Effectiveness and relevance of MR acceptance testing: results of an 8 year audit.

The effectiveness and relevance of independent acceptance testing was assessed by means of an audit of acceptance procedures for 17 MRI systems, with field strengths in the range 0.5-1.5 T, acquired over 8 years. Signal-to-noise ratio and geometric linearity were found to be the image quality parameters most likely to fall below acceptable or expected standards. These received confirmed successful corrective action in 69% of instances. Non-uniformity, ghosting and poor fat suppression were the next most common non-compliant parameters, but yielded less satisfactory outcomes. Spatial resolution was not found to be a sensitive parameter in determining acceptability. 49% of all non-compliant parameters received verifiable corrective attention. A schedule of actual acceptance criteria is presented and shown to be reasonable. Parameter failure rates were shown not to have improved with time. A safety audit of 11 of the installations revealed the most common failings to be inadequate suite layout and poor use of signs. The mean number of safety issues per installation identified as requiring attention was 5, from a questionnaire of 100 points. A number of anecdotal errors and omissions are reported. The data support the importance of an appropriate acceptance procedure for new clinical MRI equipment and for the involvement of a suitably qualified safety adviser on the project team from the outset.

[1]  D W McRobbie Short communication: The absolute signal-to-noise ratio in MRI acceptance testing. , 1996, The British journal of radiology.

[2]  D. McRobbie Radio frequency eddy current losses for an annular conductor in MRI: theory and applications. , 1993, Medical physics.

[3]  Richard A. Lerski Quality control in magnetic resonance imaging , 1998 .

[4]  Y. Zhu,et al.  Refinements of the finite-size pencil beam model of three-dimensional photon dose calculation. , 1997, Medical physics.

[5]  P. Judy,et al.  The line spread function and modulation transfer function of a computed tomographic scanner. , 1976, Medical physics.

[6]  McRobbie Dw Quantitative assessment of magnetic resonance imaging systems. , 1992 .

[7]  R M Harrison,et al.  Quality assurance for MRI: practical experience. , 2000, The British journal of radiology.

[8]  P. M. Walker,et al.  V. Multi-center trial with protocols and prototype test objects for the assessment of MRI equipment , 1988 .

[9]  D W McRobbie A three-dimensional volumetric test object for geometry evaluation in magnetic resonance imaging. , 1997, Medical physics.

[10]  T W Redpath,et al.  Estimating achievable signal-to-noise ratios of MRI transmit-receive coils from radiofrequency power measurements: applications in quality control. , 2000, Physics in medicine and biology.

[11]  L. Axel,et al.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. , 1990, Medical physics.

[12]  S K Mun,et al.  Acceptance testing of magnetic resonance imaging systems: report of AAPM Nuclear Magnetic Resonance Task Group No. 6. , 1992, Medical physics.