Uncertainty estimation of a liquid flow standard system with small flow rates

Abstract A liquid flow standard system is used to calibrate liquid volume of fuel–oil flow meters at small flow rates between 50 L/h and 700 L/h. However, the system has not been used to calibrate volume flow rate because the system is only operated with the standing-start-and-finish mode. In this study, the liquid flow standard system was rebuilt to provide a calibration service of volume flow rate by attaching two flow diverters, which can operate the system with the flying-start-and-finish mode. To evaluate its performance for volume flow metering, several techniques were introduced. First, diverter timing errors were estimated by linear regression. Second, covariance between buoyancy correction factor and water density was obtained to consider interdependency between the two measurands. Third, calibration and measurement capability (CMC) was evaluated by setting a fixed value of collected weight or elapsed time for flow diversion. Finally, several CMCs were compared to find the best measurement condition. As a result of the above approach, the CMC of the liquid flow standard system was found to be (0.10–0.52)% ( k = 2) for (50–700) L/h with a minimum collected weight at 10 kg.

[1]  T. T. Yeh,et al.  Description and Uncertainty Analysis of NISTs 20 Liter Hydrocarbon Liquid Flow Standard (20 L HLFS) | NIST , 2010 .

[2]  N Bignell,et al.  Recommended table for the density of water between 0 ?C and 40 ?C based on recent experimental reports , 2001 .

[3]  Hans-Joachim Baade,et al.  Model-based flow diverter analysis for an improved uncertainty determination in liquid flow calibration facilities , 2010 .

[4]  Hans-Joachim Baade,et al.  Water density determination in high-accuracy flowmeter calibration — Measurement uncertainties and practical aspects , 2012 .

[6]  Takashi Shimada,et al.  Development of a new diverter system for liquid flow calibration facilities , 2003 .

[7]  T. T. Yeh,et al.  Diverter Uncertainty Less Than 0.01 percent for Water Flow Calibrations , 2006 .

[9]  S. Standard GUIDE TO THE EXPRESSION OF UNCERTAINTY IN MEASUREMENT , 2006 .

[10]  W. Wagner,et al.  The IAPWS Formulation 1995 for the Thermodynamic Properties of Ordinary Water Substance for General and Scientific Use , 2002 .

[11]  Hans-Joachim Baade,et al.  Quantifying impacts on the measurement uncertainty in flow calibration arising from dynamic flow effects , 2015 .

[12]  Elsa Batista,et al.  The selection of water property formulae for volume and flow calibration , 2008 .

[13]  Masaki Takamoto,et al.  Development of weighing tank system employing rotating double wing diverter , 2006 .

[14]  D Brynn Hibbert,et al.  The uncertainty of a result from a linear calibration. , 2006, The Analyst.