Dosimetric consideration for patients with dental filling materials undergoing irradiation of oral cavity using RapidArc: challenges and solution

In this study, we investigate the effect of dental filling materials (DFM) on RapidArcTM treatment plans and delivery in a patient undergoing radiotherapy treatment. The presence of DFM creates uncertainties in CT number and causes long streaking artifacts in the reconstructed images which greatly affect the dose distribution inside the oral cavity. The influence of extensive dental filling artifacts on dose distribution was performed using a geometrically well defined head and neck IMRT verification phantom (PTW, Freiburg, Germany) together with inserts from DFM (Amalgam, 11.3 g/cm3). The phantom was scanned using Siemens SOMATOM Sensation CT simulator (Siemens AG, Germany) under standard head and neck imaging protocol (120 kV, 120 mAs, voxel size 1×1×2 mm3). Three RapidArcTM plans were created in the Varian Eclipse treatment planning System (TPS) to treat oral cavity using the same CT dataset including; 1) raw CT image, 2) streaking artifacts replaced with a mask of 10 HU and 3) 2 cm thick 6000 HU virtual filter (a volume around the teeth in TPS to mimic extra attenuation). The virtual filter thickness optimization was purely based on measured PDD data acquired with DFM and the calculation in Eclipse Planning System using direct beam. The dose delivery and distribution for the three plans was verified using Gafchromic EBT2 (International Specialty Product, Wayne, NJ, USA) film measurements. The artifact mask and virtual filter around the teeth in the planning was found very useful to reduce the discrepancies between the dose plan and delivery. From clinical point of view, these results can be helpful to understand the increase of mucositis in patient having DFM, and further investigation is underway for clinical solution.

[1]  A. Boyer,et al.  Tissue Inhomogeneity Corrections for Megavoltage Photon Beams , 2004 .

[2]  A. Niroomand-rad,et al.  Radiation dose perturbation at tissue-titanium dental interfaces in head and neck cancer patients. , 1996, International journal of radiation oncology, biology, physics.

[3]  A. Murthy,et al.  Dose perturbation resulting from gold fillings in patients with head and neck cancers. , 1979, International journal of radiation oncology, biology, physics.

[4]  P. Biggs,et al.  Effect of a femoral head prosthesis on megavoltage beam radiotherapy. , 1988, International journal of radiation oncology, biology, physics.

[5]  R. Jacob,et al.  Backscatter radiation at bone-titanium interface from high-energy X and gamma rays. , 1987, International journal of radiation oncology, biology, physics.

[6]  Luc Beaulieu,et al.  Correction of CT artifacts and its influence on Monte Carlo dose calculations. , 2007, Medical physics.

[7]  G. Ibbott,et al.  Hip prostheses during pelvic irradiation: effects and corrections. , 1988, International journal of radiation oncology, biology, physics.

[8]  P. Keall,et al.  Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63. , 2003, Medical physics.

[9]  M. Carolan,et al.  Effect of hip prostheses on radiotherapy dose. , 2000, Australasian radiology.

[10]  Vladimir Pekar,et al.  Segmentation-aided adaptive filtering for metal artifact reduction in radio-therapeutic CT images , 2004, SPIE Medical Imaging.

[11]  R Roberts,et al.  How accurate is a CT-based dose calculation on a pencil beam TPS for a patient with a metallic prosthesis? , 2001, Physics in medicine and biology.

[12]  R. Wang,et al.  Dosimetric measurement of scattered radiation from dental implants in simulated head and neck radiotherapy. , 1998, The International journal of oral & maxillofacial implants.

[13]  C. Sibata,et al.  Influence of hip prostheses on high energy photon dose distributions. , 1990, International journal of radiation oncology, biology, physics.

[14]  G. Ding,et al.  A study on beams passing through hip prosthesis for pelvic radiation treatment. , 2001, International journal of radiation oncology, biology, physics.