A new silicone hollow fiber membrane oxygenator for extracorporeal membrane oxygenation (ECMO) was developed using an ultrathin silicone hollow fiber, with a 300 microm outer diameter and a wall thickness of 50 microm. The hollow fibers were mechanically cross-wound on the flow distributor to achieve equal distribution of blood flow without changing the fiber shape. The housing, made of silicone coated acryl, was 236 mm long with an inner diameter of 60 mm. The surface area was 1.0 m2 for prototype 211, and 1.1 m2 for prototype 209. The silicone fiber length was 150 mm, and the silicone membrane packing density was 43% for prototype 211 and 36% for prototype 209. Prototype 211 has a priming volume of 208 ml, and prototype 209 has a priming volume of 228 ml. The prototype 211 oxygenator demonstrates a gas transfer rate of 120 +/- 5 ml/min (mean +/- SD) for O2 and 67 +/- 12 ml/min for CO2 under 2 L of blood flow and 4 L of O2 gas flow. Prototype 209 produced the same values. The blood side pressure drop was low compared with the silicone sheet oxygenator (Avecor, 1500ECMO). These results showed that this new oxygenator for ECMO had efficiency similar to the silicone sheet oxygenator that has a 50% larger surface area. These results suggest that the new generation oxygenator using an ultrathin silicone hollow fiber possesses sufficient gas transfer performance for long-term extracorporeal lung support.