Safety threshold of R-wave amplitudes in patients with implantable cardioverter defibrillator

Objective A safety threshold for baseline rhythm R-wave amplitudes during follow-up of implantable cardioverter defibrillators (ICD) has not been established. We aimed to analyse the amplitude distribution and undersensing rate during spontaneous episodes of ventricular fibrillation (VF), and define a safety amplitude threshold for baseline R-waves. Methods Data were obtained from an observational multicentre registry conducted at 48 centres in Spain. Baseline R-wave amplitudes and VF events were prospectively registered by remote monitoring. Signal processing algorithms were used to compare amplitudes of baseline R-waves with VF R-waves. All undersensed R-waves after the blanking period (120 ms) were manually marked. Results We studied 2507 patients from August 2011 to September 2014, which yielded 229 VF episodes (cycle length 189.6±29.1 ms) from 83 patients that were suitable for R-wave comparisons (follow-up 2.7±2.6 years). The majority (77.6%) of VF R-waves (n=13953) showed lower amplitudes than the reference baseline R-wave. The decrease in VF amplitude was progressively attenuated among subgroups of baseline R-wave amplitude (≥17; ≥12 to <17; ≥7 to <12; ≥2.2 to <7 mV) from the highest to the lowest: median deviations −51.2% to +22.4%, respectively (p=0.027). There were no significant differences in undersensing rates of VF R-waves among subgroups. Both the normalised histogram distribution and the undersensing risk function obtained from the ≥2.2 to <7 mV subgroup enabled the prediction that baseline R-wave amplitudes ≤2.5 mV (interquartile range: 2.3–2.8 mV) may lead to ≥25% of undersensed VF R-waves. Conclusions Baseline R-wave amplitudes ≤2.5 mV during follow-up of patients with ICDs may lead to high risk of delayed detection of VF. Trial registration number NCT01561144; results.

[1]  J. Healey,et al.  Wilkoff BL, et al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. , 2017, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[2]  Jagmeet P. Singh,et al.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. , 2016, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[3]  Paul A. Belk,et al.  Sinus rhythm R-wave amplitude as a predictor of ventricular fibrillation undersensing in patients with implantable cardioverter-defibrillator. , 2015, Heart rhythm.

[4]  J. Alzueta,et al.  Registro Español de Desfibrilador Automático Implantable. XI Informe Oficial de la Sección de Electrofisiología y Arritmias de la Sociedad Española de Cardiología (2014) , 2015 .

[5]  Renato Pietro Ricci,et al.  HRS Expert Consensus Statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. , 2015, Heart rhythm.

[6]  J. Healey,et al.  Delayed AICD therapy and cardiac arrest resulting from undersensing of ventricular fibrillation in a subject with hypertrophic cardiomyopathy–A case report , 2015, Indian Pacing and Electrophysiology Journal.

[7]  J. Pogue,et al.  Cardioverter defibrillator implantation without induction of ventricular fibrillation: a single-blind, non-inferiority, randomised controlled trial (SIMPLE) , 2015, The Lancet.

[8]  A. Quesada,et al.  Spanish Implantable Cardioverter-defibrillator Registry. Eleventh Official Report of the Spanish Society of Cardiology Electrophysiology and Arrhythmias Section (2014). , 2013, Revista espanola de cardiologia.

[9]  P. Friedman,et al.  Automated Vulnerability Testing Identifies Patients With Inadequate Defibrillation Safety Margin , 2012, Circulation. Arrhythmia and electrophysiology.

[10]  Martin J Schalij,et al.  Trans-venous lead removal without the use of extraction sheaths, results of >250 removal procedures. , 2012, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[11]  Felipe Alonso-Atienza,et al.  Spectral analysis of intracardiac electrograms during induced and spontaneous ventricular fibrillation in humans. , 2009, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[12]  J. Healey,et al.  Complications associated with defibrillation threshold testing: the Canadian experience. , 2008, Heart rhythm.

[13]  N. Lever,et al.  Differences in the characteristics of induced and spontaneous episodes of ventricular fibrillation. , 2007, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[14]  A. Russo,et al.  The Dilemma of ICD Implant Testing , 2007, Pacing and clinical electrophysiology : PACE.

[15]  R. Tukkie,et al.  Undersensing of VF in a Patient With Optimal R Wave Sensing During Sinus Rhythm , 2004, Pacing and clinical electrophysiology : PACE.

[16]  C. Swerdlow Implantation of Cardioverter Defibrillators Without Induction of Ventricular Fibrillation , 2001, Circulation.

[17]  D. Krum,et al.  Ventricular Fibrillation Sensing and Detection by Implantable Defibrillators: Is One Better than the Others? A Prospective, Comparative Study , 2001, Journal of cardiovascular electrophysiology.

[18]  M. Stanton,et al.  Are Routine Arrhythmia Inductions Necessary in Patients with Pectoral Implantable Cardioverter Defibrillators? , 2000, Journal of cardiovascular electrophysiology.

[19]  A. Natale,et al.  Undetected ventricular fibrillation in transvenous implantable cardioverter-defibrillators. Prospective comparison of different lead system-device combinations. , 1996, Circulation.

[20]  B. Wilkoff,et al.  Adequacy of implantable cardioverter-defibrillator lead placement for tachyarrhythmia detection by sinus rhythm electrogram amplitude. , 1995, The American journal of cardiology.

[21]  K. Ellenbogen,et al.  Measurement of ventricular electrogram amplitude during intraoperative induction of ventricular tachyarrhythmias. , 1992, The American journal of cardiology.

[22]  Douglas L. Jones,et al.  Correlation Between the Ventricular Electrogram Amplitude in Sinus Rhythm and in Ventricular Fibrillation , 1990, Pacing and clinical electrophysiology : PACE.