Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT

BackgroundTwo devices for self-measurement of blood pressure at the brachial artery—the Omron M5-I and the Omron 705IT—were evaluated according to the international protocol of the European Society of Hypertension. DesignThe international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 supplementary subjects are included making a total number of 33 subjects on which the final validation is performed. MethodsFor each subject, four blood pressure (BP) measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with three measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 differences were classified into categories (≤5, ≤10, ≤15 mmHg). The number of differences in each category was compared to the number required by the international protocol. An individual analysis was then done to determine for each subject the number of comparisons ≤5 mmHg. At least 22 of the 33 subjects should have two of their three comparisons ≤5 mmHg. ResultsThe two tested devices passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were −0.9±5.8 and −0.8±4.8 mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP) respectively for the Omron M5-I device and −0.2±4.5 and −2.0±4.8 mmHg for the Omron 705IT device. Readings for the two devices differing by less than 5, 10 and 15 mmHg for systolic and diastolic values fulfil the recommendation criteria of the international protocol as well as the individual analysis. ConclusionsThe Omron M-5I and the Omron 705IT devices pass the validation recommendations of the international protocol.

[1]  F. Mee,et al.  Comparative accuracy of six ambulatory devices according to blood pressure levels , 1993, Journal of hypertension.

[2]  A. Coats,et al.  Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. , 1994, Journal of human hypertension.

[3]  R. Asmar,et al.  Validation of the Omron HEM-907 device for blood pressure measurement , 2002, Blood pressure monitoring.

[4]  A. Zanchetti,et al.  Guidelines for the use of self-blood pressure monitoring: a summary report of the First International Consensus Conference. Groupe Evaluation & Measure of the French Society of Hypertension. , 2000, Journal of hypertension.

[5]  J. Staessen,et al.  Reference values for self-recorded blood pressure: a meta-analysis of summary data. , 1998, Archives of internal medicine.

[6]  L. Appel,et al.  Ambulatory Blood Pressure Monitoring and Blood Pressure Self-Measurement in the Diagnosis and Management of Hypertension , 1993, Annals of Internal Medicine.

[7]  F. Mee,et al.  Evaluation of blood pressure measuring devices. , 1993, Clinical and experimental hypertension.

[8]  Modesti,et al.  Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. , 1996, Blood pressure monitoring.

[9]  D. Yeshurun,et al.  Accuracy of the OMRON M4 automatic blood pressure measuring device , 2000, Journal of Human Hypertension.

[10]  E. O’Brien,et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. , 1990, Journal of hypertension.

[11]  Yutaka Imai,et al.  Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults , 2002, Blood pressure monitoring.

[12]  W. White,et al.  Evaluation of the overall efficacy of the Omron office digital blood pressure HEM‐907 monitor in adults , 2001, Blood pressure monitoring.

[13]  R. Córdoba,et al.  Validación de un monitor de automedida de la presión arterial: OMRON-HM 722C , 1997 .

[14]  O. Hanon,et al.  Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. , 1999, Blood pressure monitoring.

[15]  F. Mee,et al.  Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. , 1996, Blood pressure monitoring.

[16]  T. Pickering,et al.  Recommendations for the use of home (self) and ambulatory blood pressure monitoring. American Society of Hypertension Ad Hoc Panel. , 1996, American journal of hypertension.

[17]  T. Perry,et al.  Comparison of the oscillometric blood pressure monitor (BPM‐100 Beta ) with the auscultatory mercury sphygmomanometer , 2001, Blood pressure monitoring.

[18]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[19]  D. Sackett,et al.  IMPROVEMENT OF MEDICATION COMPLIANCE IN UNCONTROLLED HYPERTENSION , 1976, The Lancet.