Determination of parts-per-billion concentrations of dioxane in water and soil by purge and trap gas chromatography/mass spectrometry or charcoal tube enrichment gas chromatography

Two methods for the determination of 1,4-dioxane in water have been studied. The first method is a heated purge and trap gas chromatography/mass spectrometry system following salting out with sodium sulfate. The second method is an adsorption on coconut-shell charcoal and solvent desorption with carbon disulfide/methanol followed by analysis of the desorbate by gas chromatography with flame ionization detection. The first method is also successful for the determination of 1,4-dioxane in solids and sediments. The second method is shown to be successful for 2-butanone, 4-methyl-2-pentanone, and butoxyethanol in water. The two methods are compared by analyzing 15 samples by both methods and achieving similar results.

[1]  John G. Dorsey,et al.  Equations for calculation of chromatographic figures of merit for ideal and skewed peaks , 1983 .

[2]  R. Delley Series for the exponentially modified Gaussian peak shape , 1985 .

[3]  R. L. Petty Elution of adsorbed organics from graphitized carbon black , 1981 .

[4]  G. Guiochon,et al.  Determination of the Higher Moments of a Nonsymmetrical Chromatographic Signal , 1976 .

[5]  J. Kirkland,et al.  Sampling and extra-column effects in high-performance liquid chromatography; influence of peak skew on plate count calculations. , 1977, Journal of chromatographic science.

[6]  P. Carr,et al.  Graphical method for obtaining retention time and number of theoretical plates from tailed chromatographic peaks , 1981 .

[7]  B. Karger,et al.  Gradient liquid chromatography/mass spectrometry using microbore columns and a moving belt interface , 1984 .

[8]  B. Bidlingmeyer,et al.  Column efficiency measurement , 1984 .

[9]  S. Chesler,et al.  Effect of peak sensing and random noise on the precision and accuracy of statistical moment analyses from digital chromatographic data , 1971 .

[10]  W. Yau Characterizing skewed chromatographic band broadening , 1977 .

[11]  E. Grushka,et al.  Computer characterization of chromatographic peaks by plate height and higher central moments , 1969 .

[12]  I. Suffet,et al.  Interactive effects of temperature, salt concentration, and pH on head space analysis for isolating volatile trace organics in aqueous environmental samples , 1979 .

[13]  E. Toren,et al.  AXIAL DISPERSION IN COILED TUBULAR REACTORS , 1986 .

[14]  P. Carr,et al.  Errors in Exponentially Modified Gaussian Equations in the Literature , 1985 .

[15]  John G. Dorsey,et al.  A Review of the Exponentially Modified Gaussian (EMG) Function: Evaluation and Subsequent Calculation of Universal Data , 1984 .

[16]  A. Cedergren,et al.  Estimation of oxygen pressure in graphite furnaces for atomic absorption spectrometry , 1984 .

[17]  Joe P. Folley Systematic errors in the measurement of peak area and peak height for overlapping peaks , 1987 .

[18]  R. Samperi,et al.  Evaluation of graphitized carbon black cartridges for rapid organic trace enrichment from water: application to priority pollutant phenols , 1986 .

[19]  J. Funk,et al.  Retention Time and the First Time Moment in Elution Chromatography. III. General Conclusions , 1971 .

[20]  R. R. Walters,et al.  Effect of Baseline Errors on the Calculation of Statistical Moments of Tailed Chromatographic Peaks , 1984 .

[21]  J. Tyson Peak width and reagent dispersion in flow injection analysis , 1986 .

[22]  R. Sturgeon,et al.  Absorption pulse shifting in graphite furnace atomic absorption spectrometry , 1985 .

[23]  R. Delley The peak width of nearly Gaussian peaks , 1984 .