Optimization of slice sensitivity profile for radiographic tomosynthesis.

Similar to other tomographic imaging modalities, the slice sensitivity profile (SSP) is an important image quality metric for radiographic tomosynthesis. In this study, the relationship between the acquisition angular range (θ) and the SSP for the linear trajectory system was carefully investigated from both theoretical and experimental perspectives. A mathematical SSP model was derived for arbitrary points in the reconstructed volume. We used a newly developed flat-panel tomosynthesis prototype system to experimentally validate the mathematical model from 20°(±10°) to 60°(±30°) angular ranges. The SSP was measured by imaging an edge phantom placed at an angle with respect to the detector plane using the modulation transfer function degradation (MTF-d) method. In addition to the experiments, computer simulations were performed to investigate the relationship in a wider angular range (2.5° to 60°). Furthermore, image data from an anthropomorphic phantom were collected to corroborate the system analysis. All the images in this study were constructed using a 3D view-weighted cone-beam filtered backprojection algorithm (3D VW CB-FBP). The theoretical analysis reveals that the SSP of linear trajectory tomosynthesis is inversely proportional to tan(θ∕2). This theory was supported by both simulation (χ2=1.415, DF=7, p=0.985) and phantom experiment (r=0.999, p<0.001) and was further confirmed by an analysis of the reconstructed images of an anthropomorphic phantom. The results imply that the benefit of narrower SSP by increasing angular range quickly diminishes once beyond 40°. The advantages of the MTF-d method were also demonstrated.

[1]  James T Dobbins,et al.  Digital x-ray tomosynthesis: current state of the art and clinical potential. , 2003, Physics in medicine and biology.

[2]  D. Kopans,et al.  Digital tomosynthesis in breast imaging. , 1997, Radiology.

[3]  Renuka Uppaluri,et al.  Measurement of slice thickness and in-plane resolution on radiographic tomosynthesis system using modulation transfer function (MTF) , 2006, SPIE Medical Imaging.

[4]  D. Kopans,et al.  Tomographic mammography using a limited number of low-dose cone-beam projection images. , 2003, Medical physics.

[5]  C. Beaulieu,et al.  Circular tomosynthesis: potential in imaging of breast and upper cervical spine--preliminary phantom and in vitro study. , 2003, Radiology.

[6]  Richard L Webber,et al.  Application of TACT to the evaluation of total joint arthroplasty. , 2003, Medical physics.

[7]  E. Samei,et al.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device. , 1998, Medical physics.

[8]  R. J. Johns,et al.  Three dimensional roentgenography. , 1969, The American journal of roentgenology, radium therapy, and nuclear medicine.

[9]  D. Godfrey,et al.  Optimization of the matrix inversion tomosynthesis (MITS) impulse response and modulation transfer function characteristics for chest imaging. , 2006, Medical physics.

[10]  D. G. Grant Tomosynthesis: a three-dimensional radiographic imaging technique. , 1972, IEEE transactions on bio-medical engineering.

[11]  B. G. Ziedses des Plantes,et al.  Eine Neue Methode Zur Differenzierung in der Rontgenographie (Planigraphies) , 1932 .

[12]  Ismail B. Tutar,et al.  Tomosynthesis-based localization of radioactive seeds in prostate brachytherapy. , 2003, Medical physics.

[13]  J Duryea,et al.  Digital tomosynthesis of hand joints for arthritis assessment. , 2003, Medical physics.

[14]  E. Miller,et al.  An infinite number of laminagrams from a finite number of radiographs. , 1971, Radiology.

[15]  Daniel B. Kopans,et al.  Optimal acquisition techniques for digital breast tomosynthesis screening , 2006, SPIE Medical Imaging.

[16]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .