Acoustic properties of lesions generated with an ultrasound therapy system.

Methods for quantitative imaging of ultrasound propagation properties were applied to the examination of the acoustic appearance of lesions generated by high intensity focused ultrasound in excised pig livers. Single lesions, about 10 mm maximum diameter by 30 mm long, were created in each of six liver specimens. Two dimensional images (32 by 32 points) of sound speed, mean attenuation coefficient (as a function of frequency in the range 3 to 8.5 MHz) and mean backscattering coefficient (5 to 8 MHz) were obtained in 7 mm thick sections of tissue, cut to include a cross-section through the lesion. Images of these properties, presented alongside surface photographs of the samples, provided a qualitative demonstration that attenuation coefficient was the most useful and backscattering coefficient was the least useful acoustic parameter for visualizing such lesions. Quantitatively the data demonstrated significant increases in attenuation coefficient and sound speed in lesioned liver relative to normal, whereas backscattering was shown not to change in a significant manner except when undissolved gas is the mechanism for increased acoustic scattering. Samples where gas was not fully removed following lesion production gave significant increases in backscattering at the lesion centre, but the shape and size of regions of high backscattering coefficient corresponded poorly with the shape and size of the lesions, unlike attenuation and sound speed for which such correspondence was good.

[1]  Narendra T. Sanghvi,et al.  A Focused Ultrasound System for Tissue Volume Ablation in Deep Seated Brain Sites , 1986, IEEE 1986 Ultrasonics Symposium.

[2]  J B Pond,et al.  The role of heat in the production of ultrasonic focal lesions. , 1970, The Journal of the Acoustical Society of America.

[3]  J C Bamber,et al.  Effect of gaseous inclusions on the frequency dependence of ultrasonic attenuation in liver. , 1985, Ultrasound in medicine & biology.

[4]  J. Bamber,et al.  Quantitative Imaging of Acoustical and Histological Properties of Excised Tissues , 1991 .

[5]  H. Schwan,et al.  Mechanism of absorption of ultrasound in liver tissue. , 1971, The Journal of the Acoustical Society of America.

[6]  C R Hill,et al.  Ultrasonic attenuation and propagation speed in mammalian tissues as a function of temperature. , 1979, Ultrasound in medicine & biology.

[7]  P. P. Lele,et al.  An analysis of lesion development in the brain and in plastics by high-intensity focused ultrasound at low-megahertz frequencies. , 1972, The Journal of the Acoustical Society of America.

[8]  R. C. Eggleton,et al.  Threshold ultrasonic dosages for structural changes in the mammalian brain. , 1970, The Journal of the Acoustical Society of America.

[9]  C. R. Hill,et al.  Acoustic properties of normal and cancerous human liver-I. Dependence on pathological condition. , 1981, Ultrasound in medicine & biology.

[10]  Rivens Ih Quantitative studies of biological damage induced using high intensity focused ultrasound. , 1992 .

[11]  D. R. Bacon,et al.  Characteristics of a PVDF Membrane Hydrophone for Use in the Range 1-100 MHz , 1982, IEEE Transactions on Sonics and Ultrasonics.

[12]  G. Haar,et al.  High intensity focused ultrasound--a surgical technique for the treatment of discrete liver tumours. , 1989, Physics in medicine and biology.

[13]  C R Hill,et al.  Lesion development in focused ultrasound surgery: a general model. , 1994, Ultrasound in medicine & biology.

[14]  P. P. Lele,et al.  Production of deep focal lesions by focused ultrasound--current status. , 1967, Ultrasonics.