopressin in the Brain: Modulation of Activity-Dependent Water Flux in the Neocortex , 2001, The Journal of Neuroscience.

[36]  Simon Arridge,et al.  Stroke type differentiation using spectrally constrained multifrequency EIT: evaluation of feasibility in a realistic head model , 2014, Physiological measurement.

[37]  Jean Gotman,et al.  Diagnostic utility of invasive EEG for epilepsy surgery: Indications, modalities, and techniques , 2016, Epilepsia.

[38]  Carlos H. Muravchik,et al.  Electrode and brain modeling in stereo-EEG , 2012, Clinical Neurophysiology.

[39]  M. Teplan FUNDAMENTALS OF EEG MEASUREMENT , 2002 .

[40]  D. Cohen,et al.  MEG versus EEG localization test using implanted sources in the human brain , 1990, Annals of neurology.

[41]  R H Bayford,et al.  Multi-frequency electrical impedance tomography (EIT) of the adult human head: initial findings in brain tumours, arteriovenous malformations and chronic stroke, development of an analysis method and calibration , 2006, Physiological measurement.

[42]  C D Binnie,et al.  Factors limiting the application of electrical impedance tomography for identification of regional conductivity changes using scalp electrodes during epileptic seizures in humans , 2006, Physiological measurement.

[43]  Michael J Lang,et al.  Complications of subdural and depth electrodes in 269 patients undergoing 317 procedures for invasive monitoring in epilepsy , 2016, Epilepsia.

[44]  D L Collins,et al.  White-matter diffusion abnormalities in temporal-lobe epilepsy with and without mesial temporal sclerosis , 2008, Journal of Neurology, Neurosurgery, and Psychiatry.

[45]  G. Alarcón,et al.  Electrophysiological aspects of interictal and ictal activity in human partial epilepsy , 1996, Seizure.

[46]  C. Elger,et al.  Clinical Relevance of Quantified Intracranial Interictal Spike Activity in Presurgical Evaluation of Epilepsy , 2000, Epilepsia.

[47]  Andreas Dedner,et al.  A Fast Parallel Solver for the Forward Problem in Electrical Impedance Tomography , 2015, IEEE Transactions on Biomedical Engineering.

[48]  Intracranial electroencephalography and localization studies , 2012 .

[49]  James Avery,et al.  A Versatile and Reproducible Multi-Frequency Electrical Impedance Tomography System , 2017, Sensors.

[50]  G. Jackson,et al.  Long‐term seizure outcome and risk factors for recurrence after extratemporal epilepsy surgery , 2012, Epilepsia.

[51]  Théodore Papadopoulo,et al.  The adjoint method for general EEG and MEG sensor-based lead field equations , 2009, Physics in medicine and biology.

[52]  U. Heinemann,et al.  Ionic changes and alterations in the size of the extracellular space during epileptic activity. , 1986, Advances in neurology.

[53]  B. MacVicar,et al.  Imaging cell volume changes and neuronal excitation in the hippocampal slice , 1994, Neuroscience.

[54]  J Gotman,et al.  Reliability of dipole models of epileptic spikes , 1999, Clinical Neurophysiology.

[55]  Kaspar Anton Schindler,et al.  Ictal time-irreversible intracranial EEG signals as markers of the epileptogenic zone , 2016, Clinical Neurophysiology.

[56]  James Avery,et al.  Correcting electrode modelling errors in EIT on realistic 3D head models , 2015, Physiological measurement.

[57]  Feng Gao,et al.  Real-time imaging of epileptic seizures in rats using electrical impedance tomography , 2017, Neuroreport.

[58]  B. Burle,et al.  Spatial and temporal resolutions of EEG: Is it really black and white? A scalp current density view , 2015, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[59]  J. Ebersole Defining epileptogenic foci: past, present, future. , 1997, Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society.