Novel methods for 3D postoperative analysis of total knee arthroplasty using 2D-3D image registration.

BACKGROUND Radiographic evaluation of patients after total knee arthroplasty is an important tool for assessing the correct position of the implants and the axis of limb alignment because this will determine long-term durability of the implants. Recently, 2D-3D medical image registration methods are developed for 3D postoperative analysis of total knee arthroplasty. However, most of these techniques have focused only on knee implants. METHODS A 2D-3D medical image registration is implemented to compute the 3D positions of not only implants but also lower limb bones. The following 3D postoperative analysis methods for total knee arthroplasty are presented in this paper: (1) automatic calculation of relative angles of implants and bones, (2) assessment of external rotation angles of inserted implants, and (3) measurement and comparison of both flexion-extension gap balances. Finally these methods have been applied in five patients who underwent total knee replacements. FINDINGS A practical method that can evaluate the patient's knee conditions has been successfully developed. The repeatability and accuracy of 2D-3D registration were around 0.2mm as obtained from the tests using model bones. Based on the 3D information, the novel methods of postoperative analysis were proposed and successfully applied to the patients. INTERPRETATION The 3D positions for both knee implants and lower limb bones can be calculated in order to perform comprehensive postoperative analyses of total knee arthroplasty. The proposed analyses of the postoperative evaluations facilitated various 3D evaluations of the status of implants, alignment of lower limb and gap balances which were not previously feasible.

[1]  B. Beynnon,et al.  The Transepicondylar Axis Approximates the Optimal Flexion Axis of the Knee , 1998, Clinical orthopaedics and related research.

[2]  F. C. Ewald The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. , 1989, Clinical orthopaedics and related research.

[3]  D. S. Pandher,et al.  Computer-assisted navigation increases precision of component placement in total knee arthroplasty. , 2007, Clinical orthopaedics and related research.

[4]  G. Scuderi**,et al.  Rotational positioning of the femoral component in total knee arthroplasty. , 2000, The American journal of knee surgery.

[5]  Graves Ej,et al.  National Hospital Discharge Survey , 2004 .

[6]  John J. Callaghan,et al.  The adult knee , 2003 .

[7]  M. Ritter,et al.  Postoperative alignment of total knee replacement. Its effect on survival. , 1994, Clinical orthopaedics and related research.

[8]  J. Benjamin,et al.  Component alignment in total knee arthroplasty. , 2006, Instructional course lectures.

[9]  E Y Chao,et al.  Normal axial alignment of the lower extremity and load-bearing distribution at the knee. , 1990, Clinical orthopaedics and related research.

[10]  Torsten Rohlfing,et al.  Intensity-based registration algorithm for probabilistic images and its application for 2D to 3D image registration , 2002, SPIE Medical Imaging.

[11]  Leo Joskowicz,et al.  Anatomical image-based rigid registration between fluoroscopic X-ray and CT: methods comparison and experimental results , 2003, CARS.

[12]  J. J. Jacobs,et al.  Malrotation Causing Patellofemoral Complications After Total Knee Arthroplasty , 1998, Clinical orthopaedics and related research.

[13]  Stefan M. Gabriel,et al.  Three-dimensional determination of femoral-tibial contact positions under in vivo conditions using fluoroscopy. , 1998, Clinical biomechanics.

[14]  David J. Hawkes,et al.  A Comparison of 2D-3D Intensity-Based Registration and Feature-Based Registration for Neurointerventions , 2002, MICCAI.

[15]  J D Reuben,et al.  The effect of femoral component position on patellar tracking after total knee arthroplasty. , 1990, Clinical orthopaedics and related research.

[16]  Georg Matziolis,et al.  A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. , 2007, The Journal of bone and joint surgery. American volume.

[17]  Richard D. Komistek,et al.  A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images , 2003, IEEE Transactions on Medical Imaging.

[18]  Johan Bellemans,et al.  Making sense of knee arthroplasty kinematics , 2003 .

[19]  L. Whiteside,et al.  The effects of axial rotational alignment of the femoral component on knee stability and patellar tracking in total knee arthroplasty demonstrated on autopsy specimens. , 1993, Clinical orthopaedics and related research.

[20]  C. Petchprapa,et al.  Imaging of total knee arthroplasty. , 2006, Seminars in musculoskeletal radiology.

[21]  Anthony J. Petrella,et al.  Optimizing Femoral Component Rotation in Total Knee Arthroplasty , 2001, Clinical orthopaedics and related research.

[22]  Guoan Li,et al.  Investigation of in vivo 6DOF total knee arthoplasty kinematics using a dual orthogonal fluoroscopic system , 2006, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[23]  Ramin Shahidi,et al.  Evaluation of Intensity-Based 2D-3D Spine Image Registration Using Clinical Gold-Standard Data , 2003, WBIR.

[24]  Tomaz Slivnik,et al.  3-D/2-D registration of CT and MR to X-ray images , 2003, IEEE Transactions on Medical Imaging.

[25]  C. D. Gelatt,et al.  Optimization by Simulated Annealing , 1983, Science.

[26]  R A Berger,et al.  Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. , 1993, Clinical orthopaedics and related research.

[27]  Insall Jn,et al.  Rotational positioning of the femoral component in total knee arthroplasty. , 2000 .

[28]  Johan H. C. Reiber,et al.  The use of Roentgen stereophotogrammetry to study micromotion of orthopaedic implants , 2002 .