Evaluation of monolithic HPLC columns for various pharmaceutical separations: method transfer from conventional phases and batch to batch repeatability.

Methods developed on conventional particle-packed C18 columns for pilocarpine, propranolol, glibenclamide, glimepiride, insulin and their respective degradation products or related compounds were transferred from the conventional Superspher 100RP-18e column to Chromolith Performance RP-18e columns. All transfers were successful applying the same chromatographic conditions, except for insulin where the acetonitrile content of the mobile phase was reduced by 0.5%. The intraday and interday precisions for both retention time and peak area were evaluated over a wide concentration range. Results were found to be equal, or slightly better on Chromolith Performance with RSD%<1.1% in all cases. Monolithic batch to batch repeatability of both retention time and peak area, compared for monolithic columns from different batches gave an RSD% of less than 1.3%. The separation of each drug and its related products was investigated on monolithic columns at flow rates from 1 to 9 ml/min, and superior resolution was always obtained using monolithic over conventional columns at the same flow rate. A total of seven monolithic columns from four different batches were used in this study.

[1]  Alexander H Schmidt Fast HPLC for quality control of Harpagophytum procumbens by using a monolithic silica column: method transfer from conventional particle-based silica column. , 2005, Journal of chromatography. A.

[2]  K. Nakanishi,et al.  Effect of skeleton size on the performance of octadecylsilylated continuous porous silica columns in reversed-phase liquid chromatography , 1997 .

[3]  R. Plumb,et al.  Use of monolithic silica columns to increase analytical throughput for metabolite identification by liquid chromatography/tandem mass spectrometry , 2001 .

[4]  Walter A. Korfmacher,et al.  Simultaneous determination of a drug candidate and its metabolite in rat plasma samples using ultrafast monolithic column high-performance liquid chromatography/tandem mass spectrometry. , 2002, Rapid communications in mass spectrometry : RCM.

[5]  T. Higuchi,et al.  Quantitative analysis of microgram quantities of pilocarpine in aqueous solution. , 1971, Journal of pharmaceutical sciences.

[6]  D. Massart,et al.  Fast separations on monolithic silica columns: method transfer, robustness and column ageing for some case studies. , 2003, Journal of Pharmaceutical and Biomedical Analysis.

[7]  H. Wätzig,et al.  Wide concentration range investigation of recovery, precision and error structure in liquid chromatography. , 2004, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences.

[8]  F. Regnier,et al.  Potential of silica monolithic columns in peptide separations. , 2004, Journal of chromatography. A.

[9]  V. Samanidou,et al.  The use of a monolithic column to improve the simultaneous determination of four cephalosporin antibiotics in pharmaceuticals and body fluids by HPLC after solid phase extraction--a comparison with a conventional reversed-phase silica-based column. , 2004, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences.

[10]  C. Heidbreder,et al.  Amino acid neurotransmitters: separation approaches and diagnostic value. , 2002, Journal of chromatography. B, Analytical technologies in the biomedical and life sciences.

[11]  Kazuki Nakanishi,et al.  Phase Separation in Gelling Silica–Organic Polymer Solution: Systems Containing Poly(sodium styrenesulfonate) , 1991 .

[12]  R. Kostiainen,et al.  Rapid identification and quantitation of compounds with forensic interest using fast liquid chromatography-ion trap mass spectrometry and library searching. , 2003, Journal of chromatography. A.

[13]  D. Volmer,et al.  Studies on azaspiracid biotoxins. I. Ultrafast high-resolution liquid chromatography/mass spectrometry separations using monolithic columns. , 2002, Rapid communications in mass spectrometry : RCM.

[14]  R. A. Shalliker,et al.  A column capacity study of single, serial, and parallel linked rod monolithic high performance liquid chromatography columns. , 2005, Journal of chromatography. A.

[15]  J. Dempsey,et al.  Practical aspects of fast HPLC separations for pharmaceutical process development using monolithic columns , 2004 .

[16]  P. Carr,et al.  High temperature fast chromatography of proteins using a silica-based stationary phase with greatly enhanced low pH stability. , 2005, Journal of chromatography. A.

[17]  R. Plumb,et al.  Direct analysis of pharmaceutical compounds in human plasma with chromatographic resolution using an alkyl-bonded silica rod column. , 2001, Rapid communications in mass spectrometry : RCM.

[18]  E. Forgács,et al.  Characterization of plant oils on a monolithic silica column by high-performance liquid chromatography-atmospheric pressure chemical lonization-mass spectrometry , 2002 .

[19]  Kazuki Nakanishi,et al.  Simple and comprehensive two-dimensional reversed-phase HPLC using monolithic silica columns. , 2004, Analytical chemistry.

[20]  J. Meriluoto,et al.  Rapid separation of microcystins and nodularin using a monolithic silica C18 column. , 2002, Journal of chromatography. A.

[21]  G. Guiochon,et al.  Repeatability and reproducibility of retention data and band profiles on six batches of monolithic columns. , 2002, Journal of chromatography. A.

[22]  K. Nakanishi,et al.  A New Monolithic‐Type HPLC Column For Fast Separations , 2000 .

[23]  L. Goodman,et al.  The Pharmacological Basis of Therapeutics , 1941 .

[24]  K. D. Sternitzke,et al.  Improved high-performance liquid chromatographic determination of pilocarpine and its degradation products in ophthalmic solutions importance of octadecylsilane column choice , 1996 .