Levamisole: a Common Adulterant in Cocaine Street Samples Hindering Electrochemical Detection of Cocaine.

The present work investigates the electrochemical determination of cocaine in the presence of levamisole, one of the most common adulterants found in cocaine street samples. Levamisole misleads cocaine color tests, giving a blue color (positive test) even in the absence of cocaine. Moreover, the electrochemical detection of cocaine is also affected by the presence of levamisole, with a suppression of the oxidation signal of cocaine. When levamisole is present in the sample in ratios higher than 1:1, the cocaine signal is no longer detected, thus leading to false negative results. Mass spectrometry and nuclear magnetic resonance were used to investigate if the signal suppression is due to the formation of a complex between cocaine and levamisole in bulk solution. Strategies to eliminate this suppressing effect are further suggested in this manuscript. In a first approach, the increase of the pH of the sample solution from pH 7 to pH 12 allowed the voltammetric determination of cocaine in the presence of levamisole in a concentration range from 10 to 5000 μM at nonmodified graphite disposable electrodes with a detection limit of 5 μM. In a second approach, the graphite electrode was cathodically pretreated, resulting in the presence of oxidation peaks of both cocaine and levamisole, with a detection limit for cocaine of 3 μM over the linear range of concentrations from 10 to 2500 μM. Both these strategies have been successfully applied for the simultaneous detection of cocaine and levamisole in three street samples on unmodified graphite disposable electrodes.