In vivo accuracy assessment of a needle-based navigation system for CT-guided radiofrequency ablation of the liver.

Computed tomography (CT)-guided percutaneous radiofrequency ablation (RFA) has become a commonly used procedure in the treatment of liver tumors. One of the main challenges related to the method is the exact placement of the instrument within the lesion. To address this issue, a system was developed for computer-assisted needle placement which uses a set of fiducial needles to compensate for organ motion in real time. The purpose of this study was to assess the accuracy of the system in vivo. Two medical experts with experience in CT-guided interventions and two nonexperts used the navigation system to perform 32 needle insertions into contrasted agar nodules injected into the livers of two ventilated swine. Skin-to-target path planning and real-time needle guidance were based on preinterventional 1 mm CT data slices. The lesions were hit in 97% of all trials with a mean user error of 2.4 +/- 2.1 mm, a mean target registration error (TRE) of 2.1 +/- 1.1 mm, and a mean overall targeting error of 3.7 +/- 2.3 mm. The nonexperts achieved significantly better results than the experts with an overall error of 2.8 +/- 1.4 mm (n=16) compared to 4.5 +/- 2.7 mm (n=16). The mean time for performing four needle insertions based on one preinterventional planning CT was 57 +/- 19 min with a mean setup time of 27 min, which includes the steps fiducial insertion (24 +/- 15 min), planning CT acquisition (1 +/- 0 min), and registration (2 +/- 1 min). The mean time for path planning and targeting was 5 +/- 4 and 2 +/- 1 min, respectively. Apart from the fiducial insertion step, experts and nonexperts performed comparably fast. It is concluded that the system allows for accurate needle placement into hepatic tumors based on one planning CT and could thus enable considerable improvement to the clinical treatment standard for RFA procedures and other CT-guided interventions in the liver. To support clinical application of the method, optimization of individual system modules to reduce intervention time is proposed.

[1]  K. Cleary,et al.  Electromagnetic tracking for abdominal interventions in computer aided surgery , 2006, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[2]  Karl Rohr,et al.  Respiratory motion compensation for CT-guided interventions in the liver , 2008 .

[3]  Ken Masamune,et al.  Image overlay for CT-guided needle insertions , 2005, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[4]  J. Adler,et al.  Robotic Motion Compensation for Respiratory Movement during Radiosurgery , 2000, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[5]  K. Cleary,et al.  Assessment of hepatic motion secondary to respiration for computer assisted interventions. , 2002, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[6]  Hui Zhang,et al.  Two-Stage Registration for Real-Time Deformable Compensation Using an Electromagnetic Tracking Device , 2005, MICCAI.

[7]  Christophe Chefd'Hotel,et al.  Characterization of Internal Organ Motion Using Skin Marker Positions , 2004, MICCAI.

[8]  Marco Nolden,et al.  The Medical Imaging Interaction Toolkit , 2005, Medical Image Anal..

[9]  J. McGahan,et al.  Radiofrequency ablation of the liver: current status. , 2001, AJR. American journal of roentgenology.

[10]  Lena Maier-Hein,et al.  Precision Targeting of Liver Lesions with a Needle-Based Soft Tissue Navigation System , 2007, MICCAI.

[11]  Lena Maier-Hein,et al.  Soft tissue navigation using needle-shaped markers: evaluation of navigation aid tracking accuracy and CT registration , 2007, SPIE Medical Imaging.

[12]  Willi A. Kalender,et al.  A Navigation System for Minimally Invasive CT-Guided Interventions , 2005, MICCAI.

[13]  T. Vogl,et al.  Navigation-Based Needle Puncture of a Cadaver Using a Hybrid Tracking Navigational System , 2006, Investigative radiology.

[14]  M Tsuchida,et al.  CT-guided agar marking for localization of nonpalpable peripheral pulmonary lesions. , 1999, Chest.

[15]  Vladimir Kolmogorov,et al.  An Experimental Comparison of Min-Cut/Max-Flow Algorithms for Energy Minimization in Vision , 2004, IEEE Trans. Pattern Anal. Mach. Intell..

[16]  Willi A. Kalender,et al.  Needle and catheter navigation using electromagnetic tracking for computer-assisted C-arm CT interventions , 2007, SPIE Medical Imaging.

[17]  K. Cleary,et al.  Precision targeting of liver lesions using a novel electromagnetic navigation device in physiologic phantom and swine. , 2005, Medical physics.

[18]  J. Krücker,et al.  Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy. , 2007, Journal of vascular and interventional radiology : JVIR.

[19]  Berthold K. P. Horn,et al.  Closed-form solution of absolute orientation using unit quaternions , 1987 .

[20]  Philippe L. Pereira,et al.  Actual role of radiofrequency ablation of liver metastases , 2007, European Radiology.

[21]  K. Lam,et al.  Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging. , 1998, International journal of radiation oncology, biology, physics.