Electromyogram (EMG) Removal by Adding Sources of EMG (ERASE) -- A novel ICA-based algorithm for removing myoelectric artifacts from EEG -- Part 1

Electroencephalographic (EEG) recordings are often contaminated by electromyographic (EMG) artifacts, especially when recording during movement. Existing methods to remove EMG artifacts include independent component analysis (ICA), and other high-order statistical methods. However, these methods can not effectively remove most of EMG artifacts. Here, we proposed a modified ICA model for EMG artifacts removal in the EEG, which is called EMG Removal by Adding Sources of EMG (ERASE). In this new approach, additional channels of real EMG from neck and head muscles (reference artifacts) were added as inputs to ICA in order to "force" the most power from EMG artifacts into a few independent components (ICs). The ICs containing EMG artifacts (the "artifact ICs") were identified and rejected using an automated procedure. Simulation results showed ERASE removed EMG artifacts from EEG significantly more effectively than conventional ICA. Subsequently, EEG was collected from 8 healthy participants while they moved their hands to test the realistic efficacy of this approach. Results showed that ERASE successfully removed EMG artifacts (on average, about 75% of EMG artifacts were removed when using real EMGs as reference artifacts) while preserving the expected EEG features related to movement. We also tested the ERASE procedure using simulated EMGs as reference artifacts (about 63% of EMG artifacts removed). Compared to conventional ICA, ERASE removed on average 26% more EMG artifacts from EEG. These results indicate that using additional real or simulated EMG sources can increase the effectiveness of ICA in removing EMG artifacts from EEG. Combined with automated artifact IC rejection, ERASE also minimizes potential user bias.

[1]  R. Gouripeddi,et al.  Computer Methods and Programs in Biomedicine , 2022 .

[2]  Matías Zañartu,et al.  Improving EEG Muscle Artifact Removal With an EMG Array , 2020, IEEE Transactions on Instrumentation and Measurement.

[3]  Fang Xu,et al.  A cautionary tale of entropic criteria in assessing the validity of the maximum entropy principle , 2019, EPL (Europhysics Letters).

[4]  An H. Do,et al.  Hemicraniectomy in Traumatic Brain Injury: A Noninvasive Platform to Investigate High Gamma Activity for Brain Machine Interfaces , 2019, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[5]  W. David Hairston,et al.  Adding neck muscle activity to a head phantom device to validate mobile EEG muscle and motion artifact removal * , 2019, 2019 9th International IEEE/EMBS Conference on Neural Engineering (NER).

[6]  Yongcheng Li,et al.  A novel algorithm for removing artifacts from EEG data , 2018, 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[7]  Wei Wu,et al.  ARTIST: A fully automated artifact rejection algorithm for single‐pulse TMS‐EEG data , 2018, Human brain mapping.

[8]  Xueyuan Xu,et al.  The Use of Multivariate EMD and CCA for Denoising Muscle Artifacts From Few-Channel EEG Recordings , 2018, IEEE Transactions on Instrumentation and Measurement.

[9]  Rabab K. Ward,et al.  Removing Muscle Artifacts From EEG Data: Multichannel or Single-Channel Techniques? , 2016, IEEE Sensors Journal.

[10]  S. Edward Rajan,et al.  A novel approach for the elimination of artefacts from EEG signals employing an improved Artificial Immune System algorithm , 2016, J. Exp. Theor. Artif. Intell..

[11]  Raveendran Paramesran,et al.  Artifacts-matched blind source separation and wavelet transform for multichannel EEG denoising , 2015, Biomed. Signal Process. Control..

[12]  Begoña Garcia-Zapirain,et al.  EEG artifact removal—state-of-the-art and guidelines , 2015, Journal of neural engineering.

[13]  Jing Hu,et al.  Removal of EOG and EMG artifacts from EEG using combination of functional link neural network and adaptive neural fuzzy inference system , 2015, Neurocomputing.

[14]  Rabab Kreidieh Ward,et al.  A Preliminary Study of Muscular Artifact Cancellation in Single-Channel EEG , 2014, Sensors.

[15]  Xun Chen,et al.  Removal of Muscle Artifacts from Single-Channel EEG Based on Ensemble Empirical Mode Decomposition and Multiset Canonical Correlation Analysis , 2014, J. Appl. Math..

[16]  Aiguo Song,et al.  EOG Artifact Correction from EEG Recording Using Stationary Subspace Analysis and Empirical Mode Decomposition , 2013, Sensors.

[17]  Rami K. Niazy,et al.  Automatic correction of eye blink artifact in single channel EEG recording using EMD and OMP , 2013, 21st European Signal Processing Conference (EUSIPCO 2013).

[18]  Kevin Warwick,et al.  Automated Artifact Removal From the Electroencephalogram , 2013, Clinical EEG and neuroscience.

[19]  S. Muthukumaraswamy High-frequency brain activity and muscle artifacts in MEG/EEG: a review and recommendations , 2013, Front. Hum. Neurosci..

[20]  Guoping Gao,et al.  Removal of Ocular Artifacts in EEG—An Improved Approach Combining DWT and ANC for Portable Applications , 2013, IEEE Journal of Biomedical and Health Informatics.

[21]  Seán F. McLoone,et al.  The Use of Ensemble Empirical Mode Decomposition With Canonical Correlation Analysis as a Novel Artifact Removal Technique , 2013, IEEE Transactions on Biomedical Engineering.

[22]  Lotfi Senhadji,et al.  ICA-based EEG denoising: a comparative analysis of fifteen methods , 2012 .

[23]  Vera Kaiser,et al.  What does clean EEG look like? , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[24]  Jiang Li,et al.  EOG artifact removal using a wavelet neural network , 2012, Neurocomputing.

[25]  Lotfi Senhadji,et al.  Removal of muscle artifact from EEG data: comparison between stochastic (ICA and CCA) and deterministic (EMD and wavelet-based) approaches , 2012, EURASIP J. Adv. Signal Process..

[26]  Tomás Ward,et al.  Artifact Removal in Physiological Signals—Practices and Possibilities , 2012, IEEE Transactions on Information Technology in Biomedicine.

[27]  F. La Foresta,et al.  Automatic Artifact Rejection From Multichannel Scalp EEG by Wavelet ICA , 2012, IEEE Sensors Journal.

[28]  R. Oostenveld,et al.  Independent EEG Sources Are Dipolar , 2012, PloS one.

[29]  Christopher J. James,et al.  Employing spatially constrained ICA and wavelet denoising, for automatic removal of artifacts from multichannel EEG data , 2012, Signal Process..

[30]  Roberto Hornero,et al.  Quantitative Evaluation of Artifact Removal in Real Magnetoencephalogram Signals with Blind Source Separation , 2011, Annals of Biomedical Engineering.

[31]  A. Mognon,et al.  ADJUST: An automatic EEG artifact detector based on the joint use of spatial and temporal features. , 2011, Psychophysiology.

[32]  Trieu H. Pham,et al.  A test of four EOG correction methods using an improved validation technique. , 2011, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[33]  S. Olbrich,et al.  ICA-based muscle artefact correction of EEG data: What is muscle and what is brain? Comment on McMenamin et al. , 2011, NeuroImage.

[34]  R. B. Reilly,et al.  FASTER: Fully Automated Statistical Thresholding for EEG artifact Rejection , 2010, Journal of Neuroscience Methods.

[35]  Sabine Van Huffel,et al.  Source Separation From Single-Channel Recordings by Combining Empirical-Mode Decomposition and Independent Component Analysis , 2010, IEEE Transactions on Biomedical Engineering.

[36]  Akaysha Tang,et al.  Applications of Second Order Blind Identification to High-Density EEG-Based Brain Imaging: A Review , 2010, ISNN.

[37]  Sabine Van Huffel,et al.  Removal of Muscle Artifacts from EEG Recordings of Spoken Language Production , 2010, Neuroinformatics.

[38]  Daniel P. Ferris,et al.  Removal of movement artifact from high-density EEG recorded during walking and running. , 2010, Journal of neurophysiology.

[39]  Jeffrey S. Maxwell,et al.  Validation of ICA-based myogenic artifact correction for scalp and source-localized EEG , 2010, NeuroImage.

[40]  K. Jellinger,et al.  Practical Guide for Clinical Neurophysiologic Testing: EEG , 2009 .

[41]  Carlos Guerrero-Mosquera,et al.  Automatic removal of ocular artifacts from EEG data using adaptive filtering and Independent Component Analysis , 2009, 2009 17th European Signal Processing Conference.

[42]  Jeffrey S. Maxwell,et al.  Validation of regression-based myogenic correction techniques for scalp and source-localized EEG. , 2009, Psychophysiology.

[43]  Jeffrey S. Maxwell,et al.  Electromyogenic Artifacts and Electroencephalographic Inferences , 2009, Brain Topography.

[44]  Miguel Angel Mañanas,et al.  A comparative study of automatic techniques for ocular artifact reduction in spontaneous EEG signals based on clinical target variables: A simulation case , 2008, Comput. Biol. Medicine.

[45]  Mercedes Atienza,et al.  Muscle Artifact Removal from Human Sleep EEG by Using Independent Component Analysis , 2008, Annals of Biomedical Engineering.

[46]  Frédéric Grouiller,et al.  A comparative study of different artefact removal algorithms for EEG signals acquired during functional MRI , 2007, NeuroImage.

[47]  Slawomir J. Nasuto,et al.  Automatic Artefact Removal from Event-related Potentials via Clustering , 2007, J. VLSI Signal Process..

[48]  Domenico Prattichizzo,et al.  Off-line removal of TMS-induced artifacts on human electroencephalography by Kalman filter , 2007, Journal of Neuroscience Methods.

[49]  Vince D. Calhoun,et al.  A fast algorithm for one-unit ICA-R , 2007, Inf. Sci..

[50]  R. Ward,et al.  EMG and EOG artifacts in brain computer interface systems: A survey , 2007, Clinical Neurophysiology.

[51]  Terrence J. Sejnowski,et al.  Enhanced detection of artifacts in EEG data using higher-order statistics and independent component analysis , 2007, NeuroImage.

[52]  V. Krishnaveni,et al.  Removal of ocular artifacts from EEG using adaptive thresholding of wavelet coefficients , 2006, Journal of neural engineering.

[53]  Wim Van Paesschen,et al.  Canonical Correlation Analysis Applied to Remove Muscle Artifacts From the Electroencephalogram , 2006, IEEE Transactions on Biomedical Engineering.

[54]  Christopher J. James,et al.  On Semi-Blind Source Separation Using Spatial Constraints With Applications in EEG Analysis , 2006, IEEE Transactions on Biomedical Engineering.

[55]  Rémi Gribonval,et al.  A survey of Sparse Component Analysis for blind source separation: principles, perspectives, and new challenges , 2006, ESANN.

[56]  Christopher J James,et al.  Independent component analysis for biomedical signals , 2005, Physiological measurement.

[57]  Y. Tran,et al.  Using independent component analysis to remove artifact from electroencephalographic measured during stuttered speech , 2004, Medical and Biological Engineering and Computing.

[58]  R. Kass,et al.  Automatic correction of ocular artifacts in the EEG: a comparison of regression-based and component-based methods , 2004 .

[59]  L. Euginia,et al.  A wavelet based de-noising technique for ocular artifact correction of the electroencephalogram , 2004, Proceedings. The 8th Russian-Korean International Symposium on Science and Technology, 2004. KORUS 2004..

[60]  R. Kass,et al.  Automatic correction of ocular artifacts in the EEG: a comparison of regression-based and component-based methods. , 2004, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[61]  Arnaud Delorme,et al.  EEGLAB: an open source toolbox for analysis of single-trial EEG dynamics including independent component analysis , 2004, Journal of Neuroscience Methods.

[62]  Christopher J. James,et al.  Temporally constrained ICA: an application to artifact rejection in electromagnetic brain signal analysis , 2003, IEEE Transactions on Biomedical Engineering.

[63]  Patrick Berg,et al.  Artifact Correction of the Ongoing EEG Using Spatial Filters Based on Artifact and Brain Signal Topographies , 2002, Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society.

[64]  M. Scherg,et al.  Spatially constrained independent component analysis for artifact correction in EEG and MEG , 2001, NeuroImage.

[65]  Erkki Oja,et al.  Independent component analysis: algorithms and applications , 2000, Neural Networks.

[66]  Erkki Oja,et al.  Independent component approach to the analysis of EEG and MEG recordings , 2000, IEEE Transactions on Biomedical Engineering.

[67]  Jacques Duchêne,et al.  A model of EMG generation , 2000, IEEE Transactions on Biomedical Engineering.

[68]  Mark A. Girolami,et al.  An Alternative Perspective on Adaptive Independent Component Analysis Algorithms , 1998, Neural Computation.

[69]  Michael Unser,et al.  A review of wavelets in biomedical applications , 1996, Proc. IEEE.

[70]  I. Johnstone,et al.  Adapting to Unknown Smoothness via Wavelet Shrinkage , 1995 .

[71]  Pierre Comon,et al.  Independent component analysis, A new concept? , 1994, Signal Process..

[72]  E Donchin,et al.  A new method for off-line removal of ocular artifact. , 1983, Electroencephalography and clinical neurophysiology.

[73]  A. Hodgkin,et al.  A quantitative description of membrane current and its application to conduction and excitation in nerve , 1952, The Journal of physiology.

[74]  Francisco J. Pelayo,et al.  Trends in EEG-BCI for daily-life: Requirements for artifact removal , 2017, Biomed. Signal Process. Control..

[75]  Christiaan Burger,et al.  Removal of EOG artefacts by combining wavelet neural network and independent component analysis , 2015, Biomed. Signal Process. Control..

[76]  S. Debener,et al.  Towards a truly mobile auditory brain-computer interface: exploring the P300 to take away. , 2014, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[77]  Pham Trieu,et al.  Empirically Validating Fully Automated EOG Artifact Correction using Independent Components Analysis , 2012 .

[78]  Norden E. Huang,et al.  Ensemble Empirical Mode Decomposition: a Noise-Assisted Data Analysis Method , 2009, Adv. Data Sci. Adapt. Anal..

[79]  S. Romero,et al.  Ocular Reduction in EEG Signals Based on Adaptive Filtering, Regression and Blind Source Separation , 2008, Annals of Biomedical Engineering.

[80]  Wei Lu,et al.  Approach and applications of constrained ICA , 2005, IEEE Transactions on Neural Networks.

[81]  Terrence J. Sejnowski,et al.  AUTOMATIC ARTIFACT REJECTION FOR EEG DATA USING HIGH-ORDER STATISTICS AND INDEPENDENT COMPONENT ANALYSIS , 2001 .

[82]  A. J. Bell,et al.  INDEPENDENT COMPONENT ANALYSIS OF BIOMEDICAL SIGNALS , 2000 .

[83]  D F Stegeman,et al.  Muscle fiber action potential changes and surface EMG: A simulation study. , 1992, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[84]  P Berg,et al.  Dipole modelling of eye activity and its application to the removal of eye artefacts from the EEG and MEG. , 1991, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[85]  J. Schoffelen,et al.  Good Practice for Conducting and Reporting Meg Research , 2022 .