Megavoltage cone-beam computed tomography using a high-efficiency image receptor.

PURPOSE To develop an image receptor capable of forming high-quality megavoltage CT images using modest radiation doses. METHODS AND MATERIALS A flat-panel imaging system consisting of a conventional flat-panel sensor attached to a thick CsI scintillator has been fabricated. The scintillator consists of individual CsI crystals 8 mm thick by 0.38 mm x 0.38-mm pitch. Five sides of each crystal are coated with a reflecting powder/epoxy mixture, and the sixth side is in contact with the flat-panel sensor. A timing interface coordinates acquisition by the imaging system and pulsing of the linear accelerator. With this interface, as little as one accelerator pulse (0.023 cGy at the isocenter) can be used to form projection images. Different CT phantoms irradiated by a 6-MV X-ray beam have been imaged to evaluate the performance of the imaging system. The phantoms have been mounted on a rotating stage and rotated while 360 projection images are acquired in 48 s. These projections have been reconstructed using the Feldkamp cone-beam CT reconstruction algorithm. RESULTS AND DISCUSSION Using an irradiation of 16 cGy (360 projections x 0.046 cGy/projection), the contrast resolution is approximately 1% for large objects. High-contrast structures as small as 1.2 mm are clearly visible. The reconstructed CT values are linear (R(2) = 0.98) for electron densities between 0.001 and 2.16 g/cm(3), and the reconstruction time for a 512 x 512 x 512 data set is 6 min. Images of an anthropomorphic phantom show that soft-tissue structures such as the heart, lung, kidneys, and liver are visible in the reconstructed images (16 cGy, 5-mm-thick slices). CONCLUSIONS The acquisition of megavoltage CT images with soft-tissue contrast is possible with irradiations as small as 16 cGy.

[1]  C C Ling,et al.  Cone-beam CT with megavoltage beams and an amorphous silicon electronic portal imaging device: potential for verification of radiotherapy of lung cancer. , 2002, Medical physics.

[2]  Robert Jeraj,et al.  Development of a new multielement detector system for megavoltage photons , 2002, SPIE Medical Imaging.

[3]  K. Langen,et al.  Organ motion and its management. , 2001, International journal of radiation oncology, biology, physics.

[4]  Isaias D. Job,et al.  40 x 30 cm flat-panel imager for angiography, R&F, and cone-beam CT applications , 2001 .

[5]  W Swindell,et al.  A 4-MV CT scanner for radiation therapy: the prototype system. , 1982, Medical physics.

[6]  J O Deasy,et al.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. , 1993, Medical physics.

[7]  H Guan,et al.  Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multi-level scheme algebraic reconstruction technique (MLS-ART) , 1998, Physics in medicine and biology.

[8]  C. Ling,et al.  Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: a strategy for large-scale implementation for the treatment of carcinoma of the prostate. , 1997, International journal of radiation oncology, biology, physics.

[9]  K. Nakagawa,et al.  Technological features and clinical feasibility of megavoltage CT scanning , 1992, European Radiology.

[10]  Cedric X. Yu,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. , 1995, Physics in medicine and biology.

[11]  Giselher G. Lichti,et al.  Influence of radiation damage on BGO scintillation properties 1 1 Supported by the Deutsches Zentrum für Luft- und Raumfahrt e. V (DLR) under Contract 50.OG.9503.0 and partly by the Deutsche Forschungsgemeinschaft (DFG) under Contract Ri 242/12-1. , 1998 .

[12]  S. Webb,et al.  Rapid portal imaging with a high-efficiency, large field-of-view detector. , 1998, Medical physics.

[13]  A Brahme,et al.  Radiotherapeutic computed tomography with scanned photon beams. , 1987, International journal of radiation oncology, biology, physics.

[14]  T. Mackie,et al.  Megavoltage CT on a tomotherapy system. , 1999, Physics in medicine and biology.

[15]  T. Bortfeld,et al.  X-ray field compensation with multileaf collimators. , 1994, International journal of radiation oncology, biology, physics.

[16]  F F Yin,et al.  Adaptive portal CT reconstruction: a simulation study. , 2000, Medical physics.

[17]  J Dudson,et al.  A feasibility study for megavoltage cone beam CT using a commercial EPID. , 1998, Physics in medicine and biology.

[18]  D. G. Lewis,et al.  A linear array, scintillation crystal-photodiode detector for megavoltage imaging. , 1991, Medical physics.

[19]  K W Leszczynski,et al.  On few-view tomographic reconstruction with megavoltage photon beams. , 2001, Medical physics.

[20]  K Nakagawa,et al.  Real-time beam monitoring in dynamic conformation therapy. , 1994, International journal of radiation oncology, biology, physics.

[21]  R Mohan,et al.  Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation. , 1996, International journal of radiation oncology, biology, physics.

[22]  Mageras,et al.  Control, Correction, and Modeling of Setup Errors and Organ Motion. , 1995, Seminars in radiation oncology.

[23]  T Haycocks,et al.  Positioning errors and prostate motion during conformal prostate radiotherapy using on-line isocentre set-up verification and implanted prostate markers. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[24]  W Swindell,et al.  Computed tomography with a linear accelerator with radiotherapy applications. , 1983, Medical physics.

[25]  P. Evans,et al.  Optimization of the scintillation detector in a combined 3D megavoltage CT scanner and portal imager. , 1998, Medical physics.

[26]  D A Jaffray,et al.  A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets. , 1999, International journal of radiation oncology, biology, physics.

[27]  B A Groh,et al.  Tomotherapeutic portal imaging for radiation treatment verification. , 1998, Physics in medicine and biology.

[28]  W Schlegel,et al.  Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimation. , 1994, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[29]  P. Munro,et al.  Evaluation of a high-density scintillating glass for portal imaging. , 1996, Medical Physics (Lancaster).

[30]  C. Langer,et al.  Concurrent chemoradiation using paclitaxel and carboplatin in locally advanced non-small cell lung cancer. , 1999, Seminars in radiation oncology.

[31]  D. G. Lewis,et al.  A megavoltage CT scanner for radiotherapy verification. , 1992, Physics in medicine and biology.

[32]  S Mutic,et al.  Abutment region dosimetry for serial tomotherapy. , 1999, International journal of radiation oncology, biology, physics.

[33]  R Jeraj,et al.  Monte Carlo study of a highly efficient gas ionization detector for megavoltage imaging and image-guided radiotherapy. , 2002, Medical physics.

[34]  J. Kapatoes,et al.  Megavoltage CT image reconstruction during tomotherapy treatments. , 2000, Physics in medicine and biology.