The Effect of an Implantable Doppler Probe on the Salvage of Microvascular Tissue Transplants

&NA; One hundred forty‐seven flaps in 135 consecutive patients undergoing microvascular transplantation were monitored using a miniature Doppler ultrasonic probe. Using a modification of a technique described previously by Swartz, the probes were secured to the outflow vein of the flap with Vicryl mesh. Twenty instances of thrombosis or spasm were detected in 16 patients, and all flaps were salvaged (100 percent). There were four false positive and no false negative results. This probe allows for safe, continuous monitoring of flap blood flow, which permits the rapid detection and hence rapid treatment of postoperative complications. Our experience suggests that a significant improvement in the salvage rate of microvascular transplants may be attainable with the use of this device. (Plast. Reconstr. Surg. 101: 1268, 1998.)

[1]  Michael J. Miller,et al.  Implantable Venous Doppler Microvascular Monitoring: Laboratory Investigation and Clinical Results , 1994, Plastic and reconstructive surgery.

[2]  N. Jones,et al.  Intraoperative and postoperative monitoring of microsurgical free tissue transfers. , 1992, Clinics in plastic surgery.

[3]  S. Suominen,et al.  Free flap failures , 1995, Microsurgery.

[4]  T. Harashina Analysis of 200 free flaps. , 1988, British journal of plastic surgery.

[5]  H. Machens,et al.  Vascular complications after free tissue transfer , 1995, Microsurgery.

[6]  C. S. Jones,et al.  The role of emergent exploration in free-tissue transfer: a review of 150 consecutive cases. , 1990, Plastic and reconstructive surgery.

[7]  W. Shaw,et al.  Reconstruction of the lower extremity with microvascular free flaps: a 10-year experience with 304 consecutive cases. , 1989, The Journal of trauma.

[8]  M. Wood,et al.  Experience with One Hundred Consecutive Free Flaps , 1987, Annals of plastic surgery.

[9]  W. Frayer,et al.  LASER DOPPLER MEASUREMENT OF BLOOD FLOW IN THE MICROCIRCULATION , 1978, Plastic and reconstructive surgery.

[10]  K. Black,et al.  Transcutaneous PO2 in Flaps: A New Method of Survival Prediction , 1980, Plastic and reconstructive surgery.

[11]  Peter C. Neugan,et al.  Monitoring techniques for the detection of flow failure in the postoperative period , 1993 .

[12]  N. Georgiade,et al.  Photoplethysmography as an aid in determining the viability of pedicle flaps. , 1969, Plastic and reconstructive surgery.