Biomechanical comparison of conventional and anatomical calcaneal plates for the treatment of intraarticular calcaneal fractures – a finite element study

Abstract Initial stability is essential for open reduction internal fixation of intraarticular calcaneal fractures. Geometrical feature of a calcaneal plate is influential to its endurance under physiological load. It is unclear if conventional and pre-contoured anatomical calcaneal plates may exhibit differently in biomechanical perspective. A Sanders’ Type II-B intraarticular calcaneal fracture model was reconstructed to evaluate the effectiveness of calcaneal plates using finite element methods. Incremental vertical joint loads up to 450 N were exerted on the subtalar joint to evaluate the stability and safety of the calcaneal plates and bony structure. Results revealed that the anatomical calcaneal plate model had greater average structural stiffness (585.7 N/mm) and lower von Mises stress on the plate (774.5 MPa) compared to those observed in the conventional calcaneal plate model (stiffness: 430.9 N/mm; stress on plate: 867.1 MPa). Although both maximal compressive and maximal tensile stress and strain were lower in the anatomical calcaneal plate group, greater loads on fixation screws were found (average 172.7 MPa compared to 82.18 MPa in the conventional calcaneal plate). It was noted that high magnitude stress concentrations would occur where the bone plate bridges the fracture line on the lateral side of the calcaneus bone. Sufficient fixation strength at the posterolateral calcaneus bone is important for maintaining subtalar joint load after reduction and fixation of a Sanders’ Type II-B calcaneal fracture. In addition, geometrical design of a calcaneal plate should worth considering for the mechanical safety in practical usage.

[1]  X. Weng,et al.  Primary Stability of Absorbable Screw Fixation for Intra-articular Calcaneal Fractures: A Finite Element Analysis , 2015 .

[2]  Xiao Yu,et al.  The sustentaculum tali screw fixation for the treatment of Sanders type II calcaneal fracture: A finite element analysis , 1969, Pakistan journal of medical sciences.

[3]  Zhan-po Wu,et al.  A biomechanical comparison of calcaneal anatomical plate and compression bolts versus conventional AO plate in fixation of Sanders Type III calcaneus fractures , 2012 .

[4]  Shih-Liang Shih,et al.  Biomechanical analysis of foot with different foot arch heights: a finite element analysis , 2012, Computer methods in biomechanics and biomedical engineering.

[5]  R. Adelaar,et al.  Biomechanical Evaluation of a Locking and Nonlocking Reconstruction Plate in an Osteoporotic Calcaneal Fracture Model , 2011, Foot & ankle international.

[6]  H. Zwipp,et al.  Stability of Locking and Non-Locking Plates in an Osteoporotic Calcaneal Fracture Model , 2011, Foot & ankle international.

[7]  Shih-Liang Shih,et al.  Using an Optimization Approach to Design an Insole for Lowering Plantar Fascia Stress—A Finite Element Study , 2008, Annals of Biomedical Engineering.

[8]  Chun-Li Lin,et al.  Finite element analysis of plantar fascia under stretch-the relative contribution of windlass mechanism and Achilles tendon force. , 2008, Journal of biomechanics.

[9]  M. Richter,et al.  Polyaxially-locked plate screws increase stability of fracture fixation in an experimental model of calcaneal fracture. , 2006, The Journal of bone and joint surgery. British volume.

[10]  John T. Campbell,et al.  A Biomechanical Comparison of Locking and Nonlocking Plates for the Fixation of Calcaneal Fractures , 2006, Foot & ankle international.

[11]  R. Buckley,et al.  Complications Following Management of Displaced Intra-Articular Calcaneal Fractures: A Prospective Randomized Trial Comparing Open Reduction Internal Fixation With Nonoperative Management , 2003, Journal of orthopaedic trauma.

[12]  R. Buckley,et al.  Operative Compared with Nonoperative Treatment of Displaced Intra-Articular Calcaneal Fractures: A Prospective, Randomized, Controlled Multicenter Trial , 2002, The Journal of bone and joint surgery. American volume.

[13]  Amit Gefen,et al.  Stress analysis of the standing foot following surgical plantar fascia release. , 2002, Journal of biomechanics.

[14]  S. G. Agnew Calcaneal fractures: operative versus nonoperative treatment. , 1997, The Journal of trauma.

[15]  D. Thordarson,et al.  Operative vs. Nonoperative Treatment of Intra-Articular Fractures of the Calcaneus: A Prospective Randomized Trial , 1996, Foot & ankle international.

[16]  B. Sangeorzan,et al.  Extensive intraarticular fractures of the foot. Surgical management of calcaneal fractures. , 1993, Clinical orthopaedics and related research.

[17]  R. Sanders,et al.  Operative Treatment in 120 Displaced Intraarticular Calcaneal Fractures Results Using a Prognostic Computed Tomography Scan Classification , 1993, Clinical orthopaedics and related research.

[18]  R. Atkins,et al.  Intra-articular fractures of the calcaneum. Part II: Open reduction and internal fixation by the extended lateral transcalcaneal approach. , 1993, The Journal of bone and joint surgery. British volume.

[19]  B. Sangeorzan,et al.  Extensive intraarticular fractures of the foot , 1993 .

[20]  R. Buckley,et al.  Comparison of open versus closed reduction of intraarticular calcaneal fractures: a matched cohort in workmen. , 1992 .

[21]  R. Buckley,et al.  Comparison of Open Versus Closed Reduction of Intraarticular Calcaneal Fractures: A Matched Cohort in Workmen , 1992, Journal of orthopaedic trauma.

[22]  P. Essex-Lopresti The mechanism, reduction technique, and results in fractures of the OS calcis , 1952, The British journal of surgery.