Data from a different mobility particle sizer (DMPS) or an electrical aerosol analyzer (EAA) has been combined with data from an aerodynamic particle sizer (APS) and converted to obtain aerosol mass distribution parameters on a near real-time basis. A low pressure impactor (LPI), a direct and independent measure of this mass distribution, provided information for comparison. The number distribution of particles within the electrical measurement range was obtained with the DMPS and EAA. Data from the APS for particles greater than that size were used to complete the number distribution. Two methods of obtaining mass distribution parameters from this number data were attempted. The first was to convert the number data, channel by channel, to mass data and then fit a log-normal function to this new mass distribution. The second method was to fit a log-normal function to the combined number distribution and then use the Hatch-Choate equations to obtain mass parameters. Both the DMPS / APS and the EAA / APS sy...