Capacitive driven-right-leg grounding in Indirect-contact ECG measurement

For the reduction of common-mode noise level in Indirect-contact ECG (IDC-ECG) measurement, a driven-right-leg grounding method was applied to the IDC-ECG. Because the IDC-ECG does not require any direct contact between the electrodes and the human skin, it is adequate for un-constraining long-term ECG measurement at home and its various applications are now under development. However, larger 60 Hz noise induced by power line appears in IDC-ECG than in conventional ECG, that is a restriction of IDC-ECG application. In this study, the driven-right-leg ground which has been used in conventional direct-contact ECG, was adapted to the IDC-ECG measurement, by feedback of the inversion of amplified common-mode noise to the body through the conductive textile laid on the chair seat. It was shown that the level of 60Hz power line noise was reduced to about -40 dB when the driven-right-leg gain was 1000.

[1]  Ramon Pallàs-Areny,et al.  A comprehensive model for power line interference in biopotential measurements , 2000, IEEE Trans. Instrum. Meas..

[2]  A. Pedotti,et al.  ECG monitoring through environmental electrodes , 2000, 1st Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine and Biology. Proceedings (Cat. No.00EX451).

[3]  T. Tamura,et al.  Fully automated health monitoring system in the home , 1998 .

[4]  M. Ishijima Monitoring of electrocardiograms in bed without utilizing body surface electrodes , 1993, IEEE Transactions on Biomedical Engineering.

[5]  G. Breithardt,et al.  Heart rate variability: standards of measurement, physiological interpretation and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. , 1996 .

[6]  A. Malliani,et al.  Heart rate variability. Standards of measurement, physiological interpretation, and clinical use , 1996 .

[7]  Yong Gyu Lim,et al.  ECG measurement on a chair without conductive contact , 2006, IEEE Transactions on Biomedical Engineering.

[8]  Ramon Pallàs-Areny,et al.  AC-coupled front-end for biopotential measurements , 2003, IEEE Transactions on Biomedical Engineering.

[9]  Youngjoon Chee,et al.  Development of a nonintrusive blood pressure estimation system for computer users. , 2007, Telemedicine journal and e-health : the official journal of the American Telemedicine Association.

[10]  Kajiro Watanabe,et al.  Noninvasive measurement of heartbeat, respiration, snoring and body movements of a subject in bed via a pneumatic method , 2005, IEEE Transactions on Biomedical Engineering.

[11]  Choi Jin-Wook,et al.  A New Method for Unconstrained Pulse Arrival Time (PAT) Measurement on a Chair , 2006 .

[12]  John G. Webster,et al.  Driven-right-leg circuit design , 1983, IEEE Transactions on Biomedical Engineering.

[13]  R. Pallas-Areny,et al.  Interference-rejection characteristics of biopotential amplifiers: a comparative analysis , 1988, IEEE Transactions on Biomedical Engineering.

[14]  John G. Webster,et al.  Reductionl of Interference Due to Common Mode Voltage in Biopotential Amplifiers , 1983, IEEE Transactions on Biomedical Engineering.

[15]  C A Grimbergen,et al.  High-quality recording of bioelectric events , 1991, Medical and Biological Engineering and Computing.

[16]  J. Alihanka,et al.  A static charge sensitive bed. A new method for recording body movements during sleep. , 1979, Electroencephalography and clinical neurophysiology.

[17]  Yu Hen Hu,et al.  Power-Line Interference Detection and Suppression in ECG Signal Processing , 2008, IEEE Transactions on Biomedical Engineering.

[18]  Yong Gyu Lim,et al.  ECG Recording on a Bed During Sleep Without Direct Skin-Contact , 2007, IEEE Transactions on Biomedical Engineering.