Heart motion-adapted MR velocity mapping of blood velocity distribution downstream of aortic valve prostheses: initial experience.

PURPOSE To investigate blood flow velocities and shear rates at two distances downstream of an artificial aortic valve in patients. MATERIALS AND METHODS Blood velocity was quantified downstream of the valve prosthesis (for replacement after aortic valve stenosis or combined stenosis and regurgitation) in 10 patients by using a magnetic resonance (MR) cine velocity mapping method in which the imaging section position is adapted according to the excursion of the valvular plane of the heart. Two acquisitions were performed to display the blood velocity distributions one-fourth valve diameter and one valve diameter downstream of the valve and to quantify blood volumes and shear rates. RESULTS The velocity profiles measured during flow acceleration one-fourth valve diameter downstream were characterized by a distinct pattern of two lateral jets and one central jet of antegrade flow. High shear rates were found along the leaflet tips. The profiles obtained one valve diameter downstream were skewed, with varying velocity patterns among patients. Peak shear rates were found close to the vessel wall. With correction for through-plane motion of the valve, the mean apparent regurgitant fraction (+/- SD) was 14% +/- 6; the mean regurgitant fraction without correction was 9% +/- 5. CONCLUSION The described noninvasive procedure for velocity mapping enables measurements close to the valve and thus evaluation of blood flow patterns with respect to valve design in humans.

[1]  S. Riederer,et al.  Respiratory Motion of the Heart: Kinematics and the Implications for the Spatial Resolution in Coronary Imaging , 1995, Magnetic resonance in medicine.

[2]  M Weiger,et al.  Motion‐adapted gating based on k‐space weighting for reduction of respiratory motion artifacts , 1997, Magnetic resonance in medicine.

[3]  E. M. Pedersen,et al.  Magnetic resonance imaging of blood velocity distribution around St. Jude medical aortic valves in patients. , 1996, The Journal of heart valve disease.

[4]  René M. Botnar,et al.  Automatic vessel segmentation using active contours in cine phase contrast flow measurements , 1999, Journal of magnetic resonance imaging : JMRI.

[5]  M. McConnell,et al.  Prospective navigator correction of image position for coronary MR angiography. , 1997, Radiology.

[6]  A P Yoganathan,et al.  Magnetic resonance velocity imaging: a new method for prosthetic heart valve study. , 1995, The Journal of heart valve disease.

[7]  P K Paulsen,et al.  Quantitation of the turbulent stress distribution downstream of normal, diseased and artificial aortic valves in humans. , 1992, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[8]  D N Firmin,et al.  Valve and great vessel stenosis: assessment with MR jet velocity mapping. , 1991, Radiology.

[9]  P. Walker,et al.  Semiautomated method for noise reduction and background phase error correction in MR phase velocity data , 1993, Journal of magnetic resonance imaging : JMRI.

[10]  H Schmid-Schönbein,et al.  Towards a concept of thrombosis in accelerated flow: rheology, fluid dynamics, and biochemistry. , 1985, Biorheology.

[11]  A P Yoganathan,et al.  Quantification of regurgitant flow through bileaflet heart valve prostheses: theoretical and in vitro studies. , 1993, Ultrasound in medicine & biology.

[12]  P K Paulsen,et al.  Turbulent stresses downstream of three mechanical aortic valve prostheses in human beings. , 1994, The Journal of thoracic and cardiovascular surgery.

[13]  P Boesiger,et al.  Heart motion adapted cine phase‐contrast flow measurements through the aortic valve , 1999, Magnetic resonance in medicine.

[14]  H. Reul,et al.  Estimation of Shear Stress-related Blood Damage in Heart Valve Prostheses - in Vitro Comparison of 25 Aortic Valves , 1990, The International journal of artificial organs.

[15]  N H Hwang,et al.  Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. , 1984, Biorheology.

[16]  N J Pelc,et al.  Concomitant gradient terms in phase contrast MR: Analysis and correction , 1998, Magnetic resonance in medicine.

[17]  Y Wang,et al.  3D MR angiography of pulmonary arteries using realtime navigator gating and magnetization preparation , 1996, Magnetic resonance in medicine.

[18]  P. Walker,et al.  Comparison of magnetic resonance imaging and Laser Doppler Anemometry velocity measurements downstream of replacement heart valves: implications for in vivo assessment of prosthetic valve function. , 1996, The Journal of heart valve disease.

[19]  R. Pettigrew,et al.  Turbulent Fluctuation Velocity: The Most Significant Determinant of Signal Loss in Stenotic Vessels , 1995, Magnetic resonance in medicine.

[20]  M H Buonocore Blood flow measurement using variable velocity encoding in the RR interval , 1993, Magnetic resonance in medicine.

[21]  Woo Yr,et al.  In vitro pulsatile flow velocity and turbulent shear stress measurements in the vicinity of mechanical aortic heart valve prostheses. , 1985 .

[22]  P K Paulsen,et al.  Velocity fields and turbulent stresses downstream of biological and mechanical aortic valve prostheses implanted in pigs. , 1988, Cardiovascular research.

[23]  O Henriksen,et al.  Quantification of complex flow using MR phase imaging--a study of parameters influencing the phase/velocity relation. , 1992, Magnetic resonance imaging.

[24]  H Reul,et al.  Velocity and Shear Stress Distribution Downstream of Mechanical Heart Valves in Pulsatile Flow , 1989, The International journal of artificial organs.