Influence of Screw Augmentation in Posterior Dynamic and Rigid Stabilization Systems in Osteoporotic Lumbar Vertebrae: A Biomechanical Cadaveric Study

Study Design. Biomechanical cadaveric study. Objective. To determine whether augmentation positively influence screw stability or not. Summary of Background Data. Implantation of pedicle screws is a common procedure in spine surgery to provide an anchorage of posterior internal fixation into vertebrae. Screw performance is highly correlated to bone quality. Therefore, polymeric cement is often injected through specifically designed perforated pedicle screws into osteoporotic bone to potentially enhance screw stability. Methods. Caudocephalic dynamic loading was applied as quasi-physiological alternative to classical pull-out tests on 16 screws implanted in osteoporotic lumbar vertebrae and 20 screws in nonosteoporotic specimen. Load was applied using 2 different configurations simulating standard and dynamic posterior stabilization devices. Screw performance was quantified by measurement of screwhead displacement during the loading cycles. To reduce the impact of bone quality and morphology, screw performance was compared for each vertebra and averaged afterward. Results. All screws (with or without cement) implanted in osteoporotic vertebrae showed lower performances than the ones implanted into nonosteoporotic specimen. Augmentation was negligible for screws implanted into nonosteoporotic specimen, whereas in osteoporotic vertebrae pedicle screw stability was significantly increased. For dynamic posterior stabilization system an increase of screwhead displacement was observed in comparison with standard fixation devices in both setups. Conclusion. Augmentation enhances screw performance in patients with poor bone stock, whereas no difference is observed for patients without osteoporosis. Furthermore, dynamic stabilization systems have the possibility to fail when implanted in osteoporotic bone. Level of Evidence: N/A

[1]  L. Benneker,et al.  DensiProbe Spine: an intraoperative measurement of bone quality in spinal instrumentation. A clinical feasibility study. , 2013, The spine journal : official journal of the North American Spine Society.

[2]  Cheng-Kung Cheng,et al.  Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. , 2013, Medical engineering & physics.

[3]  W. Hutton,et al.  A biomechanical study of two different pedicle screw methods for fixation in osteoporotic and nonosteoporotic vertebrae. , 2012, Journal of surgical orthopaedic advances.

[4]  C. Niu,et al.  Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: A comparative study between cannulated screws with cement injection and solid screws with cement pre-filling , 2011, BMC musculoskeletal disorders.

[5]  W. Hutton,et al.  Pedicle Screw Augmentation With Polyethylene Tape: A Biomechanical Study in the Osteoporotic Thoracolumbar Spine , 2010, Journal of spinal disorders & techniques.

[6]  M. Ogon,et al.  Assessment of different screw augmentation techniques and screw designs in osteoporotic spines , 2008, European Spine Journal.

[7]  M. Blauth,et al.  Influence of osteoporosis on fracture fixation - a systematic literature review , 2008, Osteoporosis International.

[8]  R. McLain,et al.  Primary Pedicle Screw Augmentation in Osteoporotic Lumbar Vertebrae: Biomechanical Analysis of Pedicle Fixation Strength , 2007, Spine.

[9]  David P Fyhrie,et al.  Biomechanical analysis of differing pedicle screw insertion angles. , 2007, Clinical biomechanics.

[10]  J. Tan,et al.  Cement Augmentation of Vertebral Screws Enhances the Interface Strength Between Interbody Device and Vertebral Body , 2007, Spine.

[11]  S. Santavirta,et al.  The microenvironment around total hip replacement prostheses. , 2005, Clinical orthopaedics and related research.

[12]  J. Tan,et al.  Pedicle Screw Motion in the Osteoporotic Spine After Augmentation With Laminar Hooks, Sublaminar Wires, or Calcium Phosphate Cement: A Comparative Analysis , 2004, Spine.

[13]  S. Cook,et al.  Biomechanical study of pedicle screw fixation in severely osteoporotic bone. , 2004, The spine journal : official journal of the North American Spine Society.

[14]  G. Andersson,et al.  Augmentation of Pedicle Screw Fixation Strength Using an Injectable Calcium Phosphate Cement as a Function of Injection Timing and Method , 2004, Spine.

[15]  R. Keen Burden of osteoporosis and fractures , 2003, Current osteoporosis reports.

[16]  W. Walsh,et al.  Hydroxyapatite Composite Resin Cement Augmentation of Pedicle Screw Fixation , 2003, Clinical orthopaedics and related research.

[17]  A. Tamura,et al.  Natural hypothermia immediately after transient global cerebral ischemia induced by spontaneous subarachnoid hemorrhage. , 2003, Journal of neurosurgery.

[18]  H. Wilke,et al.  Enhanced primary stability through additional cementable cannulated rescue screw for anterior thoracolumbar plate application. , 2003, Journal of neurosurgery.

[19]  S. Evans,et al.  Bone cement or bone substitute augmentation of pedicle screws improves pullout strength in posterior spinal fixation , 2002, Journal of materials science. Materials in medicine.

[20]  L. Gayet,et al.  Biomechanical Study and Digital Modeling of Traction Resistance in Posterior Thoracic Implants , 2002, Spine.

[21]  W. Lu,et al.  Bioactive Bone Cement as a Principal Fixture for Spinal Burst Fracture: An In Vitro Biomechanical and Morphologic Study , 2001, Spine.

[22]  A Rohlmann,et al.  Monitoring In Vivo Implant Loads With a Telemeterized Internal Spinal Fixation Device , 2000, Spine.

[23]  A Rohlmann,et al.  Influence of load carrying on loads in internal spinal fixators. , 2000, Journal of biomechanics.

[24]  A. Rohlmann,et al.  Loads on internal spinal fixators measured in different body positions , 1999, European Spine Journal.

[25]  T. Washio,et al.  Structural Characteristics of the Pedicle and Its Role in Screw Stability , 1997, Spine.

[26]  Tait S. Smith,et al.  The Effect of Bone Quality on Pedicle Screw Loading in Axial Instability: A Synthetic Model , 1997, Spine.

[27]  T. Washio,et al.  An Experimental Study of a Combination Method Using a Pedicle Screw and Laminar Hook for the Osteoporotic Spine , 1997, Spine.

[28]  A Rohlmann,et al.  Loads on an internal spinal fixation device during walking. , 1997, Journal of biomechanics.

[29]  S. Cook,et al.  Effects of Bone Mineral Density on Pedicle Screw Fixation , 1994, Spine.

[30]  F. Kummer,et al.  Hydroxyapatite Enhancement of Posterior Spinal Instrumentation Fixation , 1994, Spine.

[31]  K. An,et al.  Pull-out strength of screws from polymethylmethacrylate cement. , 1994, The Journal of bone and joint surgery. British volume.

[32]  A. Tencer,et al.  Caudo-cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentation. , 1993, Spine.

[33]  W. Hayes,et al.  Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. , 1993, Clinical orthopaedics and related research.

[34]  J N Weinstein,et al.  Anatomic and technical considerations of pedicle screw fixation. , 1992, Clinical orthopaedics and related research.

[35]  J. Corin,et al.  Biomechanical Analysis of Pedicle Screw Instrumentation Systems in a Corpectomy Model , 1989, Spine.